The addendum addresses two questions: 1) Progress payments can be included in proposed payment terms, and 2) Clarification on Molybdenum X-ray optics requirements, specifying that Bragg-Brentano/parallel beam configuration is required, with a preference for transmission focusing optic capability for capillary measurements using Mo radiation, particularly for samples containing highly scattering elements like Fe, Co, and Gd. If vendors cannot provide the transmission option, they must propose alternative solutions for collecting data on such samples.

View the file

Other files

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Ames Laboratory RFP A25-0733 ADDENDUM NUMBER 1

ADDENDUM NUMBER 1

ADDENDUM DATE: 12/12/2024

PROJECT: RFP A25-0733

PROPOSAL DUE DATE: 12/20/2024 at 4:00 p.m. CST

ADDENDUM ISSUED DATE: 12/12/2024

The information contained in this Addendum modifies, supplements, or replaces information contained in the RFP Documents dated 12/3/2024 and is hereby made a part of the Contract Documents.

Acknowledge receipt of this Addendum with your proposal submission

Addendum consists of 1 page.

ADDENDUM INDEX

GENERAL ITEMS: Item 1

GENERAL ITEMS

1.) Question #1: Can progress payments be made and suggested in payment terms on the bid response quotation?

Answer #1: Yes, please include your proposed payment terms in your proposal.

Question #2: Please clarify what the expectation is for Molybdenum X-ray optics since none are specified. Is the expectation Bragg-Brentano/parallel beam or transmission focusing optic?

Answer #2: Operation of Molybdenum X-ray optics in Bragg-Brentano/parallel beam configuration is required. The ability to use transmission focusing optic for capillary measurements using Mo radiation is preferred because our polycrystalline research samples often contain highly scattering elements such as Fe, Co, and Gd. If a vendor cannot provide transmission option for Mo radiation, acceptable solutions for collecting data on such samples must be provided.

File details come from the government source that posted it. Updated .