This document details the salient characteristics of the PAMAS SBSS Particle Counter, a precision measurement device with multiple volumetric sensor configurations. The device supports various international test methods including ISO 4406, SAE AS 4059 E, and NAS 1638, and offers pressure ranges from 5-10 bar with three different volumetric sensor models (HCB-LD-50/50, HCB-LD-25/25, and SLS-25/25) capable of measuring particle sizes from 0.5-400 µm. Each sensor has distinct size range capabilities and maximum particle concentration thresholds, with flow rates ranging from 10-25 ml/min.
The counter features advanced technical specifications, including 8-32 adjustable size channels with freely selectable thresholds, a backlit graphical LC-display with 320x240 pixel resolution, an integrated 32-column thermoprinter, and RS-232C data transfer capabilities. It supports multiple power supply voltages (100V-230V, 50-60Hz) and has compact dimensions of 400mm x 300mm x 600mm, making it versatile for various technical and industrial particle analysis applications.
View the file
Other files for this federal contract opportunity
| File | Type | Posted |
|---|---|---|
| SPMYM225Q2620 Benchtop Particle Counter.pdf |
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
PAMAS SBSS Particle Counter Salient Characteristics
1. Applicable Test Methods: ISO 4406, SAE AS 4059 E, NAS 1638, GJB 420, GOST 17216, NAVAIR 01-0A-17
2. Pressure range
a. Standard accessory pump: up to 5 bar
b. High pressure pump or external pressure supply: up to 10 bar
3. PAMAS HCB-LD-50/50 Volumetric Sensor:
a. Size ranges:
• 1-200 µm (ISO 21501-3)
• 1-400 µm: option on request
• 1-100 µm (ISO 4402)
• 4-70 µm(c) according to ISO 11171
b. Max. particle concentration:
• 24,000 P/ml at a flow rate of 25 ml/min
4. PAMAS HCB-LD-25/25 Volumetric Sensor:
a. Size ranges:
• 1-200 µm (ISO 21501-3)
• 1-100 µm (ISO 4402)
• 4-70 µm(c) according to ISO 11171
b. Max. particle concentration:
• 120,000 P/ml at a flow rate of 25 ml/min
5. PAMAS SLS-25/25 Volumetric Sensor:
a. Size ranges:
• 0.5-20 µm (ISO 21501-2)
• 1-20 µm (ISO 4402)
• 1.5-25 µm(c) (ISO 11171)
b. Max. particle concentration:
• 13.000 P/ml at a flow rate of 10 ml/min
6. Dimensions: 400 mm x 300 mm x 600 mm (W x D x H)
7. Counter
a. 8, 16 or 32 free adjustable size channels; the thresholds are free selectable within the measuring range of the particle sensor
b. Backlit graphical LC-display with 320 x 240 pixel resolution.Data printout: 32 column thermo printer
c. Integrated 32 column thermoprinter
d. Data transfer: 8 Bit ASCII code, RS-232C port
e. Power supply via alternating current: 100 V, 115 V, 220 V, 230 V, 50–60Hz