This document is a Request for Quote (RFQ) for three open-path nitrous oxide analyzers to be used in eddy covariance flux measurements. The project requires high-resolution, field-deployable analyzers capable of simultaneous measurement of nitrous oxide (N2O) and water vapor (H2O) with specific technical requirements such as 0.1-second time resolution, -20°C to 45°C operating temperature range, and 0.2% of ambient level resolution at 10 Hz. The solicitation is a 100% Small Business Set-Aside with NAICS code 334516 (Analytical Laboratory Instrument Manufacturing). Quotes are due by 2:00 PM CDT on August 19, 2024, and the Government anticipates awarding a Firm Fixed Price contract. Evaluation will be based on the lowest-priced, technically-acceptable quote that also has satisfactory past performance.
View the file
Other files for this federal contract opportunity
| File | Type | Posted |
|---|---|---|
| Clauses and Provisions.pdf |
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
SPECIFICATIONS
Project: OPEN-PATH NITROUS OXIDE ANALYZER FOR CONTINUOUS MEASUREMENT
IN EDDY COVARIANCE FLUX MEASUREMENTS
GENERAL INFORMATION
1.0 Scope of Work: Vendor is to provide three open-path nitrous oxide analyzers for continuous measurement of N₂O and H₂O with 0.1 second time resolution, designed for use in eddy covariance flux measurements.
2.0 Background: The accurate measurement of nitrous oxide (N₂O) and water vapor
(H₂O) fluxes is critical for understanding greenhouse gas emissions in agricultural fields. This project requires three high-resolution, open-path instrument capable of simultaneous N₂O and H₂O measurements. The analyzers must be field-deployable, robust, and capable of operating under a range of environmental conditions. This purchase is funded through the Inflation Reduction Act and supports the Midwest Greenhouse Gas Fluxes objective under the ARS-NRCS inter-agency agreement.
CONTRACTOR REQUIREMENTS
3.0 Technical Requirements/Tasks:
The requirement is for analyzers in integrate into an existing sensor system to provide continuous measures of nitrous oxide and water vapor fluxes. The system must include the ability to measure nitrous oxide and water vapor fluxes from crop ecosystems.
Specifications:
The system must meet the following specifications:
Greenhouse Gas Configuration: N₂O and H₂O
Data Measurement Interval: 0.1 second
Resolution: 0.2% of ambient level @ 10 Hz (N₂O 0.7 ppb @ typical ambient level)
Operating Temperature Range: -20 °C to 45 °C
Operating Pressure Range: 78 to 100 kPa
Relative Humidity Range: 0 to 100%
Detection Method: Wavelength Modulation Spectroscopy, 2f detection
Power Requirements: 6 to 8.5 V DC
Power Consumption: 2 watts nominally, 6 – 8 watts with mirror heating
Size: Open path cell dimensions: no more than 10 cm (4 in) dia., 55.9 cm (22 in) height
Attached control box dimensions: No larger than 21.6 cm (8.5 in) height x 11.4 cm (4.5 in) x
15.2 cm (6 in)
Weight: No more than 3.5 kg (7.5 lb)
Optical Path: Maximum 30.5 cm (12 in) physical path and minimum 15 m optical measurement path
Additional Features: Automated clearing of particulates from mirrors
Integration and Connectivity: Compatible with auxiliary items such as sonic anemometers and dataloggers.