SOW TMDE Calibration.pdf
PDF 332 KB Posted
- Attached to
- TMDE Calibrations-Naval Medical Center San Diego
- Solicitation number
- HT941025Q2051
- Issued by
-
Defense Health Agency
About this file
This Statement of Work (SOW) details service and maintenance requirements for 171 pieces of Test, Measurement, and Diagnostic Equipment (TMDE) at Naval Medical Center San Diego. The contractor must perform comprehensive calibration and preventive maintenance twice annually (May and November) in compliance with NIST ISO/IEC 17025 standards, US Navy regulations, and Joint Commission requirements. Key performance requirements include maintaining a 95% equipment uptime rate, ensuring 95% of periodic maintenance is completed within 30 days of the due date, and using only fully qualified field engineers/technicians with OEM or equivalent training.
The scope covers all labor, parts, materials, and equipment necessary to service government-owned medical testing devices, with specific emphasis on maintaining high-risk and low-risk medical equipment to ensure patient safety. Contractors must submit detailed field service reports within 72 hours of service completion, use NIST-traceable test equipment, and obtain prior approval for any third-party service vendors. The maintenance must preserve the original design and functional capabilities of the equipment, with all replacement parts being new or OEM-refurbished, and all work performed between 0800-1700 Pacific Standard Time.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| Combined Synopsis HT941025Q2051 TMDE Calibrations-BIOMED repair-Final.pdf |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
STATEMENT OF WORK
SERVICE AND MAINTENANCE
1. Scope: The contractor shall provide all services, labor, parts, materials, and equipment necessary for the service and maintenance for the government owned
TMDE Equipment units as listed on Appendix A, "LABEL FOR APPENDIX" at
Naval Medical Center San Diego (NMCSD).
Contractor shall service and maintain the equipment within NIST ISO/IEC 17025, in accordance with the US, State, and Local laws and regulations, US Navy regulations, and Joint Commission requirements.
2. Definitions: As used throughout this work statement, the following terms shall have the meaning set forth below:
2.1. Service and maintenance: (a) any modification, adjustment, or replacement of the equipment that corrects a malfunction and effectively brings the equipment into material conformity with the technical specification for that equipment as designated by the original equipment manufacturer (OEM) and NIST (National
Institute of Standards and Technology), (b) a procedure or routine that, when observed in the regular, proper operation of the equipment, avoids the material adverse effect of the applicable nonconformity, (c) any measures performed that assure the continued operation of the equipment or returns the equipment to its designed efficiency and capacity, (d) replacement of a piece of equipment on a one-for-one basis in order to provide a substitute of equal or greater value in lieu of a damaged or defective device for compensation, return, or refund of equipment.
2.2. Contracting Officer (KO): A warranted representative of the federal government who has the express authority to enter into, administer, and/or terminate contracts and make related determinations and findings. The KO is the only person authorized to approve changes to any terms, conditions, or requirements of the contract.
2.3. Technical Representative (TR): A Government employee appointed by the KO and responsible for the inspection, evaluation, and acceptance of products and services provided in accordance with the terms and conditions of the contract.
3. Location:
Biomedical Repair Department
Building One, Ground Floor
Naval Medical Center San Diego
34800 Bob Wilson Drive
San Diego, CA 92134-5000
4. Requirements:
4.1. General Performance Requirements:
Biomedical Equipment Repairs test measurement and diagnostic equipment’s (TMDE) annual inspection, Calibration. Consist of 171 pieces of equipment that is vital to keeping Naval Medical Center San Diego as they are used to service High risk and Low risk medical equipment and allowing to follow the Joint Commission and ensure the patient safety.
4.1.1. The Contractor shall provide equipment maintenance services to include all periodic planned maintenance and unscheduled repairs in accordance with OEM specifications and in compliance with Beacon Point standards and NIST ISO/IEC
17025, to include repairs necessitated by operator error.
4.1.2. The Contractor may use third party service vendors not limited to the
Original Equipment Manufacturer(OEM), however, these vendors must be approved in advance by the KO and TR. The Contractor must submit a list of any sub-contractors they intend to use as third-party service vendors to the KO and TR for approval. Third party service vendors shall not perform any work under this contract prior to such approval.
4.1.3. Ensure that only fully qualified field engineers and/or technicians who have gone through OEM or equivalent third-party service schools for the equipment shall be employed in the performance of all work performed under this contract.
Service personnel shall have sufficient experience to demonstrate proficiency in maintenance of the product. Upon request, Contractor shall provide written assurance of the competency their personnel as well as a list of credentials for approved field service personnel. The KO and/or TR may authenticate the training requirements and request training certificates or credentials from the service vendors at any time.
4.1.4. Test equipment: Test equipment calibration shall be traceable to National
Institutes of Standards and Technology (NIST traceable). Any efforts conducted without equipment within the calibration intervals is not performance in accordance with this contract. Upon request by TR, the Contractor and/or service vendors shall provide a copy of the current calibration certification of all test equipment to be used, or which has been used, on NMCSD equipment.
4.1.5. Field Service Reports: Contractor shall furnish a written or electronic vendor supplied Field Service Report (FSR) to the Duty Staff at NMCSD Biomed. All
FSR's shall be submitted within 72 hours of completion of service and shall be sent to the following email address:
usn.nmcsd-contractadministrator@health.mil
The contractor or their representative shall complete the FSR to include, at a minimum, the following information:
1. Contractor Name
2. Technician's printed name, telephone number, and signature
3. Date and Time of Arrival
4. ECN (to be provided by Navy) and Serial Number of Equipment
5. Time Expended Repairing/Service: Labor Hours, Rate, Materials
6. Summary of Work Performed and Accepted by End-User, with Government
Representative's Printed name and Signature)
Completed Field Service Reports are required prior to acceptance of any invoice.
Failure to submit a Field Service
Report to NMCSD Biomed can result in delay of payment or rejection of invoice from Government. It is highly recommended that the contractor attach a copy of the Field Service Report to their
Wide Area Workflow Invoice in addition to the required submission to NMCSD Biomed in order to expedite invoicing process.
4.2. NMCSD Preventive Maintenance Requirements:
4.2.1. Preventive maintenance shall be performed in a manner which will minimize corrective maintenance and will include but is not limited to: calibration, adjustment, lubrication, updating software for security, safety, or system functionality issues, testing required to maintain a safe environment of care, electrical safety testing, output testing, replacement of batteries and other shelf life components on a scheduled basis (which becomes corrective maintenance if the component fails due to lack of preventive maintenance).
mailto:usn.nmcsd-contractadministrator@health.mil
4.2.4. Calibrations shall be performed at a minimum of 2 time per year, with these scheduled calibrations (May and NOV) to be coordinated with NMCSD BIOMED after contract award.
4.2.5. Calibration Completion: Contractor must ensure that 95% or more of all periodic maintenance is performed within 30 days of due date, and that 100% of periodic maintenance is performed within 60 days of the due date.
4.2.6. Equipment in Use: If, for reasons beyond the Contractor’s control, equipment is not made available for maintenance for the entire 30 days before the maintenance due date, and 60 days following due date, the Contractor shall immediately notify the COR.
4.2.7. Equipment Unable to Locate (UTL): If the Contractor is unable to locate an item within the maintenance window specified, the Contractor shall immediately notify the TR. The TR may determine that the property has been disposed of or otherwise inactive or unavailable for use and may adjust the equipment covered under this Statement of Work accordingly.
4.2.8. 100% Calibration: Under the circumstances of 4.2.6. and/or 4.2.7., the TR and Contractor shall arrange a means of complying with the requirements for preventive maintenance at no additional cost to the government. Equipment determined missing, lost, or stolen shall be exempt from maintenance.
4.2.9. Uptime Rate: The Contractor shall maintain an overall uptime rate of 95% or better for all equipment covered by this effort. Individual uptime rate for specific equipment, if required, will be as specified on Appendix A.
4.2.10. Uptime Calculation: Uptime is calculated on a rolling 6-month basis. Until the Contractor has at least 6 months of performance history with NMCSD, the uptime shall be calculated cumulatively for the entire duration of the contract.
4.3. NMCSD Corrective Maintenance Requirements:
4.3.1. Not required
4.3.3. Equipment improvements/modifications shall be made only upon BIOMED written approval and direction.
4.3.4. Notify BIOMED immediately upon receipt of OEM safety recall notices or upon receipt of recall notices pertaining to any parts or materials utilized by contractor on the equipment.
4.3.5. Ensure that the original design and functional capabilities shall not be changed, modified, or altered without express written authorization from
BIOMED.
4.3.7. Furnish OEM approved lubricants and lubricate wear points within the equipment in accordance with OEM guidelines.
4.3.8. Replacement parts shall be covered by the contractor for the duration of the contract. Extend to the Government all commercial warranties on replacement parts. All replacement parts or supplies utilized shall be new parts or OEM refurbished parts. OEM refurbished parts shall only be accepted upon written request from the contractor to NMCSD Biomed and shall only be utilized/installed upon written approval from NMCSD Biomed to the contractor.
4.4. End-User Requirements:
4.4.3. Contractor shall perform Calibration on the (171) Biomed TMDE units at least 2 times per year, with a minimum of 6 months between preventive maintenance inspections.
4.4.4. Contractor shall keep department personnel informed as to the status/condition of equipment during PM and the repair status/condition when equipment is down or out for repair. Points of contact for the department are:
Biomedical Repair Contracts team at 619-532-8017 or usn.nmcsd-contractadministrator@health.mil .
4.5. Response Time and Labor:
4.5.2. Contractor shall use commercially reasonable efforts to provide on-site support within two (2) business days from notification by NMCSD Biomed, unless otherwise noted in Section 4.4.
4.5.3. Work required in the performance of this contract shall be performed between the hours of 0800 and 1700, Pacific Standard Time. All compensation for labor from the hours of 0800 to 1700 Pacific Standard Time is provided for in this contract and included in the contract price. Contractor may be permitted to arrange for the performance outside of these hours via coordination with NMCSD Biomed at no additional-cost-to-the-government.
4.5.4. Any work that would require overtime must be approved, in writing, in advance by the KO.
5. Travel and Related Expenses: Contractor shall be responsible for its service-related trip expenses, including round trip travel, mileage, and overnight living expenses.
6. Shipment: The contractor shall not exceed two (2) business days for shipment of all equipment coming from and/or going to NMCSD Biomed minus exceptions or conditions as previously noted in this Statement of Work. All shipping expenses/costs are provided for in this contract and included in the contract price.
7. Liability. N/A
8. Utilities, Facilities, and Equipment Access:
8.1. The contractor may use Government utilities (e.g., electrical power, compressed air, water) that are available and required for any service performed under this contract while at Government facilities. The facilities engineer however, must approve of any contractor equipment which would connect to these utilities to ensure compatibility with NMCSD utilities.
8.2. Any outages required to perform services must be approved by NMCSD
Facilities Management prior to initiation of work and shall be performed per
Facilities' instruction. Contractor may request Facilities assistance in performing these outages if required.
9. Security and Safety Requirements:
9.1. The Contractor is required to submit/complete Naval Base San Diego Base access requirements for access to NMCSD. Contractor shall contact Naval Base
San Diego Security Department at 619-556-1653/5336 or at https://www.cnic.navy.mil/regions/cnrsw/installations/navbase_san_diego/om/forc e_protection/pass_and_decal.html in order to obtain the necessary clearances to perform this contract. All costs for completing the requirements for access are the responsibility of the contractor.
https://www.cnic.navy.mil/regions/cnrsw/installations/navbase_san_diego/om/force_protection/pass_and_decal.html https://www.cnic.navy.mil/regions/cnrsw/installations/navbase_san_diego/om/force_protection/pass_and_decal.html
While waiting for authorized access, contractor shall, at a minimum, contact
NMCSD Biomed not later than 1400 on the day prior to a scheduled visit to submit the required information to complete the NMCSD Temporary Security
Pass. In case of emergent repairs or service, the contractor shall wait at the
NMCSD Main Gate for escort from Biomed personnel.
9.2. Identification, Parking, and Smoking Regulations at NMCSD: Field service personnel shall wear visible photo identification at all times while on the premises of NMCSD. It is the responsibility of the field service personnel to park in the appropriate designated parking areas. Information on parking is available from NMCSD Security personnel. Smoking is prohibited inside any buildings of NMCSD and is restricted to designated smoking locations only. Possession of weapons, alcohol, or pharmaceuticals that are not over the counter and not prescribed to the individual/family is prohibited on the premises.
Enclosed containers, including tool kits, are subject to search.
Violations of regulations may result in citation answerable in the United States
(Federal) District Court. Violation of any policy which results in lost time for
Contractor or his sub-contractors is the sole responsibility of the Contractor, and shall not be charged to the government. Contractor employees shall demonstrate appropriate behavior and dress at all times.
12. Invoicing: All invoicing shall be done in accordance with this statement of work and in accordance with the contract. Failure to do so may delay or interfere with payment for services rendered.
13. NMCSD Points of Contact: For all matters concerning performance of this contract, the primary point of contact shall be:
Biomedical Repair Division, Material Management Department
Building One, Ground Floor, Room GD-18H1
Naval Medical Center San Diego
34800 Bob Wilson Drive
San Diego, CA 92134
TEL: 619-532-8017
Fax: 619-532-8013
Email: usn.nmcsd-contractadministrator@health.mil
Ecn Mfr. Serial No. Nameplate Model Nomenclature
024120 HLTA-40625 HLTA-40
THERMOMETER, ELECTRONIC,
THERMISTOR/THERMOCOUPLE
032356 76661379 77111 MULTIMETER, DIGITAL
032358 66181200 77 MULTIMETER, DIGITAL
034355 918614 MPS450 SIMULATOR, CARDIAC, ELECTROCARDIOGRAPHY
034356 9189026 RF303 TESTER, ELECTROSURGICAL UNIT
034361 9187035 SIGMAPACE 1000 TESTER, EXTERNAL CARDIAC PACEMAKER
034362 9187036 SIGMAPACE 1000 TESTER, EXTERNAL CARDIAC PACEMAKER
034364 12784 IDA 4 PLUS CALIBRATOR, INFUSION DROP-RATE
034483 92040020 87-V MULTIMETER, DIGITAL
034487 9204033 VT PLUS HF CALIBRATOR, RESPIRATORY THERAPY UNIT
035314 208054DFV C.A. 25 PHOTOTACHOMETER
035315 208055DFV C.A. 25 PHOTOTACHOMETER
035535 367300-473 500 PHOTOTACHOMETER
035536 M367300-499 500 PHOTOTACHOMETER
037865 4150713002 PTS2000 TESTER, VENTILATOR
039079 4150735009 PTS2000 TESTER, VENTILATOR
098062 9757030 7000DP TESTER, DEFIBRILLATOR
098532 3027 PS320 SIMULATOR, CARDIAC, HEART SOUND
098533 3001 PS320 SIMULATOR, CARDIAC, HEART SOUND
098591 97250161 TRMS 177 MULTIMETER, DIGITAL
098592 97250162 TRMS 177 MULTIMETER, DIGITAL
098593 97250163 TRMS 177 MULTIMETER, DIGITAL
102389 S10000268 DSTA-40
THERMOMETER, ELECTRONIC,
THERMISTOR/THERMOCOUPLE
102390 S10000405 HLTA-40
THERMOMETER, ELECTRONIC,
THERMISTOR/THERMOCOUPLE
103507 9975052 ESA620
TESTER, ELECTRICAL SAFETY/MEDICAL DEVICE
PERFORMANCE
103508 9975054 ESA620
TESTER, ELECTRICAL SAFETY/MEDICAL DEVICE
PERFORMANCE
103509 9975050 ESA620
TESTER, ELECTRICAL SAFETY/MEDICAL DEVICE
PERFORMANCE
104204 9974035 ESA 612
TESTER, ELECTRICAL SAFETY/MEDICAL DEVICE
PERFORMANCE
104205 9975012 ESA 612
TESTER, ELECTRICAL SAFETY/MEDICAL DEVICE
PERFORMANCE
104206 9975014 ESA 612
TESTER, ELECTRICAL SAFETY/MEDICAL DEVICE
PERFORMANCE
104897 10350148 87-V TRUE RMS MULTIMETER, DIGITAL
104898 10350149 87-V TRUE RMS MULTIMETER, DIGITAL
104900 10350150 87-V TRUE RMS MULTIMETER, DIGITAL
104923 2973 SIMCUBE SC-5 SIMULATOR, CARDIAC, ELECTROCARDIOGRAPHY
104924 2974 SIMCUBE SC-5 SIMULATOR, CARDIAC, ELECTROCARDIOGRAPHY
104998 46440823 77/AN MULTIMETER, DIGITAL
105557 27064 170
TESTER, ELECTRICAL SAFETY/MEDICAL DEVICE
PERFORMANCE
105885 9369004 NIBP ANALYZER TESTER, PNEUMATIC
105888 9344005 NIBP ANALYZER TESTER, PNEUMATIC
106139 10-102 15-150 FLOWMETER, GAS
106142 102 PV350 PRESSURE METER, DIGITAL
106143 101 PV350 PRESSURE METER, DIGITAL
106144 114317 7542 PRESSURE METER, DIGITAL
106170 401358 LIGHT METER METER, LIGHT
107035 1198010 2247700 TESTER, EXTERNAL CARDIAC PACEMAKER
107585 708327 300831 FLOWMETER, GAS
107749 27128 ZSL 600 SCALE, CLINICAL, PRECISION
108368 143183 UW4 POWER METER, ULTRASOUND, DIAGNOSTIC UNIT
108372 3382 MODEL 113A SIMULATOR, CARDIAC, ELECTROCARDIOGRAPHY
108975 25354450 VIC-212 BALANCE, ELECTRONIC, PRECISION
109259 918000060 M2 PRESSURE METER, DIGITAL
109285 Z165145 461893 PHOTOTACHOMETER
109289 8846023 BP PUMP 2 SIMULATOR, NONINVASIVE BLOOD PRESSURE
111406 OX3197 OX-1 SIMULATOR, TRANSDUCER OUTPUT, PULSE OXIMETRY
111408 OX3196 OX-1 SIMULATOR, TRANSDUCER OUTPUT, PULSE OXIMETRY
113244 26354227 AY612 BALANCE, ELECTRONIC, PRECISION
113245 26354342 AY612 BALANCE, ELECTRONIC, PRECISION
113262 SL01622 3600I TESTER, VENTILATOR
113935 11-0102-07 AFG 1000N DYNAMOMETER, MULTIPURPOSE, PUSH-PULL
114606 SL01613 3600I TESTER, VENTILATOR
114607 SL01617 3600I TESTER, VENTILATOR
115159 4663 6625-01-580-9615 SIMULATOR, CARDIAC, ELECTROCARDIOGRAPHY
115502 1808036 ESU303RS-U TESTER, ELECTROSURGICAL UNIT
115936 OX4367 OX-1 SIMULATOR, TRANSDUCER OUTPUT, PULSE OXIMETRY
116007 1848056 7000 TESTER, DEFIBRILLATOR
119670 29209 ZSL600 SCALE, CLINICAL, PRECISION
121250 OX2273 OX-1 SIMULATOR, TRANSDUCER OUTPUT, PULSE OXIMETRY
124438 73781485 ESU-2400 TESTER, ELECTROSURGICAL UNIT
124439 73781484 ESU-2400 TESTER, ELECTROSURGICAL UNIT
129665 6704 SIMCUBE SC-5 SIMULATOR, NONINVASIVE BLOOD PRESSURE
129666 6700 SIMCUBE SC-5 SIMULATOR, NONINVASIVE BLOOD PRESSURE
129667 6701 SIMCUBE SC-5 SIMULATOR, NONINVASIVE BLOOD PRESSURE
129668 6702 SIMCUBE SC-5 SIMULATOR, NONINVASIVE BLOOD PRESSURE
129670 6707 SIMCUBE SC-5 SIMULATOR, NONINVASIVE BLOOD PRESSURE
129677 OX6875 OXSIM OX-1 SIMULATOR, TRANSDUCER OUTPUT, PULSE OXIMETRY
129678 OX6871 OXSIM OX-1 SIMULATOR, TRANSDUCER OUTPUT, PULSE OXIMETRY
129679 OX6872 OXSIM OX-1 SIMULATOR, TRANSDUCER OUTPUT, PULSE OXIMETRY
129682 SL8-3534 SIMSLIM SL-8 SIMULATOR, MULTIPARAMETER
129683 SL8-3532 SIMSLIM SL-8 SIMULATOR, MULTIPARAMETER
129684 OX6874 OXSIM OX-1 SIMULATOR, TRANSDUCER OUTPUT, PULSE OXIMETRY
129685 OX6878 OXSIM OX-1 SIMULATOR, TRANSDUCER OUTPUT, PULSE OXIMETRY
129686 SL8-3529 SIM SLIM SL-8 SIMULATOR, MULTIPARAMETER
129687 SL8-3528 SIMSLIM SL-8 SIMULATOR, MULTIPARAMETER
129688 SL8-3527 SIMSLIM SL-8 SIMULATOR, MULTIPARAMETER
129689 SL8-3531 SIMSLIM SL8 SIMULATOR, MULTIPARAMETER
129690 SL8-3533 SIMSLIM SL-8 SIMULATOR, MULTIPARAMETER
129691 OX6877 OXSIM OX-1 SIMULATOR, TRANSDUCER OUTPUT, PULSE OXIMETRY
132532 H295808 461995 PHOTOTACHOMETER
132533 H302963 461995 PHOTOTACHOMETER
132534 H302920 461995 PHOTOTACHOMETER
132535 H291865 461995 PHOTOTACHOMETER
132536 H302917 461995 PHOTOTACHOMETER
132622 2983030 3979409 SIMULATOR, CARDIAC, ELECTROCARDIOGRAPHY
132623 2984051 3979409 SIMULATOR, CARDIAC, ELECTROCARDIOGRAPHY
132876 SL8-776 SIMSLIM SL-8 SIMULATOR, MULTIPARAMETER
132966 29310057 2550519 MULTIMETER, DIGITAL
132967 29310059 2550519 MULTIMETER, DIGITAL
132968 29310056 2550519 MULTIMETER, DIGITAL
133456 A504506 129003 SIMULATOR, TEMPERATURE
133457 A504509 129003 SIMULATOR, TEMPERATURE
133458 A504529 129003 SIMULATOR, TEMPERATURE
133493 A393580 129003 SIMULATOR, TEMPERATURE
135334 3581051 2247700 TESTER, EXTERNAL CARDIAC PACEMAKER
135335 3582017 2247700 TESTER, EXTERNAL CARDIAC PACEMAKER
135475 3582030 4375516 TESTER, ELECTRICAL SAFETY
135476 3583055 4375516 TESTER, ELECTRICAL SAFETY
135477 3585003 4375516 TESTER, ELECTRICAL SAFETY
135478 3585004 4375516 TESTER, ELECTRICAL SAFETY
135534 33802 2500 METER, PRESSURE
135535 33803 2500 METER, PRESSURE
135536 33804 2500 METER, PRESSURE
135537 33805 2500 METER, PRESSURE
135538 33806 2500 METER, PRESSURE
136441 3643026 2794057 SIMULATOR, MULTIPARAMETER
140508 32018 NEOBLUE
PHOTOTHERAPY UNIT, VISIBLE LIGHT,
HYPERBILIRUBINEMIA
143686 14455 53870 RADIOMETER, PHOTOTHERAPY
143687 14456 53870 RADIOMETER, PHOTOTHERAPY
143696 7699 500075 BILIRUBINOMETER
143697 7701 500075 BILIRUBINOMETER
144843 39800016 4722620 TESTER, INFANT INCUBATOR/WARMER
145696 40811446003 4081G TESTER, VENTILATOR
145699 DM9860328 199XRAY OSCILLOSCOPE
145700 20180412004 04453-000 FLOWMETER, LIQUID, INSPECTION/MAINTENANCE
145701 20180412001 04453-000 FLOWMETER, LIQUID, INSPECTION/MAINTENANCE
145702 20180412002 04453-000 FLOWMETER, LIQUID, INSPECTION/MAINTENANCE
145703 20180412003 04453-000 FLOWMETER, LIQUID, INSPECTION/MAINTENANCE
145946 4149036 2811919 TESTER, DEFIBRILLATOR
145947 4145057 2811919 TESTER, DEFIBRILLATOR
145948 4159054 2811919 TESTER, DEFIBRILLATOR
145949 4158100 3158544 TESTER, DEFIBRILLATOR
145950 4158103 3158544 TESTER, DEFIBRILLATOR
145951 4158107 3158544 TESTER, DEFIBRILLATOR
148115 4338054 4111101 SIMULATOR, TRANSDUCER OUTPUT, PULSE OXIMETRY
148116 4298052 4111101 SIMULATOR, TRANSDUCER OUTPUT, PULSE OXIMETRY
148117 4353063 4111101 SIMULATOR, TRANSDUCER OUTPUT, PULSE OXIMETRY
151323 43730262 2074974 MULTIMETER, DIGITAL
151324 43730261 2074974 MULTIMETER, DIGITAL
151325 43730267 2074974 MULTIMETER, DIGITAL
151326 43730257 2074974 MULTIMETER, DIGITAL
151327 43730264 2074974 MULTIMETER, DIGITAL
151328 43730101 2074974 MULTIMETER, DIGITAL
164514 4246 ST-1 TESTER, ELECTRICAL SAFETY
164515 4245 ST-1 TESTER, ELECTRICAL SAFETY
165989 56548702 5282536 OSCILLOSCOPE
165990 56358703 5282536 OSCILLOSCOPE
165991 58118701 5282536 OSCILLOSCOPE
0M06E5 121751 QED-6H TESTER, DEFIBRILLATOR
0M165D 157515 3000 TESTER, DEFIBRILLATOR
0M2B46 80070270 8000A MULTIMETER, DIGITAL
0M2B4C 80070256 8000A MULTIMETER, DIGITAL
0M3AC0 DM6070871 1420 OSCILLOSCOPE
0M3AD4 49341070 8000A MULTIMETER, DIGITAL
0M3AD6 61201146 8000A MULTIMETER, DIGITAL
0M3C08 1141 CUFFLINK TESTER, PNEUMATIC
0M3C1A 945 4538 TESTER, ELECTROSURGICAL UNIT
0M3C29 99504 RS-232 TESTER, PNEUMATIC
0M3C34 11085 DPM III CALIBRATOR, PRESSURE TRANSDUCER
0M3C3A 944 77 TESTER, ELECTROSURGICAL UNIT
0M3C67 45106 DPM III CALIBRATOR, PRESSURE TRANSDUCER
0M3C76 260-00889 3025 TESTER, FUNCTION GENERATOR
166968 6096743 VT650 CALIBRATOR, RESPIRATORY THERAPY UNIT
166992 5370 FT-2KIT TESTER, INFUSION PUMP
166993 5360 FT-2KIT TESTER, INFUSION PUMP
166981 5052 ST-1
TESTER, ELECTRICAL SAFETY/MEDICAL DEVICE
PERFORMANCE
166979 14187 SC-5 SIMULATOR, NONINVASIVE BLOOD PRESSURE
166980 12151 OX-2 SIMULATOR, CARDIAC
166982 10203 SL-8 SIMULATOR, MULTIPARAMETER
042129 053096067 1726 Tachometer
114617 24048 neoBlue Meter Phototherapy Radiometer
114617 17037 neoBlue Sensor Phototherapy Sensor
162993 733124200 ULT-2020 Leakage Current Tester