MINI SPECTROMETERS (qty 25) Federal contract opportunity
Solicitation number
140G0326Q0024
Issued by
Department of the Interior US Geological Survey Office of Acquisitions and Grants

About this file

This document contains technical specifications for spectrometers required for operational scanning UV spectroscopy of SO2 and halogen oxides. The spectrometers must meet the following requirements: small package size with combined length, width, and height less than 30 centimeters; light weight under 1 kilogram; low power consumption below 2.5 watts; spectral resolution of 0.5 nanometers or better; USB 2.0 communication interface; SMA fiber connectivity; adequate pixel mapping with more than 2000 pixels; spectral range exceeding 150 nanometers; UV sensitivity with CMOS detector technology; and adequate stray-light rejection such that pixels below 300 nanometers maintain less than 0.5 percent of maximum intensity.

View the file

On GovTribe

Work with this file on GovTribe

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

The technical specifications of the spectrometers required for use in operational scanning UV spectroscopy of SO2 and halogen oxides are as follows: