US DEPARTMENT OF ENERGY NETL MASS SPECTROMETER SPECIFICATION
Items:
1. One (1) - Thermo Scientific Prima BT Laboratory Bench-Top Gas Analysis Mass Spectrometer
· 50 Micron pin-hole molecular leak
· 75 Micron Micro Capillary Assembly (standard - 10 ml/min)
· 115V 50/60Hz Configuration
· Connection Type – Fractional (1/4 inch) connection fittings
· Rapid Multi-Stream Sampler (RMS) with 16 inlet ports (for up to 15 samples, with one port reserved for calibration gases), heated up to 80 deg C, includes digital sample flow meter and microcapillary
· Automatic Calibration - solenoid valve calibration panel assembly with 6 ports for calibration gases
· Ion chamber treatment to maximize response to OH and S containing components
· Service Case comprising ion source, vacuum gauge, seals kit, fuse kit, tools, spare rotary pump oil (not included where Fomblin Rotary Pump option is used), leak, pressure/flow sensor board, microcapillary and FEP tubing
2. Delivery to:
| U.S Department of Energy |
| National Energy Technology Laboratory |
| 3610 Collins Ferry Road |
| Morgantown, WV 26508 |
| Ion Source |
| Enclosed Electron Impact with Dual Filaments, temperature controlled (settable over range 120-200 degrees C, to ± 0.1 degrees C) |
| Analyzer Type |
| Scanning Laminated Electromagnet, 6 cm radius, 80 degrees deflection |
| Mass Range |
| Adjustable, default is 1-150 amu at 1000 eV ion acceleration voltage, (at 750 eV ion acceleration voltage, mass range is 1-200 amu) |
| Resolution |
| Switchable between two collector resolving slits, resolving powers of 60 (1mm) and 20 (4 mm) are standard. Optionally 140/85 (0.36 mm/0.69 mm) or 100/45 (0.56 mm/1.45 mm) or 140/45 (0.36 mm/1.45 mm) may be fitted |
| Mass scale stability |
| Measured at mass 28 < 0.013 amu over 24 hours |
| Peak Shape |
| At 60 resolution, the ratio of the width of the flat-top (99% height width) to the base peak width (5% height width) 0.4 |
| Abundance sensitivity |
| <250 ppm for 27/28 |
| Detector |
| Faraday and optional Faraday/SEM dual detector |
| Inlet Type |
| Capillary with Molecular leak and bypass (standard configuration) |
| Vacuum System |
| Turbomolecular Pump and Rotary Pump |
| Sample Flow |
| Digitally measured and recorded for each stream for any instrument with RMS option |
| Precision* |
| <0.1% relative (typical, application dependent) |
| Linearity* |
| <1% rel over decade change in concentration (typical, application dependent) |
| Dynamic range* |
| Faraday: 10 ppm – 100% (theoretical, application dependent) |
Optional single MCP SEM: 1 ppm – 1000 ppm (theoretical, application dependent) Optional twin MCP SEM: 10 ppb – 1000 ppm (theoretical, application dependent)
| Stability* |
| <1% relative over 1 week (typical, application dependent) |
Additional Specifications
1. The mass spectrometer shall employ magnetic sector scanning technology for maximum stability and calibration intervals. All peaks over the entire mass range of the mass spectrometer shall be capable of being monitored with a single Faraday detector rather than requiring multiple detectors and a limited magnet scan range.
2. The mass spectrometer shall be capable of automatically selecting from among at least 15 different inlet streams under software control. This rapid multi-stream sampler should be able to heated up to 80 degrees C to avoid any condensation from the process stream. It should also be equipped with a digital flow meter.
3. The unit should include the control for the calibration panel and RMS temperature. The stream selector for the RMS should also be software controlled.
4. The mass spectrometer shall be designed and built to ISO 9001 standards.
5. Gases shall be detectable from 100% to better than 20 ppm using a Faraday detector.
6. The mass spectrometer should be equipped with automatic calibration system - solenoid valve assembly with 6 ports for automatic calibration with up to 6 calibration gases. The automated system is required to allow the mass spectrometer to automatically shut off gas streams from inlets when they are not being sampled.
7. The price of the mass spectrometer shall include the initial installation and testing by the vendor and travel costs associated with the installation. The installation, testing, and travel shall be shown as separate line items in the proposal.
8. The mass spectrometer shall be capable of unattended stand-alone operation in semi climate-controlled environments and must maintain 1% absolute accuracy and stability, without recalibration, during a temperature variation of 20 degrees F over 12 hours, for all peaks with normalized concentrations greater than 1% volume concentrations in the absence of interfering peaks. In climate-controlled environments, the same accuracy and stability shall apply for a period of at least 24 hours.
9. The mass spectrometer shall include dual filaments which are software selectable.
10. The mass spectrometer shall allow at least 32 derived values to be calculated from "raw data" (detector outputs directly proportional to the ion currents) as well as from gas concentrations.
11. The mass spectrometer shall be capable of covering the mass range from 1 to 150. A low-resolution slit, with a mass resolution (defined as the mass at the center of the peak divided by the peak full width in mass units at 10% maximum height) of 25 or less shall be available for peaks in the mass range of 1 to 20. A high-resolution slit shall be available for peaks with masses greater than 20. Switchover from the low-resolution to the high-resolution slit shall be done automatically during runs in which both low-mass and high-mass peaks are monitored.
12. The mass spectrometer shall be able to sample gases (e.g. room air) at truly ambient pressures.
13. Drawings and manuals for the system should be supplied electronically.
14. A serial port (RS232) should be provided for connection to a computer. Additionally, two configurable serial ports should be provided for Modbus communication.
15. Electrical requirements : 120 VAC/60 Hz.
16. The vendor should provide at least 18 months of warranty from the date of delivery and 12 months from the date of instrument commissioning.
17. Service kit (Ion source, vacuum gauge, seals, fuses, micro capillary, tools, etc) should be included
18. The mass spectrometer should have fractional (1/4 inch) connection fittings.
19. The mass spectrometer should be able to analyze gas sample stream from the stream containing the following components with a desirable precision.
| Species | Desired composition range |
| Nominal range (mol%) | |
| Precision (std deviation) mol% |
| analyzed |
| (mol%) |
| H2 |
| 5-60 |
| 17-55 |
| 0.05 |
| co |
| 5-25 |
| 10-20 |
| 0.05 |
| CO2 |
| 0.5-10 |
| 3-6 |
| 0.01 |
| N2 |
| 10-50 |
| 15-45 |
| 0.05 |
| Ar |
| 1-25 |
| 1 or 10-20 |
| 0.02 |
| CH4 |
| 0.1-30 |
| <5 |
| 0.05 |
| H2S |
| <1 |
| <0.5 |
| 0.001 |
| 02 |
| 0-21 |
| <1 |
| 0.01 |
Ethylene, propylene, butene, benzene may also present in the stream but it is not required to analyze these species.