This is a Statement of Requirement (SOR) for the procurement of an Audiology Vestibular Testing Rotary Chair by the Naval Medical Center Camp Lejeune (NMCCL) Audiology Department. The requirement specifies one (1) Vestibular Rotary Chair with Videonystagmography (VNG) option and Air Caloric Addition, including a computer, TV monitor, ocular motor stimulus, and software. The equipment will be stationary and hardwired for electrical connection. The chair is intended to evaluate and identify inner ear and neurological balance disorders as part of comprehensive vestibular evaluations. The government also requires installation labor, one day of on-site training, and removal and disposal of the existing Neuro-Kinetics Chair/System.

The technical specifications require equipment equivalent to or branded as the Orion Auto-Traverse Vestibular Rotary Chair with VNG, with minimum specifications including a RCS-365 missile-grade torque motor with 365 ft-lbs peak torque, maximum angular velocity of 400°/sec, and maximum acceleration of 1200°/sec². The rotational motion system must include dual-axis tilt (LARP/RALP) with tilt speeds of 10-15 seconds for full reposition and maximum tilt angles of 45°. The eye tracking system must feature I-Portal Infrared Video-Oculography (VOG) with a 100 Hz sampling rate (standard) and 250 Hz high-speed mode capability, with specified horizontal and vertical spatial resolution requirements. Position accuracy must be within 0.2°, encoder resolution of 20,000 counts/rev, and gimbal configuration must be fully motorized with software-controlled dynamic repositioning to align with specific canal planes.

View the file

Other files

File Type Posted
Solicitation - HT940626QE055.pdf PDF
Volume I _TC _Company Name.docx DOCX document

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Statement of Requirement (SOR) Project Title: Audiology Vestibular Testing Rotary Chair Location: Naval Medical Center Camp Lejeune, Audiology Department, NC Quantity: One (1)

1. Scope and Functionality Background: The Audiology Department at Naval Medical Center Camp Lejeune (NMCCL) requires one (1) Vestibular Rotary Chair with the Videonystagmography (VNG) option with Air Caloric Addition, Computer, TV Monitor, Ocular Motor Stimulus and Software. Labor, installation and training required for the new rotary chair.

Feature
Requirement Detail
Primary Function
Evaluates and identifies inner ear and neurological balance disorders as a critical component of a comprehensive vestibular evaluation.
Mobility
Stationary. Electrical Hardwired.
Software
This is a standalone manufactured software system compatible with either chair.
Installation
Vestibular Rotary Chair
Training
One day on-site
Removal
Removal and disposal of current Neuro-Kinetics Chair/System

2. Technical Specifications The equipment must include, at a minimum, the following hardware and software components for Brand Name or Equivalent to Orion Auto-Traverse Vestibular Rotary Chair with the VNG:

Minimum Specification

Rotational Motion System
● Motor Type: RCS-365 missile-grade torque motor

● Peak Torque: 365 ft-lbs

● Maximum Angular Velocity: 400°/sec

● Maximum Acceleration: 1200°/sec²

● Position Accuracy: ±0.2°

● Encoder Resolution: 20,000 counts/rev

Tilt System
● Axis: Dual-axis tilt (LARP/RALP)

● Tilt Speed: 10–15 seconds full reposition

● Max Tilt Angle: ±45°

● Positional Repeatability: ±1.5°

Gimbal Configuration:
● Fully motorized, software-controlled

● Dynamic repositioning to align with specific canal planes

Eye Tracking System: I-Portal® Infrared Video-Oculography (VOG)

● Sampling Rate: 100 Hz (standard); 250 Hz (high-speed mode)

● Spatial Resolution: ○ Horizontal/Vertical: <0.1°

● Torsional: <0.5°

● Eye Coverage: Binocular tracking

● Slip Ring: Fiber-optic rotary transmission for uninterrupted signal

Visual Stimulus Display:
● Display Type: VR-style integrated headset

● Resolution: 1440 x 2560 per eye

● Refresh Rate: 60 Hz

● Visual Field: >100°

● Diopter Adjustment: ±6D

● Calibration: Auto or manual 5-point

Patient Accommodation:
● Weight Capacity: Up to 400 lbs (181 kg)

● Dimensions: Fits 3rd–97th percentile adult patients

● Pediatric Use: Compatible with child restraint insert

Safety Systems:
● 5-point Harness System

● Adjustable Head Holder: Lateral and vertical positioning

● Leg Stabilization: Adjustable separators and straps

● Emergency Stop: Operator and patient accessible

● Real-Time Monitoring Camera: 8” LCD interface for clinician

● Two-Way Audio: Secure wireless headset system

Software Platform:

● Protocol Types: SHA, VOR, VSR, COR, OKN, SOT, unilateral centrifugation, etc.

● Pre-programmed and customizable protocols (40+)

● SQL Database with encrypted storage

● Data Export: PDF, CSV, HL7-compatible

● Interoperability: Integrates with Dx200/300 goggles for multi-device networks

User Interface:

● Control Terminal: Industrial touchscreen workstation

● Connectivity: USB 3.0, Ethernet

● Update Method: Secure OTA (over-the-air) firmware & protocol updates

3. Environmental and Electrical Requirements

Parameter
Specification
I-Portal NOTC Pro+ system
The system requires an electrical service disconnect that must be a 20A fused disconnect fused for 20A. A L15-30R receptacle is necessary within 3ft of the planned location of the electrical console. Hardwire directly to the electrical console.
Voltage
208v
Phase
Triple Phase (Phase 3)
Room Size
17ft x 11ft

4. Sourcing, Warranty, and Maintenance

Category
Requirement
Authorized Sourcing
All equipment must be new. Reconditioned or used equipment is not acceptable. Equipment must be sourced directly from the Original Equipment Manufacturer (OEM) or an OEM-authorized dealer/distributor.
Verification
Documentation verifying authorized dealer/distributor status must be provided to the Government upon request.
Preventive Maintenance
The vendor is responsible for all preventive maintenance and service in accordance with the OEM schedule until the expiration of the warranty period.
Service Scope
Maintenance includes all inspections, calibrations, parts replacement, and software/firmware updates required to keep the unit in a manufacturer-supported condition.
Parts & Tools
All maintenance actions must be performed using OEM-approved parts and diagnostic tools.
Warranty
Shall provide a standard Original Equipment Manufacturer (OEM) warranty for all delivered products.

5. Training Requirements

Requirement
Description
Format
In-person training.
Timing
To be conducted at the time of installation.
Objective
Familiarization with hardware components and navigation of new software settings for clinical staff.

6. Cybersecurity and Information Technology This equipment is intended for standalone operation to ensure mission security and patient data integrity without reliance on external network infrastructure.

Requirement
Description
Network Connectivity
The unit shall not be integrated with any internal government network systems. It must function entirely as a standalone device.
Data Export
The system must be capable of exporting completed examination results to a CD-ROM in a PDF summary format.
Remote Viewing
Remote reviewing/viewing capabilities must be facilitated via the dedicated offline viewing station or secure media transfer (CD-ROM).
Authorization to Operate (ATO)
Because this equipment will not be maintained on or connected to a government network, it does not require an ATO.

7. Delivery, Installation, and Acceptance All components of the Vestibular Rotary Chair/System must be delivered and fully operational within the specified period following the contract award to ensure no lapse in audiology diagnostic capabilities.

Milestone
Requirement / Timeline
Delivery Location
Naval Medical Center Camp Lejeune (NMCCL)Attn: Audiology Department 100 Brewster Blvd, Camp Lejeune, NC 28547
Base Access
Personnel Identification- All vendor personnel must present a valid U.S. Passport or a REAL ID-compliant driver’s license/ID card to gain base access. Vehicle Access- Vehicles must be registered to the vendor/delivery company with current proof of insurance. If you use rental, a valid rental agreement must be presented. Failure to meet these security and access requirements will result in denied base entry and delivery delays, for which the vendor is entirely responsible.
Delivery Timeline
All equipment shall be delivered within 60 days after contract award.
Installation
On-site installation and configuration must be completed by the vendor within 5 business days of delivery.
Coordination
The vendor must coordinate delivery and installation dates at least 48 hours in advance with the NMCCL Point of Contact (POC) and Biomedical Engineering (BIOMED) to ensure site readiness.
Final Acceptance
The Government will have 10 business days following installation and the initial training session to inspect the equipment and verify full functionality before final acceptance is granted.