Statement of Work for Preventative Maintenance of Microscopes.pdf

PDF 68 KB Posted

Attached to
Preventative Maintenance of Microscopes
Solicitation number
1333MG26Q0007
Issued by

About this file

This is a Statement of Work for annual preventative maintenance services for microscopes at the NOAA Research Laboratory in Charleston, South Carolina.

The contractor shall perform preventative maintenance and alignment on twenty-nine (29) microscopes, including multiple units of Nikon, Olympus, Zeiss, Motic, and Leica models. Minimum requirements encompass comprehensive optical and mechanical evaluation, cleaning of all optical components, collimation to factory specifications, disassembly and lubrication of adjustment assemblies, cleaning and lubrication of all moving parts and slideways, alignment of phase rings and condensers, illuminator inspection and cleaning, nosepiece and stage mechanism servicing, focus mechanism overhaul as needed, and alignment of light sources. The contractor must maintain detailed office records by serial number and component information for each microscope and provide individual service records. The period of performance consists of a base year from September 30, 2026 through September 29, 2027, followed by four optional years extending through September 29, 2031. IT security requirements note that Assessment and Authorization requirements under 48 CFR 1352.239-72 do not apply and no Security Accreditation Package is required. For technical questions and to schedule laboratory access, contractors should contact Paul Pennington at 843-593-0043 or Katy Chung at 843-709-6490.

View the file

Other files

File Type Posted
Amendment 0001.pdf PDF
QUESTIONS AND ANSWERS - Amendment 0001.pdf PDF
NEW MICROSCOPE LIST - Amendment 0001.pdf PDF
Statement of Work - Amendment 0001.pdf PDF
Sol_1333MG26Q0007.pdf PDF
WH1313.pdf PDF
SAM.gov Wage Determination.pdf PDF

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Statement of Work for Preventative Maintenance of Microscopes

The NOAA Research Laboratory at 331 Fort Johnson Road, Charleston S.C. has a need for annual

Preventative Maintenance (PM) of twenty-nine (29) microscopes listed below:

1 each PM and Alignment of a Nikon Eclipse 55i Brightfield and Fluorescence

1 each PM and Alignment of a Motic Panthera C Trinocular Compound LED Microscope

1 each PM and Alignment of Zeis Primo Star iLED Brightfield and Fluorescence

1 each PM and Alignment of Zeiss Stemi 508 Stereo Microscope with K EDU

1 each PM and Alignment of Olympus SZ61 Stereo Microscope

1 each PM and Alignment of Olympus MVX10 Macro Zoom Fluorescence Microscope

1 each PM and Alignment of Olympus CHS Biological Microscope

1 each PM and Alignment of Nikon SMZ1500

2 each PM and Alignment of Olympus BX51

2 each PM and Alignment of Ziess Inverted

1 each PM and Alignment of Leica Stereo

2 each PM and Alignment of Olympus Stereo

1 each PM and Alignment of Olympus BX60 with Phase and Fluorescence

1 each PM and Alignment of Olympus SZH

1 each PM and Alignment of Nikon TS100 with Fluorescence

1 each PM and Alignment of Nikon E400

4 each PM and Alignment of Olympus SZ61

1 each PM and Alignment of Olympus SZ50

5 each PM and Alignment of microscope

Scope: The contractor shall perform preventative maintenance on the above twenty-nine (29) microscopes which shall include the following minimum requirements:

• Evaluate optical and mechanical performance prior to any service.

• Clean all optical components including mirrors, prisms, objectives, observation tubes,

• Check for damage to objective front lenses.

• Collimate stage, sub-stage and lamp house to factory specifications. Check moving parts and fine focusing mechanism.

• Disassemble, clean, repack and lubricate coarse and fine adjustment assembly

• Clean and lubricate all slideways and moving parts

• Clean, align and lubricate sub-stage condensers.

• Check co axle X, Y coordinate stage mechanism. Disassemble and lubricate as needed.

• Disassemble, clean, and lubricate nosepiece carriage assembly

• Remove, clean, and align all phase rings, condensers and diaphragms to factory specifications.

• Remove and clean illuminator, check for wiring condition, clean filters and mirror and insure correct functions of field diaphragms as necessary.

• Remove bodies and check for cleanliness of prisms, clean and lubricate interpupillary adjustment, check for correct prism alignment.

• Reassemble all operating components. Check for parcentering of condenser, nosepiece, and binocular body. Insure parcentering of low through high power objectives.

• Check for and adjust, as needed, parfocality between objectives.

• Set all stops to ensure maximum performance and prevent objective or specimen damage.

• Align halogen, mercury or xenon light sources as necessary.

• Coarse and Fine focus overhaul as needed.

• Clean outside area of microscope.

• Maintain office records of microscopes by serial number of components including lamp type to aid in minimal down time should repair be needed.

• Provide individual service records on microscopes

Period of Performance:

Base Yr: September 30, 2026 – September 29, 2027 Opt Yr 1: September 30, 2027 – September 29, 2028 Opt Yr 2: September 30, 2028 – September 29, 2029 Opt Yr 3: September 30, 2029 – September 29, 2030 Opt Yr 4: September 30, 2030 – September 29, 2031

IT Security Requirements:

The Assessment and Authorization (A&A) requirements of Clause 48 CFR 1352.239-72 do not apply, and a Security Accreditation Package is not required.

For technical questions, and to schedule Lab access for service, contact Paul Pennington, 843-593-0043

OR Katy Chung at cell 843-709-6490.

File details come from the government source that posted it. Updated .