About this file

This document amends a request for proposal for a new mammography unit at the Phoenix Indian Medical Center. The amendment extends the offer due date to August 11, 2023 while keeping the same time. It removes the uninterruptible power supply from the scope of work and only requires a power surge protector or power conditioner to protect equipment from generator testing. Large imaging files will be provided on multiple compact discs or digital video discs rather than a thumb drive since the medical center lacks access to an online file sharing service.

View the file

Other files

File Type Posted
Amendment 9 Trade-In Information.docx DOCX document
Amendment 8 Attachment A - SOW Mammography.docx DOCX document
Amendment-6 Attachment A - SOW Mammography.docx DOCX document
Amendment 6 Questions and Responses.docx DOCX document
Amendment-5 Site Visit.docx DOCX document
Amendment 4 Grey Market.docx DOCX document
Amend-3 Questions Responses.docx DOCX document
Amend-2 Attachment A - SOW Mammography.docx DOCX document
Amend-2 Questions.docx DOCX document
Amend-1 RFP-23-PHX-30 PIMC Mammography.docx DOCX document
Amend-1 Attachment A - SOW Mammography.docx DOCX document
Amend-1 Questions and Responses.docx DOCX document
Attachment A - SOW Mammography.docx DOCX document
Attachment F - Clause Admin Data.docx DOCX document
Attachment E - Price Schedule.xlsx XLSX spreadsheet
Scope of Work Mammography Refresh.docx DOCX document
Attachment C - VistA_Imaging_DICOM.pdf PDF
Attachment B - PWS Service and PM.rtf RTF text file
RFP-23-PHX-30 PIMC Mammography.docx DOCX document
Attachment D - ISA_DEA Template.pdf PDF
Attachment G - 2023 AZ Tax Exemption Letter.pdf PDF
Show all 21

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Amendment 7: Answer Questions and Extend Offer Due Date Amendment 7:

1. Extend Offer Due Date to August 11, 2023 and the time remains the same.

2. There was discussion of removing the UPS from the scope of work, is that change taking place? Yes, we are removing the UPS for the mammography unit. Only included a power surge protector or power conditioner. Goal is to protect the equipment from generator testing.

3. You requested images via. CD in your scope of work. The images are way too large to burn on a CD. They are also too large to burn on a DVD, however it may possible we could do multiple and make it work. Usually for this type of request we use a Thumb Drive (from a trusted source) or a link to a box folder. The box folder would be our first choice, very easy to manage. Yes multiple CD’s or DVDs are fine. No thumb drives. PIMC does not have access to a box folder.

File details come from the government source that posted it. Updated .