The required flow cytometer must feature four lasers: Violet (405nm/80mW), Blue (488nm/50mW), Yellow Green (561nm/30mW), and Red (638nm/50mW). Key specifications include fixed integrated optics and quartz flow cell design with >1.3 numerical aperture, flow cell dimensions of 420μm x 180μm internal diameter, and scatter resolution capabilities of <300nm for Blue Side Scatter and 80nm polystyrene particles for Violet Side Scatter. The system must achieve carryover rates of <1.0% in Single Tube Format and <0.5% in Plate Loader Format. Physical requirements specify dimensions of 42.5cm x 42.5cm x 34cm for the cytometer and 14cm x 35.6cm x 35.6cm for tanks/holder, with a cytometer weight of 23.4kg. The unit must operate on 100-240V power drawing 150-250W, within a temperature range of 15-27°C non-condensing.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| SATPC0037770 Tab 07 Capability Statement FBO Posting.pdf |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Statement of Work
Objective/Requirement:
Purchase flow cytometer to detect and measure the physical and chemical characteristics of a population of cells or particles.
Laser Specs:
Laser Wavelength Power
Violet 405 nm 80 mW
Blue 488 nm 50 mW
Yellow Green 561 nm 30 mW
Red 638 nm 50 mW
Other Specs
Fixed integrated optics and quartz flow cell design with >1.3numerical aperture.
Flow Cell dimensions: 420 μm x 180 μm internal diameter
SCATTER RESOLUTION
Blue (488 nm) Side Scatter Resolution: <300 nm
Violet (405 nm) Side Scatter Resolution (VSSC): 80 nm polystyrene particles
CARRYOVER
Single Tube Format: < 1.0%
Plate Loader Format: < 0.5%
DIMENSIONS (W X D X H)
Cytometer
42.5 cm x 42.5 cm x 34 cm
Tanks and Holder
14 cm x 35.6 cm x 35.6 cm
WEIGHT
Cytometer: 23.4 kg
POWER SPECIFICATIONS
Voltage: 100–240 V Power: 150 -250 W
OPERATING TEMPERATURE NON-CONDENSING
15-27 °C