6 March 2018
NRL 5524
PURCHASE DESCRIPTION
Multifunction Microwave Analyzer
Amendment
GMG: MWHHAN-18B
Solicitation No: SPRMM1-18-R-HA24
1.0 GENERAL The following Purchase Description (PD) identifies the requirements for a hand held multifunction microwave analyzer capable of performing the functions of a spectrum analyzer, signal source, vector network analyzer, cable and antenna analyzer, vector voltmeter, interference analyzer, channel scanner, and power meter up to a frequency of 40 GHz. .
2.0 CLASSIFICATION This equipment shall meet the requirements of MIL-PRF- 28800F class 2 for Navy shipboard, submarine, shore, and Marine Corps applications.
3.0 OPERATIONAL REQUIREMENTS The specifications provided below are the minimum requirements and accuracies that will meet the Government's need. The analyzer shall at least meet these requirements and accuracies and may provide better performance. All accessories, adapters, and cables necessary for performance stated in section 3 (example: calibration kit) shall be included.
3.1 Common Specifications (unless otherwise specified):
3.1.1 Input/output connectors: The connectors shall be a native 2.4 mm (m) or 2.92 (m). If
2.4 mm, adapter to 2.92 (m) shall be provided.
3.1.2 Maximum Input level: +23 dBm with no external attenuation over full frequency range.
25 W continuous with external attenuation from DC to 18 GHz (attenuators shall be included), performance specifications below not applicable when attenuator in use.
3.1.3 Range: 2 MHz to 40 GHz
3.1.4 System Impedance: 50 ohms nominal.
3.2 Spectrum Analyzer
3.2.1 Frequency:
3.2.1.1 Range: 100 kHz to 40 GHz
3.2.1.2 Accuracy: ±1 ppm
3.2.2 Amplitude:
3.2.2.1 Measurement Range: DANL to +20 dBm
3.2.2.2 DANL (at RBW of 1 kHz and frequency of 2 MHz – 40 GHz): < -105 dBm
3.2.2.3 Frequency response: +/-2 dB across frequency range.
3.2.3 Markers: Two independent markers
3.2.3.1 Single mode: One marker displays absolute frequency & amplitude.
3.2.3.2 Delta mode: Two markers display frequency and amplitude difference between two markers.
3.2.3.3 Resolution: Frequency 1 Hz, amplitude 0.1 dB.
3.2.3.4 Accuracy: Frequency ±1 ppm, amplitude ±1 dB.
3.2.3.5 Peak search: automatic amplitude peak search function.
3.2.4 Span Modes: 0 Hz (zero span) and settable from 10 Hz to 40 GHz.
3.2.5 Resolution Bandwidth: Selectable from at least 30 Hz to 3 MHz.
3.2.6 Video Bandwidth: Selectable from at least 10 Hz to 3 MHz.
3.3 Source
3.3.1 Modes:
3.3.1.1 Tracking generator (swept frequency).
3.3.1.2 CW (Source frequency is independent of the receiver frequency).
3.3.2 Frequency Range: 2 MHz to 40 GHz.
3.3.3 Amplitude:
3.3.3.1 High Power: ≥-3 dBm typical.
3.3.3.2 Low Power: ≤-20dBm typical.
3.4 Network Analyzer, full 2-Port S Parameters
3.4.1 Frequency Range: 1 MHz to 40 GHz
3.4.2 Frequency Accuracy: ±1 ppm
3.4.3 High Test Port Power Out: ≥-3 dBm
3.4.4 Low Test Port Power Out: ≤-20dBm typical
3.4.5 Measurement Speed: ≤650 μs/pt
3.4.6 Noise Magnitude: <0.01 dB
3.4.7 System Dynamic Range: ≥85 dB
3.4.8 Uncertainty S21, S12 at -40dB Magnitude: <0.3 dB; <2°
3.4.9 Uncertainty S11, S22 at -5dB Magnitude: <0.55 dB; <5°
3.4.10 Maximum Input level (port 1 and port 2): +23 dBm
3.5 Cable & Antenna Analyzer
3.5.1 Frequency Range: 1 MHz to 40 GHz
3.5.2 Dynamic Range: ≥85 dB
3.5.3 Return Loss: 0 to 100 dB with 0.01 dB resolution
3.5.4 VSWR: 0 to 100 with 0.01 resolution
3.5.5 Measurements: Return loss, VSWR, insertion loss/transmission, one-port cable loss, and distance-to-fault (DTF).
3.6 Power Meter
3.6.1 Internal:
3.6.1.1 Internal channelized power measurement capability using the input connector, no external sensor allowed.
3.6.1.2 Power accuracy: ±2 dB.
3.6.2 External:
3.6.2.1 Equipment shall be capable of using external USB power sensors over a frequency range of 50 MHz to 40 GHz (external USB sensor is not required).
3.7 Vector Voltmeter
3.7.1 A vector voltmeter function with the capability of measuring phase shift and electrical length of a device, ratio of magnitude and phase of two channels A/B and B/A, transmission and reflection.
3.7.2 Frequency Range: 300 kHz to 40 GHz
3.7.3 Ratio Accuracy (A/B and B/A): ±1 dB
3.8 Interference Analyzer
3.8.1 An interference analyzer function that provides the identification of interfering signals using a spectrogram or waterfall display and record and playback capability to allow signal capture and offline processing.
3.8.2 Markers: Time and delta time.
3.9 Channel Scanner
3.9.1 The capability to perform channel power measurements to verify signal coverage, identify potential interference issues and optimize network performance. Each channel shall be settable with different frequencies and bandwidths for multiple signal analysis.
3.9.2 Trace Storage/Playback: Records channel power measurement and stores internally.
Capability to recall and playback.
3.10 Display Minimum 6.5 inch color VGA/LED backlit.
4.0 GENERAL REQUIREMENTS
4.1 Warm Up Time: The warm up time shall not exceed 60 minutes.
4.2 Temperature
4.2.1 Operating Temperature: -10°C to +50°C at non-condensing RH up to 95%
4.2.2 Non-operating Temperature: -51°C to +71°C at non-condensing RH up to 95%
4.3 Power Source
4.3.1 Internal DC Battery: Internal rechargeable batteries and a charger shall be provided.
The batteries shall be installed in an externally accessible compartment as specified in MIL-PRF-28800F.
4.3.1.1 Battery operation: Minimum operating time shall be 3.5 hours following a maximum recharge time of 4 hours. The test set shall be operational while being charged.
4.3.1.2 Charger: The charger shall operate from the 50 Hz and 60 Hz power source as specified in MIL-PRF-28800F.
4.3.2 Battery Status Indicator: A battery life indicator shall be incorporated into the display.
4.3.3 Auto Power Off/Standby: The test set shall have an automatic turn off or manual standby feature. The feature shall be activated when the absence of keyboard activity exceeds an interval of 10 to 30 minutes or when manually activated.
4.3.4 Battery Restrictions Per MIL-PRF-28800F, Lithium and Mercury batteries are prohibited without prior authorization. If equipment qualifies for one or both of the exceptions stated below, then the offeror shall submit the required information at the time of submission of proposals. If equipment does not qualify for an exception, then a request for approval for the use of Lithium and Mercury batteries shall be submitted with production lot delivery, after contract award. Approval shall apply only to the specific model proposed.
Exception 1: Per NAVSEA S9310-AQ-SAF-010, the use of non-rechargeable lithium coin cell batteries meeting the following criteria is approved for all uses and does not require individual testing and review by the Naval Ordnance Safety and Security Activity (NOSSA).
- Unmodified, commercially available off-the-shelf (COTS) item;
- Used in single-cell configuration;
- Maximum nominal output of 3 Volts;
- Maximum rated capacity of 1 Ah.
The coin cell manufacturer, part number, nominal voltage, and rated capacity shall be provided at the time of submission of proposals.
Exception 2: Per NAVSEA S9310-AQ-SAF-010, the use of COTS electronics and equipment powered by lithium ion rechargeable (secondary) batteries meeting the following criteria is approved for all uses and does not require individual testing and review by NOSSA.
- Unmodified, COTS battery;
- Underwriters Laboratories (UL) listed;
- Used in the device as recommended by the manufacturer. Modifications to the devices may only be made in accordance with the manufacturer’s recommendations;
e.g., addition of memory;
- Recharged only by devices expressly designed for recharge of the specific battery in use;
- No more than four cells in series (less than or equal to 18-Volt output);
- Rated for no more than 100 Watt-hours, as listed in the manufacturer’s specification or calculated by multiplying the capacity in Ah by the maximum working (nominal performance) voltage.
The lithium ion rechargeable cell manufacturer, part number, configuration (number of cells in series), nominal voltage and rated capacity shall be provided at the time of submission of proposals
4.4 Calibration Interval: The calibration interval shall be 12 months minimum. At the end of this interval, the test set shall be within all accuracy requirements specified herein with a 85% or greater confidence factor.
4.5 Calibration Procedure: The procedure, software, and special interfaces/adapters that are needed for calibration of the test set shall be provided.
4.6 Digital Interface: A RJ-45 connector to support 10/100 Base-T Ethernet interface for remote access and control shall be provided.
4.7 Dimensions: The test set shall have a dimension of less than 24 cm x 30 cm x 11 cm
(9.5 in x 12 in x 4.3 in).
4.8 Weight: The test set shall have a weight of no greater than 4.6 kg (10 lbs), without batteries or accessories.
4.9 Soft Carrying Case: A soft carrying case according to MIL-PRF- 28800F shall be provided for carrying the test set in the operational environment. The soft carrying case shall be capable of accommodating the test set, accessories, and the operator’s manual or quick start guide.
4.10 Technical Manual: The maintenance philosophy for this unit shall be level 2 (per MIL-
PRF-28800F) and require maintenance to the module level of the unit. The technical manual shall conform to the level 2 maintenance philosophy. This level would be used for most equipment where maintenance and repair is an expected phase of equipment lifecycle. Board level maintenance and troubleshooting information is required. A Use and Installation manual (Operator’s Manual) shall be provided separately.
Maintenance and Servicing manual shall be provided to two levels of maintenance, unit operational verification level and the module level.
Information required for performance verification shall include:
• Instructions to verify equipment performance,
• List the equipment required for verification tests,
• Step-by-step instructions for test connections,
• Acceptable result criteria,
• Calibration information,
• Self-test routines.
Maintenance information shall include:
• Parts lists to the component level,
• Schematics and component layout drawings,
• Block and schematic diagrams.
• List of required test equipment and connection diagrams, and
• Sequential instructions for disassembly, repair, replacement, and reassembly shall be provided.
• Board level maintenance and troubleshooting information,
• Step-by-step instructions for troubleshooting and fault isolation,
• Expected signal levels,
• Test data sheets will be included, and as required,
• The instructions will define localizing a defective circuit card.
Parts lists shall include:
• Parts lists shall be shown on illustrations or a separate listing that includes an index or reference to other illustrations.
• Part number, cage code, and generic description.
The technical manual shall be provided in both printed and electronic formats. The printed format shall be otherwise normally provided. The electronic format shall be in Portable Document Format (PDF) - ISO 32000-1:2008. Two separate CD/DVDs are required, one shall contain the Use and Installation manual and one shall contain the Maintenance and Service manual.
The statement of copyright release:
“REPRODUCTION AND DISTRIBUTION OF THIS TECHNICAL MANUAL IS
AUTHORIZED FOR U.S. GOVERNMENT PURPOSES.”
shall be printed in the first two pages of each technical manual and on the surface of the CD/DVD supplied.
4.11 Training Material: Training material that demonstrates the features, basic operations and procedures with step-by-step instructions for using the test set shall be provided.
The training material shall be provided in technical manual formats.
4.12 Additional Requirements
4.12.1 Human Readable Identification Labeling:
4.12.1.1 Equipment: Per MIL-PRF-28800F, a human readable label shall be provided for all production lot units conforming with MIL-STD-130N and permanently affixed on the equipment in an easily readable location. Required fields on the label are; CAGE, part number, and serial number. Size of the label shall conform to the size of the equipment.
4.12.1.2 Case: Per MIL-PRF-28800F, a human readable metal plate shall be provided for all production lot units conforming with MIL-STD-130N and permanently affixed to the front of the transit case. Required fields on the label are; CAGE, part number, and serial number. Size of the label shall conform to the size of the case. Pressure sensitive adhesive transfer tape is required to hold the plate to the hard transit case such as 3M™ 9472LE. Soft transit cases also require labeling with permanent placement such as a metal plate affixed with rivets, screws or adhesives.
4.12.2 Shipping container: For production lot units the package or carton containing the equipment for shipment shall be marked per MIL-STD-129P.
4.12.3 Other Additional Requirements: Shall be specified in the CDRL of the solicitation.