This statement of work outlines the requirements for annual sterilizer chamber cleaning services at multiple medical facilities for the Defense Health Agency. The contractor shall provide cleaning services for nine sterilizers located at three sites, including the main operating room and SPD at Navy Medicine Readiness and Training Command Lemoore and dental clinics at NMRTC Lemoore and Naval Air Station Fallon. Services include professionally cleaning the autoclaves and sterilization units on an annual basis to remove mineral buildup and contaminants, meeting all regulatory standards for chamber cleaning. The contractor must notify facilities before arriving and use only manufacturer-approved cleaning solutions, equipment, and parts. All labor, travel expenses, and replacement of parts and supplies are the responsibility of the contractor. Work will not be performed on ten specified federal holidays.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| Combined Synopsis - Sterilizer Cleaning HT941024Q2008.pdf |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
STATEMENT OF WORK
1.0 GENERAL REQUIREMENTS
1.1 The effort required hereunder shall be performed in accordance with this Statement of Work and in accordance with all other terms and conditions set forth herein.
1.2 The contractor shall provide all services, materials, and equipment necessary for services to provide annual sterilizer chamber cleaning for the Main Operating Room and SPD for NMRTC Lemoore and all the 2 Dental clinics locations under NMRTC Lemoore Area of Responsibility. The scope of work performed under these specifications includes the furnishing of all labor, and parts to perform all services listed to ensure continued operation at their designed efficiency and capacity.
2.0 BACKGROUND
2.1 Currently, NMRTC Lemoore MOR and 2 outlying Dental clinics (listed below) are equipped with Steam Sterilizers (Autoclaves).
The Steam Sterilizers are typically used for healthcare or industrial applications and are machines that uses steam under pressure to kill harmful bacteria, viruses, fungi, and spores on items that are placed inside a pressure vessel. The basic principle of steam sterilization, as accomplished in an autoclave, is to expose each item to direct steam contact at the required temperature and pressure for the specified time.
Thus, there are four parameters of steam sterilization: steam, pressure, temperature, and time. The ideal steam for sterilization is dry saturated steam and entrained water (dryness fraction ≥97%). Pressure serves as a means to obtain the high temperatures necessary to quickly kill microorganisms. Specific temperatures must be obtained to ensure the microbicidal activity.
The two common steam-sterilizing temperatures are 121°C (250°F) and 132°C (270°F). These temperatures (and other high temperatures) must be maintained for a minimal time to kill microorganisms. Recognized minimum exposure periods for sterilization of wrapped healthcare supplies are 30 minutes at 121°C (250°F) in a gravity displacement sterilizer or 4 minutes at 132°C (270°F) in a prevacuum sterilizer. At constant temperatures, sterilization times vary depending on the type of item (e.g., metal versus rubber, plastic, items with lumens), whether the item is wrapped or unwrapped, and the sterilizer type.
The two basic types of steam sterilizers (autoclaves) are the gravity displacement autoclave and the high-speed prevacuum sterilizer.
In the former, steam is admitted at the top or the sides of the sterilizing chamber and, because the steam is lighter than air, forces air out the bottom of the chamber through the drain vent. The gravity displacement autoclaves are primarily used to process laboratory media, water, pharmaceutical products, regulated medical waste, and nonporous articles whose surfaces have direct steam contact.
For gravity displacement sterilizers the penetration time into porous items is prolonged because of incomplete air elimination. This point is illustrated with the decontamination of 10 lbs. of microbiological waste, which requires at least 45 minutes at 121°C because the entrapped air remaining in a load of waste greatly retards steam permeation and heating efficiency
3.0 SERVICES
3.1 The vendor shall professionally clean all the autoclaves and other units that’s listed on an annually basis that meets all regulatory compliance for annual chamber cleaning standards. The accumulation of minerals, and other contaminants on the sterilization process on chamber walls from the incoming steam that builds up and cause discoloration and can hinder performance or even contaminate sterilized packs should be removed to the best of the contractor’s ability. The company must provide a 24-hour phone call and/or email to make sure any autoclave that is to be cleaned is turned off at the end of the day, before the cleaning technician arrives. All cleaning solutions, equipment, and/or parts should be approved for use by the sterilizers manufacturer and provided by the contractor to perform cleaning.
3.2 Steam Sterilizers (Autoclaves) and other sterilization units will require cleaning at all but not limited to the following locations:
3.3 Navy Medicine Readiness and Training Command Lemoore
937 Franklin Avenue, NAS Lemoore, CA 93246
3.3.1 Annual Sterilizer Cleaning for the following Sterilizers
3.3.2 Sterilizer, Amsco 400 Serial Number: 0330314-14
3.3.3 Sterilizer, Amsco 400 Serial Numbers: 030241910
3.3.4 Sterilizer, Amsco 400 Serial Numbers: 030241911
3.3.5 Sterilizer, Amsco V116 Serial Number: 0329911-03
3.4 Navy Medicine Readiness and Training Command Lemoore (Dental Clinic)
926 Franklin Avenue, NAS Lemoore, CA 93246
3.4.1 Annual Sterilizer Cleaning for the following Sterilizers
3.4.2 Sterilizers, Amsco 400 Serial Numbers: 0326814-04
3.4.3 Sterilizers, Amsco 400 Serial Numbers: 0327314-10
3.5 Naval Air Station Fallon Dental Clinic
4755 Pasture Rd. Building #299, Fallon, NV 89496
3.5.1 Annual Sterilizer Cleaning for the following Sterilizers
3.5.2 Sterilizer, Century Serial Number: 0117405-13
3.5.3 Sterilizer, Castle/Getinge 433 Serial Number: URA013127
3.6 The contractor is not required to provide service on the following days:
3.6.1 NEW YEARS DAY
3.6.2 MARTIN LUTHER KING, JR. BIRTHDAY
3.6.3 PRESIDENT’S DAY
3.6.4 MEMORIAL DAY
3.6.4 JUNETEENTH DAY
3.6.5 INDEPENDENCE DAY
3.6.6 LABOR DAY
3.6.7 COLUMBUS DAY
3.6.8 VETERANS DAY
3.6.9 THANKSGIVING DAY
3.6.10 CHRISTMAS DAY
3.11 The Vendor shall provide notice telephone call and/or via e-mail to Technical Point of Contact (listed below) within 24 hours of completion of service, along with a checklist showing completion of each task. The Vendor must be professional, clean in appearance, wear ID badge as required and must Sign In/Out at the Quarter Deck log during each visit.
4.0 COMPLIANCE WITH APPLICABLE LAWS AND STANDARDS
4.1 The vendor shall comply with all Federal, State and Local laws, regulations, and governmental orders, and services pursuant to this Agreement. The vendor shall comply with all applicable standards established by the Joint Commission.
The vendor’s staff, affiliate personnel and any other subcontractors providing services shall be properly trained and credentialed and properly reviewed according to the requirements set forth by the Joint Commission.
5.0 PARTS/SUPPLIES
5.1 All maintenance items and supplies must be replenished automatically or as needed. Shipping and handling must be the vendor’s responsibility. Parts and supplies provided under this service contract shall be guaranteed to be equal in all respects, including performance, interchangeability, durability, and quality to the OEM parts when new or as presently recommended by the contractor.
6.0 LABOR
6.1 All compensation for labor, parts and materials must be included in the contract price.
7.0 GOVERNMENT PERSONNEL
7.1 NMRTC Lemoore employees will not perform or attempt to perform any cleaning of the autoclaves.
8.0 TRAVEL AND RELATED EXPENSES
8.1 The contractor is responsible for its related trip expenses including round trip travel, mileage, and all other necessary expenses related to travel. The contractor must be escorted on the base of NMRTC Lemoore and Naval Air Station Fallon.