This is a Statement of Work (SOW) for NASA Glenn Research Center's procurement of a Scanivalve scanned pressure system.
NASA Glenn Research Center requires a complete Scanivalve pressure measurement and calibration system to be delivered to their facility in Cleveland, Ohio within 14 weeks from order. The system includes four DSA5002 pressure scanners in different ranges (5.00 PSID, 15.00 PSID, 45.00 PSID, and 100.00 PSID), each with 16-channel capacity, 0.063" bulged tubulations, ±0.04% full scale accuracy, TCP/IP communications, and 24-bit A/D converters with temperature compensation from 0°C to 70°C. Supporting components include three PDM5004 power distribution modules (providing 24VDC output across four channels with 110-220 VAC input), twelve power/serial/trigger cables (156147-01), and twelve ethernet cables (156145-01). The calibration subsystem comprises one SPC4050 2-channel pressure calibrator with IS-33 transducers (0.008% FS accuracy) featuring low and medium pressure servo regulators with multiple transducer ranges, one SPC4050PLU-2 pneumatic logic unit for 2-channel 90 psi operation, one 70380-02 pneumatic distribution panel, and one 73680 portable 4-range calibration system that integrates all calibration components into a durable transport case. The vendor must provide a one-year warranty against defects in materials and workmanship from the original delivery date, with any nonconforming items to be handled as specified by NASA.
View the file
Other files for this federal contract opportunity
| File | Type | Posted |
|---|---|---|
| SATPC0041344 Tab 10 RFQ Open Market Template.pdf | ||
| SATPC0041344 Tab 06 BNJ..pdf |
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
STATEMENT OF WORK
1. Objective/Requirements
NASA Glenn Research Center requires the purchase of a Scanivalve scanned pressure system.
2. Characteristics, Scope, and Specs
LINE PART NO. DESCRIPTION QTY.
1 DSA5002/063T Part Number: DSA5002/063T - 5.00PSID Description: DSA5002 Standard Differential (Single Reference), 16 Channel Pressure Scanner -
5.00PSID, QD
Input Header With 0.063 OD Tubulations
(PN 20009-02)
Fitting Kit
2 DSA5002/063T Part Number: DSA5002/063T - 15.00PSID Description: DSA5002 Standard Differential (Single Reference), 16 Channel Pressure Scanner -
15.00PSID,
QD Input Header With 0.063 OD Tubulations (PN 20009-02) Fitting Kit
3 DSA5002/063T Part Number: DSA5002/063T - 45.00PSID Description: DSA5002 Standard Differential (Single Reference), 16 Channel Pressure Scanner - 45.00PSID, QD Input Header With
0.063 OD Tubulations (PN:20009-02) Fitting
Kit
3 DSA5002/063T Part Number: DSA5002/063T - 100.00PSID Description: DSA5002 Standard Differential (Single Reference), 16 Channel Pressure Scanner - 100.00PSID, QD Input Header With 0.063 OD Tubulations (PN 20009-02) Fitting Kit
PDM5004 PDM5004 Power Distribution Module for up to 4 modules (DC Power Cables sold separately)
156147-01
DSA5000
156147-01 DSA5000 Power / Serial / Trigger Cable, 10 ft long
156145-01
DSA5000
156145-01 DSA5000 Ethernet Cable, 10 ft long
SPC4050 SPC4050 2-Channel Pressure Calibrator (SPECIAL: with IS-33, 0.008� FS Accuracy), with internal barometer Channel A: Low Pressure Servo Regulator
(LPSVR)
Transducer 1: 5psig (Cal’d 5.75psig) Transducer 2: 15psig (Cal’d 17psig) Channel B: Medium Pressure Servo Regulator (MPSVR) Transducer 1: 50psig (Cal’d 56psig) Transducer 2: 100psig (Cal’d 115psig)
SPC4050PLU-2 SPC4050PLU-2 Pneumatic Logic Unit, 2- Channel (90psi
CTL)
70380-02 70380-02 Pneumatic Distribution Panel, SPC4050, 2
Channel, 90 psi Control
73680 73680 4-Range Portable Calibration System with items 8-10 installed. Including Pneumatic / Electrical interface panel, internal pneumatic and electrical connections, testing and documentation. REF: Sales Dwg 80068-01
System shall support the following pressure scanners:
Scanivalve DSA5002/ 063T – 5.00 PSID pressure scanner with the following salient specifications:
• Scanner range: 5 psid
• Fitted with valve for purge and re-zero.
• Fitted with 0.063” bulged tubulations
• Engineering unit data stream up to 5,000Hz (Samples/Channel/Second)
• Accuracy of ±0.04% Full Scale
• Non-multiplexed, synchronous scanning
• TCP/IP communications
• 24-bit A/D converter per channel
• 0 degC to 70 degC temperature compensated range
Scanivalve DSA5002/ 063T – 15.00 PSID pressure scanner with the following salient specifications:
• Scanner range: 15 psid
• Fitted with valve for purge and re-zero.
• Fitted with 0.063” bulged tubulations
• Engineering unit data stream up to 5,000Hz (Samples/Channel/Second)
• Accuracy of ±0.04% Full Scale
• Non-multiplexed, synchronous scanning
• TCP/IP communications
Scanivalve DSA5002/ 063T – 45.00 PSID pressure scanner with the following salient specifications:
• Scanner range: 45 psid
• Fitted with valve for purge and re-zero.
• Fitted with 0.063” bulged tubulations
• Engineering unit data stream up to 5,000Hz (Samples/Channel/Second)
• Accuracy of ±0.04% Full Scale
• Non-multiplexed, synchronous scanning
• TCP/IP communications
• Scanivalve DSA5002/ 063T – 100.00 PSID pressure scanner with the following salient specifications:
• Scanner range: 100 psid
• Fitted with valve for purge and re-zero.
• Fitted with 0.063” bulged tubulations
• Engineering unit data stream up to 5,000Hz (Samples/Channel/Second)
• Accuracy of ±0.04% Full Scale
• Non-multiplexed, synchronous scanning
• TCP/IP communications
System shall support the following power distribution module:
Scanivalve PDM5004 with the following salient specifications:
• Provides DC power for multiple Scanivalve devices
• 4 output channels
• 110-220 VAC 50-60 Hz (90-264VAC)
• 24VDC per output
• 0 to 60°C temperature operation range (95% typ. efficiency)
• Desktop or rack-mountable
System shall support the following cables:
Scanivalve 156147-01 with the following salient specifications:
• For use with Scanivalve DSA5002 pressure scanner and PDM5004 power distribution module
• Provides power, serial, and trigger signal
• 10 ft. length
Scanivalve 156145-01 with the following salient specifications: For use with Scanivalve DSA5002 pressure scanner
• Provides ethernet communications
• 10 ft. length
System shall support the following pressure controller/ calibrator:
Scanivalve SPC4050 with the following salient specifications:
• ChA and ChB transducers - IS-33, 0.008% FS Accuracy
• Internal barometer channel
• Channel A: Low Pressure Servo Regulator (LPSVR) Transducer 1: 5psig (Cal
5.75psig)
• Transducer 2: 15psig (Cal 17psig)
• Channel B: Medium Pressure Servo Regulator (MPSVR) Transducer 1: 50psig
(Cal 56psig)
• Transducer 2: 100psig (Cal 115psig)
• Removeable/ Interchangeable transducers
• Color touch screen
• Used for automated DSA5002 calibrations
• Bench or rack-mount
System shall support the following pneumatic logic unit:
Scanivalve SPC4050PLU-2 Pneumatic Logic Unit with the following salient specifications:
• 2 – channel, 90 psi control pressure
• Positive or below atmospheric (vacuum) calibrations
• Zero offset correction logic
• Leak and response testing on pressure scanners
System shall support the following pneumatic distribution panel:
Scanivalve 70380-02 Pneumatic Distribution Panel with the following salient specifications:
• 2 – channel, 90 psi control pressure
• Provides all necessary supply and control pressure to SPC4050 and
SPC4050PLU-2
System shall support the following portable calibration unit/ case:
Scanivalve 73680 4-range Portable Calibration System with the following salient specifications:
• SPC4050 installed
• SPC4050PLU-2 installed
• 70380-02 installed
• Durable case for the above items to allow for ease of transport
• Protection for sensitive instrumentation in transit
Warranty
• The vendor shall provide a warranty against defects in materials and workmanship for a period of 1 year following completion of work. Any nonconformities identified by NASA as defects in materials and workmanship from the original delivery shall be handled as Non-Conforming Deliverables as described above.
3. Place of Performance NASA Glenn Research Center 21000 Brookpark Rd Cleveland OH 44135-3127
4. Period of Performance
14 weeks ARO.