About this file
This document is a Salient Characteristics specification for a formalin laboratory recycler required by the VA New Mexico (Albuquerque) Healthcare System. The specification identifies the B/R Instrument Corporation Pureform Formalin Recycler (Product #2100) as the brand name or equal product, with one unit and installation plus on-site training required.
The recycler must meet detailed technical specifications including a minimum 10-gallon boiler and receiver capacity with internal shelving, operation from a single standard electrical outlet, and a mobile, accessible boiler design. The unit requires an air-cooled condenser without glass parts, full accessibility for on-site service, cabinet enclosure, user touch screen controls, built-in post filter, and a vapor sensor with automatic shutdown capability for excess vapor detection. Additional requirements include an overtemperature shutdown device, consistent recycling to 10-12% purity, room temperature readiness for re-buffering, and indefinite recycling capability without formalin degradation over time. This specification supports Solicitation Number 36C26226Q1124, a Request for Quotation issued by the Department of Veterans Affairs Veterans Health Administration Veterans Integrated Service Network 22, set aside for small business concerns on a one-time acquisition basis for the Pathology and Laboratory department.
View the file
Other files for this federal contract opportunity
| File | Type | Posted |
|---|---|---|
| 36C26226Q1124 0001_1.docx | DOCX document | |
| AMENDMENT to RFQ - 36C26226Q1124 0001 - LAB RECYCLER.docx | DOCX document | |
| VENDOR QUESTIONS - LAB RECYCLER.xlsx | XLSX spreadsheet | |
| RFQ 36C26226Q1124 - LAB RECYCLER.docx | DOCX document | |
| 36C26226Q1124.docx | DOCX document |
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
VA New Mexico (Albuquerque) Healthcare System Formalin Laboratory Recycler
1. BRAND NAME OR EQUAL
| BRAND |
| PRODUCT DESCRIPTION |
| PRODUCT # |
| QTY |
B/R Instrument Corporation
Pureform Formalin Recycler, Installation and on-site training
2. SALIENT CHARACTERISTICS
Obtain histology formalin recycler. The recycler must provide the following characteristics:
· The boiler and receiver size must hold a minimum of 10 gallons.
· The unit must have internal shelving for the receiver.
· The unit must require only one standard electrical power outlet.
· The unit must include a mobile and accessible boiler.
· The unit must use an air cooled condenser without glass parts.
· The unit must be fully accessible for on-site service.
· The unit must be enclosed in a cabinet
· The unit must be capable of taking power from standard electrical outlets.
· The unit must have user touch screen control for operation.
· The unit must have a built-in post filter.
· The unit must include vapor sensor to monitor air and alarm with automatic shutdown in case of excess vapor detection.
· The unit must have an overtemperature shutdown device.
· The unit must consistently recycle back to 10-12% purity.
· The unit must be room temperature and ready to re-buffer for use.
· The unit must recycle indefinitely without degradation of the formalin overtime.