SATPC0036049 Tab 04 5 SOW Redacted.pdf
PDF 120 KB Posted
- Attached to
- Spare Netscanner modules for the Netscanner system installed at W7 Test Facility in NASA GRC. Federal contract opportunity
- Solicitation number
- 80NSSC24878586Q-R1
About this file
This document is a Statement of Work (SOW) for the procurement of spare modules for the Netscanner system installed at the W7 test facility in the NASA Glenn Research Center. The SOW details the required components and salient specifications, including Intelligent Pressure Scanners from Pressure Systems Inc. with specific model numbers, compatibility requirements, performance characteristics, and warranty terms. The related federal contract opportunity is a Request for Quote (RFQ) with Solicitation Number 80NSSC24878586Q-R1, seeking these spare Netscanner modules. This is a brand name requirement from TE Connectivity, and vendors must be authorized resellers for TE Connectivity to be eligible. The procuring agency is the National Aeronautics and Space Administration Shared Services Center.
View the file
Other files for this federal contract opportunity
| File | Type | Posted |
|---|---|---|
| SATPC0036049 Tab 06 BNJ_Redacted.pdf | ||
| SATPC0036049 Tab 10 RFQ Open Market (FO).pdf |
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
1/2
Electronically Scanned Pressure System Components Description and Salient Characteristics
July 2024
Listed below are the components required for a spare modules for the Netscanner system installed at W7 test facility in NASA GRC
The required components with salient specifications are listed below:
System shall support the following scanners of varying pressure ranges
Intelligent Pressure Scanner, Pressure Systems Inc. Model 9916 with the following salient specifications:
Qty 1. 9916-0701060000 Qty 1. 9916-0801060000 Qty 1. 9916-0901060000 Compatible for use with a Pressure Systems Inc. Model 98RK scanner interface rack.
Scanner shall contain an on board microprocessor and 16 pressure transducers, each with internal EEPROM memory chip for storing factory calibration data.
Unit shall output pressure data in engineering units, digitally compensated for zero, span, nonlinearity and temperature effects.
Scanner shall be configured for full scale measurement range from a common reference input.
Transducer accuracy, after re-zero, +/-0.05% full scale.
Transducers shall be able to withstand 2X over pressure of full scale range without reducing sensor performance (Proof Pressure).
Able to withstand 3X over pressure of full scale range without damaging sensor (Burst
Pressure).
Scanner shall have an operating range of 0 to 50 degrees C, the transducer reading shall be internally compensated to correct for temperature effects over this operating range of 0 to 50 deg C with a total thermal error of +/-0.001% F.S./deg C.
Scanner shall provide front access for input of the pressure measurements.
Scanner shall have 0.063” bulged tubulation fittings mounted on a quick disconnect plate for input of pressure measurements.
Scanner shall be capable of measuring at a rate of 100HZ per channel.
Scanner shall communicate via 10baseT Ethernet using TCP/IP or UDP/IP protocols.
Scanners shall have internal manifold valving and pneumatic calibration and purge pressure inputs that allow for on-line pneumatic re-zero & span calibrations, on-line purge operation and on-line leak checks of the pressure measurement transducers.
Scanner shall contain necessary solenoids to shift valve positions upon software commands.
2/2
Scanners shall be capable of reading rack source and purge pressures.
Modules must come with a manufacturer’s warranty of 1 year on parts and labor.
Vendor must be an authorized reseller for TE Connectivity.
File details come from the government source that posted it. Updated .