Specifications/Configuration
Plexon OmniPlex Neural Data Acquisition System System should be able to receive neural data from 64 or 128 channels coming from inserted electrodes.
System must be able to digitize the entire wide-band signal at 40kHz with a 16-bit A/D converter, with simultaneous sampling.
Unit must be able to be mounted in a 19 inch rack
S- and V-probes multi-channel linear electrode Must contain 24 or 32 platinum/iridium recording sites in a linear array Electrode body must be stainless steel with a 100-120mm (customizable) total probe length a 15um electrode diameter, with customizable inter electrode spacing between 50 and 200um
N-Form arrays Must contain 100 platinum/iridium recording sites on 16 shanks placed in a 4 x 4 array Shanks must be no larger than 125 uM and be spaced 400 uM apart Shanks must be 1.75 mm in length