This Statement of Work (SOW) details service and maintenance requirements for 171 pieces of Test, Measurement, and Diagnostic Equipment (TMDE) at Naval Medical Center San Diego. The contractor must perform comprehensive calibration and preventive maintenance twice annually (May and November) in compliance with NIST ISO/IEC 17025 standards, US Navy regulations, and Joint Commission requirements. Key performance requirements include maintaining a 95% equipment uptime rate, ensuring 95% of periodic maintenance is completed within 30 days of the due date, and using only fully qualified field engineers/technicians with OEM or equivalent training.

The scope covers all labor, parts, materials, and equipment necessary to service government-owned medical testing devices, with specific emphasis on maintaining high-risk and low-risk medical equipment to ensure patient safety. Contractors must submit detailed field service reports within 72 hours of service completion, use NIST-traceable test equipment, and obtain prior approval for any third-party service vendors. The maintenance must preserve the original design and functional capabilities of the equipment, with all replacement parts being new or OEM-refurbished, and all work performed between 0800-1700 Pacific Standard Time.

View the file

Other files

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

STATEMENT OF WORK

SERVICE AND MAINTENANCE

1. Scope: The contractor shall provide all services, labor, parts, materials, and equipment necessary for the service and maintenance for the government owned

TMDE Equipment units as listed on Appendix A, "LABEL FOR APPENDIX" at

Naval Medical Center San Diego (NMCSD).

Contractor shall service and maintain the equipment within NIST ISO/IEC 17025, in accordance with the US, State, and Local laws and regulations, US Navy regulations, and Joint Commission requirements.

2. Definitions: As used throughout this work statement, the following terms shall have the meaning set forth below:

2.1. Service and maintenance: (a) any modification, adjustment, or replacement of the equipment that corrects a malfunction and effectively brings the equipment into material conformity with the technical specification for that equipment as designated by the original equipment manufacturer (OEM) and NIST (National

Institute of Standards and Technology), (b) a procedure or routine that, when observed in the regular, proper operation of the equipment, avoids the material adverse effect of the applicable nonconformity, (c) any measures performed that assure the continued operation of the equipment or returns the equipment to its designed efficiency and capacity, (d) replacement of a piece of equipment on a one-for-one basis in order to provide a substitute of equal or greater value in lieu of a damaged or defective device for compensation, return, or refund of equipment.

2.2. Contracting Officer (KO): A warranted representative of the federal government who has the express authority to enter into, administer, and/or terminate contracts and make related determinations and findings. The KO is the only person authorized to approve changes to any terms, conditions, or requirements of the contract.

2.3. Technical Representative (TR): A Government employee appointed by the KO and responsible for the inspection, evaluation, and acceptance of products and services provided in accordance with the terms and conditions of the contract.

3. Location:

Biomedical Repair Department

Building One, Ground Floor

Naval Medical Center San Diego

34800 Bob Wilson Drive

San Diego, CA 92134-5000

4. Requirements:

4.1. General Performance Requirements:

Biomedical Equipment Repairs test measurement and diagnostic equipment’s (TMDE) annual inspection, Calibration. Consist of 171 pieces of equipment that is vital to keeping Naval Medical Center San Diego as they are used to service High risk and Low risk medical equipment and allowing to follow the Joint Commission and ensure the patient safety.

4.1.1. The Contractor shall provide equipment maintenance services to include all periodic planned maintenance and unscheduled repairs in accordance with OEM specifications and in compliance with Beacon Point standards and NIST ISO/IEC

17025, to include repairs necessitated by operator error.

4.1.2. The Contractor may use third party service vendors not limited to the

Original Equipment Manufacturer(OEM), however, these vendors must be approved in advance by the KO and TR. The Contractor must submit a list of any sub-contractors they intend to use as third-party service vendors to the KO and TR for approval. Third party service vendors shall not perform any work under this contract prior to such approval.

4.1.3. Ensure that only fully qualified field engineers and/or technicians who have gone through OEM or equivalent third-party service schools for the equipment shall be employed in the performance of all work performed under this contract.

Service personnel shall have sufficient experience to demonstrate proficiency in maintenance of the product. Upon request, Contractor shall provide written assurance of the competency their personnel as well as a list of credentials for approved field service personnel. The KO and/or TR may authenticate the training requirements and request training certificates or credentials from the service vendors at any time.

4.1.4. Test equipment: Test equipment calibration shall be traceable to National

Institutes of Standards and Technology (NIST traceable). Any efforts conducted without equipment within the calibration intervals is not performance in accordance with this contract. Upon request by TR, the Contractor and/or service vendors shall provide a copy of the current calibration certification of all test equipment to be used, or which has been used, on NMCSD equipment.

4.1.5. Field Service Reports: Contractor shall furnish a written or electronic vendor supplied Field Service Report (FSR) to the Duty Staff at NMCSD Biomed. All

FSR's shall be submitted within 72 hours of completion of service and shall be sent to the following email address:

usn.nmcsd-contractadministrator@health.mil

The contractor or their representative shall complete the FSR to include, at a minimum, the following information:

1. Contractor Name

2. Technician's printed name, telephone number, and signature

3. Date and Time of Arrival

4. ECN (to be provided by Navy) and Serial Number of Equipment

5. Time Expended Repairing/Service: Labor Hours, Rate, Materials

6. Summary of Work Performed and Accepted by End-User, with Government

Representative's Printed name and Signature)

Completed Field Service Reports are required prior to acceptance of any invoice.

Failure to submit a Field Service

Report to NMCSD Biomed can result in delay of payment or rejection of invoice from Government. It is highly recommended that the contractor attach a copy of the Field Service Report to their

Wide Area Workflow Invoice in addition to the required submission to NMCSD Biomed in order to expedite invoicing process.

4.2. NMCSD Preventive Maintenance Requirements:

4.2.1. Preventive maintenance shall be performed in a manner which will minimize corrective maintenance and will include but is not limited to: calibration, adjustment, lubrication, updating software for security, safety, or system functionality issues, testing required to maintain a safe environment of care, electrical safety testing, output testing, replacement of batteries and other shelf life components on a scheduled basis (which becomes corrective maintenance if the component fails due to lack of preventive maintenance).

mailto:usn.nmcsd-contractadministrator@health.mil

4.2.4. Calibrations shall be performed at a minimum of 2 time per year, with these scheduled calibrations (May and NOV) to be coordinated with NMCSD BIOMED after contract award.

4.2.5. Calibration Completion: Contractor must ensure that 95% or more of all periodic maintenance is performed within 30 days of due date, and that 100% of periodic maintenance is performed within 60 days of the due date.

4.2.6. Equipment in Use: If, for reasons beyond the Contractor’s control, equipment is not made available for maintenance for the entire 30 days before the maintenance due date, and 60 days following due date, the Contractor shall immediately notify the COR.

4.2.7. Equipment Unable to Locate (UTL): If the Contractor is unable to locate an item within the maintenance window specified, the Contractor shall immediately notify the TR. The TR may determine that the property has been disposed of or otherwise inactive or unavailable for use and may adjust the equipment covered under this Statement of Work accordingly.

4.2.8. 100% Calibration: Under the circumstances of 4.2.6. and/or 4.2.7., the TR and Contractor shall arrange a means of complying with the requirements for preventive maintenance at no additional cost to the government. Equipment determined missing, lost, or stolen shall be exempt from maintenance.

4.2.9. Uptime Rate: The Contractor shall maintain an overall uptime rate of 95% or better for all equipment covered by this effort. Individual uptime rate for specific equipment, if required, will be as specified on Appendix A.

4.2.10. Uptime Calculation: Uptime is calculated on a rolling 6-month basis. Until the Contractor has at least 6 months of performance history with NMCSD, the uptime shall be calculated cumulatively for the entire duration of the contract.

4.3. NMCSD Corrective Maintenance Requirements:

4.3.1. Not required

4.3.3. Equipment improvements/modifications shall be made only upon BIOMED written approval and direction.

4.3.4. Notify BIOMED immediately upon receipt of OEM safety recall notices or upon receipt of recall notices pertaining to any parts or materials utilized by contractor on the equipment.

4.3.5. Ensure that the original design and functional capabilities shall not be changed, modified, or altered without express written authorization from

BIOMED.

4.3.7. Furnish OEM approved lubricants and lubricate wear points within the equipment in accordance with OEM guidelines.

4.3.8. Replacement parts shall be covered by the contractor for the duration of the contract. Extend to the Government all commercial warranties on replacement parts. All replacement parts or supplies utilized shall be new parts or OEM refurbished parts. OEM refurbished parts shall only be accepted upon written request from the contractor to NMCSD Biomed and shall only be utilized/installed upon written approval from NMCSD Biomed to the contractor.

4.4. End-User Requirements:

4.4.3. Contractor shall perform Calibration on the (171) Biomed TMDE units at least 2 times per year, with a minimum of 6 months between preventive maintenance inspections.

4.4.4. Contractor shall keep department personnel informed as to the status/condition of equipment during PM and the repair status/condition when equipment is down or out for repair. Points of contact for the department are:

Biomedical Repair Contracts team at 619-532-8017 or usn.nmcsd-contractadministrator@health.mil .

4.5. Response Time and Labor:

4.5.2. Contractor shall use commercially reasonable efforts to provide on-site support within two (2) business days from notification by NMCSD Biomed, unless otherwise noted in Section 4.4.

4.5.3. Work required in the performance of this contract shall be performed between the hours of 0800 and 1700, Pacific Standard Time. All compensation for labor from the hours of 0800 to 1700 Pacific Standard Time is provided for in this contract and included in the contract price. Contractor may be permitted to arrange for the performance outside of these hours via coordination with NMCSD Biomed at no additional-cost-to-the-government.

4.5.4. Any work that would require overtime must be approved, in writing, in advance by the KO.

5. Travel and Related Expenses: Contractor shall be responsible for its service-related trip expenses, including round trip travel, mileage, and overnight living expenses.

6. Shipment: The contractor shall not exceed two (2) business days for shipment of all equipment coming from and/or going to NMCSD Biomed minus exceptions or conditions as previously noted in this Statement of Work. All shipping expenses/costs are provided for in this contract and included in the contract price.

7. Liability. N/A

8. Utilities, Facilities, and Equipment Access:

8.1. The contractor may use Government utilities (e.g., electrical power, compressed air, water) that are available and required for any service performed under this contract while at Government facilities. The facilities engineer however, must approve of any contractor equipment which would connect to these utilities to ensure compatibility with NMCSD utilities.

8.2. Any outages required to perform services must be approved by NMCSD

Facilities Management prior to initiation of work and shall be performed per

Facilities' instruction. Contractor may request Facilities assistance in performing these outages if required.

9. Security and Safety Requirements:

9.1. The Contractor is required to submit/complete Naval Base San Diego Base access requirements for access to NMCSD. Contractor shall contact Naval Base

San Diego Security Department at 619-556-1653/5336 or at https://www.cnic.navy.mil/regions/cnrsw/installations/navbase_san_diego/om/forc e_protection/pass_and_decal.html in order to obtain the necessary clearances to perform this contract. All costs for completing the requirements for access are the responsibility of the contractor.

https://www.cnic.navy.mil/regions/cnrsw/installations/navbase_san_diego/om/force_protection/pass_and_decal.html https://www.cnic.navy.mil/regions/cnrsw/installations/navbase_san_diego/om/force_protection/pass_and_decal.html

While waiting for authorized access, contractor shall, at a minimum, contact

NMCSD Biomed not later than 1400 on the day prior to a scheduled visit to submit the required information to complete the NMCSD Temporary Security

Pass. In case of emergent repairs or service, the contractor shall wait at the

NMCSD Main Gate for escort from Biomed personnel.

9.2. Identification, Parking, and Smoking Regulations at NMCSD: Field service personnel shall wear visible photo identification at all times while on the premises of NMCSD. It is the responsibility of the field service personnel to park in the appropriate designated parking areas. Information on parking is available from NMCSD Security personnel. Smoking is prohibited inside any buildings of NMCSD and is restricted to designated smoking locations only. Possession of weapons, alcohol, or pharmaceuticals that are not over the counter and not prescribed to the individual/family is prohibited on the premises.

Enclosed containers, including tool kits, are subject to search.

Violations of regulations may result in citation answerable in the United States

(Federal) District Court. Violation of any policy which results in lost time for

Contractor or his sub-contractors is the sole responsibility of the Contractor, and shall not be charged to the government. Contractor employees shall demonstrate appropriate behavior and dress at all times.

12. Invoicing: All invoicing shall be done in accordance with this statement of work and in accordance with the contract. Failure to do so may delay or interfere with payment for services rendered.

13. NMCSD Points of Contact: For all matters concerning performance of this contract, the primary point of contact shall be:

Biomedical Repair Division, Material Management Department

Building One, Ground Floor, Room GD-18H1

Naval Medical Center San Diego

34800 Bob Wilson Drive

San Diego, CA 92134

TEL: 619-532-8017

Fax: 619-532-8013

Email: usn.nmcsd-contractadministrator@health.mil

Ecn Mfr. Serial No. Nameplate Model Nomenclature

024120 HLTA-40625 HLTA-40

THERMOMETER, ELECTRONIC,

THERMISTOR/THERMOCOUPLE

032356 76661379 77111 MULTIMETER, DIGITAL

032358 66181200 77 MULTIMETER, DIGITAL

034355 918614 MPS450 SIMULATOR, CARDIAC, ELECTROCARDIOGRAPHY

034356 9189026 RF303 TESTER, ELECTROSURGICAL UNIT

034361 9187035 SIGMAPACE 1000 TESTER, EXTERNAL CARDIAC PACEMAKER

034362 9187036 SIGMAPACE 1000 TESTER, EXTERNAL CARDIAC PACEMAKER

034364 12784 IDA 4 PLUS CALIBRATOR, INFUSION DROP-RATE

034483 92040020 87-V MULTIMETER, DIGITAL

034487 9204033 VT PLUS HF CALIBRATOR, RESPIRATORY THERAPY UNIT

035314 208054DFV C.A. 25 PHOTOTACHOMETER

035315 208055DFV C.A. 25 PHOTOTACHOMETER

035535 367300-473 500 PHOTOTACHOMETER

035536 M367300-499 500 PHOTOTACHOMETER

037865 4150713002 PTS2000 TESTER, VENTILATOR

039079 4150735009 PTS2000 TESTER, VENTILATOR

098062 9757030 7000DP TESTER, DEFIBRILLATOR

098532 3027 PS320 SIMULATOR, CARDIAC, HEART SOUND

098533 3001 PS320 SIMULATOR, CARDIAC, HEART SOUND

098591 97250161 TRMS 177 MULTIMETER, DIGITAL

098592 97250162 TRMS 177 MULTIMETER, DIGITAL

098593 97250163 TRMS 177 MULTIMETER, DIGITAL

102389 S10000268 DSTA-40

THERMOMETER, ELECTRONIC,

THERMISTOR/THERMOCOUPLE

102390 S10000405 HLTA-40

THERMOMETER, ELECTRONIC,

THERMISTOR/THERMOCOUPLE

103507 9975052 ESA620

TESTER, ELECTRICAL SAFETY/MEDICAL DEVICE

PERFORMANCE

103508 9975054 ESA620

TESTER, ELECTRICAL SAFETY/MEDICAL DEVICE

PERFORMANCE

103509 9975050 ESA620

TESTER, ELECTRICAL SAFETY/MEDICAL DEVICE

PERFORMANCE

104204 9974035 ESA 612

TESTER, ELECTRICAL SAFETY/MEDICAL DEVICE

PERFORMANCE

104205 9975012 ESA 612

TESTER, ELECTRICAL SAFETY/MEDICAL DEVICE

PERFORMANCE

104206 9975014 ESA 612

TESTER, ELECTRICAL SAFETY/MEDICAL DEVICE

PERFORMANCE

104897 10350148 87-V TRUE RMS MULTIMETER, DIGITAL

104898 10350149 87-V TRUE RMS MULTIMETER, DIGITAL

104900 10350150 87-V TRUE RMS MULTIMETER, DIGITAL

104923 2973 SIMCUBE SC-5 SIMULATOR, CARDIAC, ELECTROCARDIOGRAPHY

104924 2974 SIMCUBE SC-5 SIMULATOR, CARDIAC, ELECTROCARDIOGRAPHY

104998 46440823 77/AN MULTIMETER, DIGITAL

105557 27064 170

TESTER, ELECTRICAL SAFETY/MEDICAL DEVICE

PERFORMANCE

105885 9369004 NIBP ANALYZER TESTER, PNEUMATIC

105888 9344005 NIBP ANALYZER TESTER, PNEUMATIC

106139 10-102 15-150 FLOWMETER, GAS

106142 102 PV350 PRESSURE METER, DIGITAL

106143 101 PV350 PRESSURE METER, DIGITAL

106144 114317 7542 PRESSURE METER, DIGITAL

106170 401358 LIGHT METER METER, LIGHT

107035 1198010 2247700 TESTER, EXTERNAL CARDIAC PACEMAKER

107585 708327 300831 FLOWMETER, GAS

107749 27128 ZSL 600 SCALE, CLINICAL, PRECISION

108368 143183 UW4 POWER METER, ULTRASOUND, DIAGNOSTIC UNIT

108372 3382 MODEL 113A SIMULATOR, CARDIAC, ELECTROCARDIOGRAPHY

108975 25354450 VIC-212 BALANCE, ELECTRONIC, PRECISION

109259 918000060 M2 PRESSURE METER, DIGITAL

109285 Z165145 461893 PHOTOTACHOMETER

109289 8846023 BP PUMP 2 SIMULATOR, NONINVASIVE BLOOD PRESSURE

111406 OX3197 OX-1 SIMULATOR, TRANSDUCER OUTPUT, PULSE OXIMETRY

111408 OX3196 OX-1 SIMULATOR, TRANSDUCER OUTPUT, PULSE OXIMETRY

113244 26354227 AY612 BALANCE, ELECTRONIC, PRECISION

113245 26354342 AY612 BALANCE, ELECTRONIC, PRECISION

113262 SL01622 3600I TESTER, VENTILATOR

113935 11-0102-07 AFG 1000N DYNAMOMETER, MULTIPURPOSE, PUSH-PULL

114606 SL01613 3600I TESTER, VENTILATOR

114607 SL01617 3600I TESTER, VENTILATOR

115159 4663 6625-01-580-9615 SIMULATOR, CARDIAC, ELECTROCARDIOGRAPHY

115502 1808036 ESU303RS-U TESTER, ELECTROSURGICAL UNIT

115936 OX4367 OX-1 SIMULATOR, TRANSDUCER OUTPUT, PULSE OXIMETRY

116007 1848056 7000 TESTER, DEFIBRILLATOR

119670 29209 ZSL600 SCALE, CLINICAL, PRECISION

121250 OX2273 OX-1 SIMULATOR, TRANSDUCER OUTPUT, PULSE OXIMETRY

124438 73781485 ESU-2400 TESTER, ELECTROSURGICAL UNIT

124439 73781484 ESU-2400 TESTER, ELECTROSURGICAL UNIT

129665 6704 SIMCUBE SC-5 SIMULATOR, NONINVASIVE BLOOD PRESSURE

129666 6700 SIMCUBE SC-5 SIMULATOR, NONINVASIVE BLOOD PRESSURE

129667 6701 SIMCUBE SC-5 SIMULATOR, NONINVASIVE BLOOD PRESSURE

129668 6702 SIMCUBE SC-5 SIMULATOR, NONINVASIVE BLOOD PRESSURE

129670 6707 SIMCUBE SC-5 SIMULATOR, NONINVASIVE BLOOD PRESSURE

129677 OX6875 OXSIM OX-1 SIMULATOR, TRANSDUCER OUTPUT, PULSE OXIMETRY

129678 OX6871 OXSIM OX-1 SIMULATOR, TRANSDUCER OUTPUT, PULSE OXIMETRY

129679 OX6872 OXSIM OX-1 SIMULATOR, TRANSDUCER OUTPUT, PULSE OXIMETRY

129682 SL8-3534 SIMSLIM SL-8 SIMULATOR, MULTIPARAMETER

129683 SL8-3532 SIMSLIM SL-8 SIMULATOR, MULTIPARAMETER

129684 OX6874 OXSIM OX-1 SIMULATOR, TRANSDUCER OUTPUT, PULSE OXIMETRY

129685 OX6878 OXSIM OX-1 SIMULATOR, TRANSDUCER OUTPUT, PULSE OXIMETRY

129686 SL8-3529 SIM SLIM SL-8 SIMULATOR, MULTIPARAMETER

129687 SL8-3528 SIMSLIM SL-8 SIMULATOR, MULTIPARAMETER

129688 SL8-3527 SIMSLIM SL-8 SIMULATOR, MULTIPARAMETER

129689 SL8-3531 SIMSLIM SL8 SIMULATOR, MULTIPARAMETER

129690 SL8-3533 SIMSLIM SL-8 SIMULATOR, MULTIPARAMETER

129691 OX6877 OXSIM OX-1 SIMULATOR, TRANSDUCER OUTPUT, PULSE OXIMETRY

132532 H295808 461995 PHOTOTACHOMETER

132533 H302963 461995 PHOTOTACHOMETER

132534 H302920 461995 PHOTOTACHOMETER

132535 H291865 461995 PHOTOTACHOMETER

132536 H302917 461995 PHOTOTACHOMETER

132622 2983030 3979409 SIMULATOR, CARDIAC, ELECTROCARDIOGRAPHY

132623 2984051 3979409 SIMULATOR, CARDIAC, ELECTROCARDIOGRAPHY

132876 SL8-776 SIMSLIM SL-8 SIMULATOR, MULTIPARAMETER

132966 29310057 2550519 MULTIMETER, DIGITAL

132967 29310059 2550519 MULTIMETER, DIGITAL

132968 29310056 2550519 MULTIMETER, DIGITAL

133456 A504506 129003 SIMULATOR, TEMPERATURE

133457 A504509 129003 SIMULATOR, TEMPERATURE

133458 A504529 129003 SIMULATOR, TEMPERATURE

133493 A393580 129003 SIMULATOR, TEMPERATURE

135334 3581051 2247700 TESTER, EXTERNAL CARDIAC PACEMAKER

135335 3582017 2247700 TESTER, EXTERNAL CARDIAC PACEMAKER

135475 3582030 4375516 TESTER, ELECTRICAL SAFETY

135476 3583055 4375516 TESTER, ELECTRICAL SAFETY

135477 3585003 4375516 TESTER, ELECTRICAL SAFETY

135478 3585004 4375516 TESTER, ELECTRICAL SAFETY

135534 33802 2500 METER, PRESSURE

135535 33803 2500 METER, PRESSURE

135536 33804 2500 METER, PRESSURE

135537 33805 2500 METER, PRESSURE

135538 33806 2500 METER, PRESSURE

136441 3643026 2794057 SIMULATOR, MULTIPARAMETER

140508 32018 NEOBLUE

PHOTOTHERAPY UNIT, VISIBLE LIGHT,

HYPERBILIRUBINEMIA

143686 14455 53870 RADIOMETER, PHOTOTHERAPY

143687 14456 53870 RADIOMETER, PHOTOTHERAPY

143696 7699 500075 BILIRUBINOMETER

143697 7701 500075 BILIRUBINOMETER

144843 39800016 4722620 TESTER, INFANT INCUBATOR/WARMER

145696 40811446003 4081G TESTER, VENTILATOR

145699 DM9860328 199XRAY OSCILLOSCOPE

145700 20180412004 04453-000 FLOWMETER, LIQUID, INSPECTION/MAINTENANCE

145701 20180412001 04453-000 FLOWMETER, LIQUID, INSPECTION/MAINTENANCE

145702 20180412002 04453-000 FLOWMETER, LIQUID, INSPECTION/MAINTENANCE

145703 20180412003 04453-000 FLOWMETER, LIQUID, INSPECTION/MAINTENANCE

145946 4149036 2811919 TESTER, DEFIBRILLATOR

145947 4145057 2811919 TESTER, DEFIBRILLATOR

145948 4159054 2811919 TESTER, DEFIBRILLATOR

145949 4158100 3158544 TESTER, DEFIBRILLATOR

145950 4158103 3158544 TESTER, DEFIBRILLATOR

145951 4158107 3158544 TESTER, DEFIBRILLATOR

148115 4338054 4111101 SIMULATOR, TRANSDUCER OUTPUT, PULSE OXIMETRY

148116 4298052 4111101 SIMULATOR, TRANSDUCER OUTPUT, PULSE OXIMETRY

148117 4353063 4111101 SIMULATOR, TRANSDUCER OUTPUT, PULSE OXIMETRY

151323 43730262 2074974 MULTIMETER, DIGITAL

151324 43730261 2074974 MULTIMETER, DIGITAL

151325 43730267 2074974 MULTIMETER, DIGITAL

151326 43730257 2074974 MULTIMETER, DIGITAL

151327 43730264 2074974 MULTIMETER, DIGITAL

151328 43730101 2074974 MULTIMETER, DIGITAL

164514 4246 ST-1 TESTER, ELECTRICAL SAFETY

164515 4245 ST-1 TESTER, ELECTRICAL SAFETY

165989 56548702 5282536 OSCILLOSCOPE

165990 56358703 5282536 OSCILLOSCOPE

165991 58118701 5282536 OSCILLOSCOPE

0M06E5 121751 QED-6H TESTER, DEFIBRILLATOR

0M165D 157515 3000 TESTER, DEFIBRILLATOR

0M2B46 80070270 8000A MULTIMETER, DIGITAL

0M2B4C 80070256 8000A MULTIMETER, DIGITAL

0M3AC0 DM6070871 1420 OSCILLOSCOPE

0M3AD4 49341070 8000A MULTIMETER, DIGITAL

0M3AD6 61201146 8000A MULTIMETER, DIGITAL

0M3C08 1141 CUFFLINK TESTER, PNEUMATIC

0M3C1A 945 4538 TESTER, ELECTROSURGICAL UNIT

0M3C29 99504 RS-232 TESTER, PNEUMATIC

0M3C34 11085 DPM III CALIBRATOR, PRESSURE TRANSDUCER

0M3C3A 944 77 TESTER, ELECTROSURGICAL UNIT

0M3C67 45106 DPM III CALIBRATOR, PRESSURE TRANSDUCER

0M3C76 260-00889 3025 TESTER, FUNCTION GENERATOR

166968 6096743 VT650 CALIBRATOR, RESPIRATORY THERAPY UNIT

166992 5370 FT-2KIT TESTER, INFUSION PUMP

166993 5360 FT-2KIT TESTER, INFUSION PUMP

166981 5052 ST-1

TESTER, ELECTRICAL SAFETY/MEDICAL DEVICE

PERFORMANCE

166979 14187 SC-5 SIMULATOR, NONINVASIVE BLOOD PRESSURE

166980 12151 OX-2 SIMULATOR, CARDIAC

166982 10203 SL-8 SIMULATOR, MULTIPARAMETER

042129 053096067 1726 Tachometer

114617 24048 neoBlue Meter Phototherapy Radiometer

114617 17037 neoBlue Sensor Phototherapy Sensor

162993 733124200 ULT-2020 Leakage Current Tester

File details come from the government source that posted it. Updated .