Statement of Work for LabRAM IIH and RAM.docx
DOCX document 25 KB Posted
- Solicitation number
- N00174-25-SIMACQ-M4-0027
About this file
This Statement of Work (SOW) details maintenance and repair services for two Resodyn mixers at Naval Surface Warfare Center Indian Head Division (NSWC IHD). The government intends to sole-source a firm fixed-price contract to Resodyn Acoustic Mixers for servicing a LabRAM IIH and RAM-5 mixer, which have experienced functional issues and lack recent maintenance. Specific requirements include diagnostic work, power and vacuum system repairs, replacement of components like accelerometers and hydraulic hoses, operational checks of multiple systems, and calibration of accelerometers.
The contract has a three-month performance period and will be executed at NSWC IHD facilities in Indian Head, MD. The procurement is being conducted under FAR 13.106(b) as a single-source acquisition, with Resodyn identified as the only vendor capable of meeting requirements. Interested parties may submit alternative capabilities by May 20, 2025, at 8:00 AM EST to Natella Calliste, with responses evaluated to determine potential competitive procurement. The contract's NAICS code is 334513, and the resulting agreement will be a firm fixed-price contract negotiated using non-competitive procedures under 10 U.S.C. 2304(c)(1).
View the file
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Statement of Work for LabRAM II H and RAM-5 Service
1.0 Introduction
The Naval Surface Warfare Center Indian Head Division (NSWC IHD) requires maintenance and repair services on Resodyn LabRAM IIH and RAM-5 mixers. The LabRAM IIH and RAM-5 mixers are currently used for energetic mixing, as a quicker and safer alternative to conventional mixing. The LabRAM IIH has had some issues with vacuum function and after attempting to fix, there was no power to the mixer and could not be fixed by NSWC IHD personnel. The contractor will determine what is wrong with the LabRAM IIH. After fixing the issue, preventative maintenance will be performed on the mixer. Additionally, the RAM-5 mixer has not been serviced since 2019 and Resodyn personnel will perform a diagnostic check and perform necessary preventative maintenance. NSWC Indian Head will provide an electrician to assist Resodyn Acoustic Mixers Incorporated.
2.0 Scope
The Contractor shall identify and perform required maintenance and repair services on two identified Resodyn mixers.
Equipment: LabRAM IIH (P/N H100000140) and RAM-5 (RAM 5020 P/N 902600) Purpose:
1. To perform diagnostic work on the LabRAM IIH.
2. To resolve all LabRAM IIH power and vacuum issues.
3. To perform preventative maintenance on the LabRAM IIH and RAM-5 mixers.
4. Provide recommendations for any additional maintenance needed.
End Expectations:
1. Restore LabRAM IIH functionality with power and ability to run mixer.
2. Restore LabRAM IIH vacuum function.
3. Resodyn to train engineers and operators on vacuum best practices and troubleshooting.
4. Ensure LabRAM IIH and RAM-5 preventative maintenance completed, and mixer status is up to date.
5. Resodyn to provide a list of any further maintenance recommendations after inspection of LabRAM IIH and RAM-5 mixers.
3.0 Requirement
The Contractor shall perform the following services for each of the mixers listed below:
LabRAM II H (P/N H100000140)
· Install replacement power supply
· Install replacement Accelerometer
· Install replacement Bellows
· Operational check of Purge System
· Operational check of Chiller System
· Operational check of HMI
· Operational check of Vacuum System
· Operational check of Resonator and Springs
· Operational check of Vessels and Vessel Holder
· Calibration of Accelerometer
RAM-5 (RAM 5020 P/N 902600)
· Change Oil, Filter, and Breather
· Install replacement Accelerometer
· Install replacement Z Accelerometer
· Install replacement Hydraulic Hoses
· Install replacement JV Parts
· Install replacement bellows
· Operational check and Inspection of the Purge System
· Operational check and Inspection of the Chiller System
· Operational check and Inspection of the Electrical System
· Operational check and Inspection of the Vacuum System
· Operational check and Inspection of the Vessels
· Calibration of Accelerometer
4.0 Security Requirements
The Contractor shall coordinate base access in advance of scheduled Maintenance with the Technical Point of Contact.
5.0 Place of Performance
Naval Surface Warfare Center, Indian Head Division M Department
BLDG 1119
Indian Head, MD 20640
M Department
BLDG 905
Indian Head, MD 20640
6.0 Period of Performance
The Period of Performance shall be three months from the date of award.
7.0 Technical Point of Contact