Salient Characteristics - LC Replacement.pdf
PDF 57 KB Posted
- Solicitation number
- FA7000RFI0522
About this file
This document is a Salient Characteristics specification for a Liquid Chromatography system replacement that must be fully compatible with an existing Sciex QTRAP 5500 Mass Spectrometer system.
The primary requirement is a complete Liquid Chromatography (LC) system featuring a system controller, parallel-type double plunger pump with maximum pressure of 15,200 PSI and variable flow rate settings, degassing unit, and autosampler capable of handling injection volumes from 0.1 µL to 50 µL with compatibility for both sample vials and microtiter plates. The column oven must maintain temperatures ranging from 10°C below room temperature to 100°C with ±0.5% accuracy. The system must be fully and natively compatible with the existing Sciex QTRAP 5500 Mass Spectrometer and include all necessary installation components (tubing, cables, solvent bottles, reservoir tray) and covered labor for installation and connection. Additional required components include a solvent selection kit with two dual solvent selection valves integrated into the pumps, a 2 Column Switching Kit with 2 position/6 port valves and valve drive attachment, a high pressure conversion kit, a PFAS Kit with covered installation labor, a Sciex imaged computer with appropriate specifications, a Sciex OS software license to control all LC-MS system features, and full system training delivered by a Sciex Applications Support Scientist.
View the file
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Salient Characteristics
Liquid Chromatography Replacement (Sciex Compatible)
1. Liquid Chromatography System with the following requirements:
a. System controller allowing communication to computer and associated components
b. Parallel-type double plunger pump with maximum pressure of 15,200 PSI and variable flow rate settings
c. Degassing unit
d. Autosampler with injection volume range of 0.1 uL – 50 uL and compatible to be used with sample vials and microtiter plates
e. Column Oven with temperature range of 10C below room temperature up to 100C with accuracy of ±0.5%
f. Liquid Chromatography system is fully and natively compatible with Sciex QTRAP 5500
Mass Spectrometer system
g. All necessary components for installation (tubing, cables, solvent bottles, reservoir tray) are included
h. Covered labor for installation of the system and connection to existing Mass
Spectrometer system
2. Solvent selection kit
a. Two dual solvent selection valves that are directly integrated into the two pumps
3. 2 Column Switching Kit
a. 2 position/6 port valves, 1 valve drive column oven attachment with all tubing/fittings needed for installation
4. High pressure conversion kit
5. PFAS Kit and covered labor for installation
6. Sciex imaged computer with the necessary specifications to run instrument software
7. Sciex OS software license needed to control all features of the LC-MS system
8. Full system training with Sciex Applications Support Scientist
File details come from the government source that posted it. Updated .