SOW Attachment Testing Chart Requirements.docx

DOCX document 14 KB Posted

Attached to
F103--Sterile Processing Service (SPS) Water testing Federal contract opportunity
Solicitation number
36C24423Q0031
Issued by
Department of Veterans Affairs Veterans Health Administration Veterans Integrated Service Network 4

View the file

Other files for this federal contract opportunity

On GovTribe

Work with this file on GovTribe

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Table 1—Categories and recommended levels of water quality for medical device reprocessing

Type of Water

Utility Water1)
Critical Water

Water Use

Flushing/Washing/Rinsing
Final Rinse2)/Steam

Specifications:

Units

Hardness
mg/L
< 1503)
< 1
Conductivity (mg/L = ppm)
S/cm
< 500
< 10

pH4)

6 – 9
5 – 7
Chlorides
mg/L
< 250
< 1
Bacteria
cfu/mL
n/a
<105)
< 10
Endotoxin
EU/mL
n/a
<205)
< 10

File details come from the government source that posted it. Updated .