This is a technical specification sheet for a Vacuum Instruments Corporation MS-40 GT Dry Helium Leak Detector being procured by NASA Shared Services Center. The document details precise requirements for a specialized leak detection instrument with capabilities including sensitivity of 4.0 x 10-11 atm cc/sec in Direct Mode, 4.0 x 10-10 atm cc/sec in Reverse Mode, and 5.0 x 10-7 atm cc/sec in Sniff Mode.

The equipment must feature a 61 liter/second turbomolecular high vacuum pump, 16 CFM drag/diaphragm dry internal pump, and operate on 115 VAC, 60 Hz power. Required features include one-button startup, fully automatic operation, internal NIST traceable temperature compensated calibrator, built-in diagnostics, and dual magnetic sector mass spectrometer with resolving power of 14. The vendor must be either the Original Equipment Manufacturer (OEM) or an OEM-authorized vendor, with delivery required within 10-12 weeks after award. This procurement is set aside for small businesses, and vendors must include an OEM Authorized Reseller Letter with their quote.

View the file

Other files

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Specifications for Vacuum Instruments Corporation MS-40 GT Dry Helium Leak Detector

Sensitivity

Response time

Leak Range

Start-up Time

Leak Ranging

Testable masses

Filaments

Max Inlet Pressure

Display

Units

Communication

Required Features

High Vacuum Pump

Internal Pump

Power

Vendor Requirement

Delivery

4.0 x 10-11 atm cc/sec in Direct Mode

4.0 x 10-10 atm cc/sec in Reverse Mode

5.0 x 10-7 atm cc/sec in Sniff Mode

Less than 2 sec Direct Mode, 3 sec Reverse Mode

1.0 x 10-3 to 4.0 x 10-11 (10 to 4x10-11 atm cc/sec w/external pump)

Less than 3 minutes

Auto/Manual

He 3 and He 4

Gross- atmospheric w/external rough pump; 7.5 torr in Fine

Alpha/Numeric atm cc/s, mbar l/sec, Pa M3/sec

Parallel printer port, RS232, Chart recorder, audio, remote

One button start-up, fully automatic operation, tuning & calibration, internal NIST traceable temperature compensated calibrator, built-in diagnostics, dual magnetic sector mass spectrometer (resolving power of 14), auto-zeroing w/continuous background compensation

61 liter/second turbomolecular pump

16 CFM drag/diaphragm dry pump

115 VAC, 60 Hz

Vendor shall be the OEM or OEM authorized vendor

10-12 Weeks after receipt of award lquave Highlight lquave Highlight