Table 1—Categories and recommended levels of water quality for medical device reprocessing
Type of Water
| Utility Water1) |
| Critical Water |
Water Use
| Flushing/Washing/Rinsing |
| Final Rinse2)/Steam |
Specifications:
Units
| Hardness |
| mg/L |
| < 1503) |
| < 1 |
| Conductivity (mg/L = ppm) |
| S/cm |
| < 500 |
| < 10 |
pH4)
| 6 – 9 |
| 5 – 7 |
| Chlorides |
| mg/L |
| < 250 |
| < 1 |
| Bacteria |
| cfu/mL |
| n/a |
| <105) |
| < 10 |
| Endotoxin |
| EU/mL |
| n/a |
| <205) |
| < 10 |