This document contains responses to bidder questions submitted for RFP 134-202627-001 regarding a biopharmaceutical automated fill machine isolator being procured by Wilson Community College, operating the Biomanufacturing Education & Skills Training (BEST) Center of Eastern North Carolina. The isolator will serve as a training device for manual aseptic processing and drug product manufacturing techniques, with proposals due November 10, 2026 at 2:00 PM ET and written questions due October 27, 2026. The system must accommodate 2 ml and 25 ml injection vials with automated filling capabilities of up to 40 units per minute at 1% accuracy, mechanical stopper insertion, and integrated containment featuring glove ports and rapid transfer ports. The awardee will be responsible for design, fabrication, assembly, integration, programming, Factory Acceptance Testing, delivery, installation, startup, commissioning, functional and acceptance testing, training, and documentation delivery. The anticipated contract award is December 2026, with performance commencing after contract execution and an anticipated performance period of one year.
The RFP uses a Best Value evaluation methodology ranked by Technical Approach, Vendor References, Vendor Experience, Financial Statements, and Cost in order of importance. Pricing shall be submitted on a milestone basis across ten deliverable phases, with all costs included in the proposal price. Key clarifications addressed in the question responses include confirmation that chilled water requirements for HVAC systems are acceptable if needed for environmental control, though the current design does not include piping chilled water to the fill room; only one sampling point is required for training purposes to conduct both passive sampling and particle monitoring; a peristaltic pump dosing system is preferred for training purposes; and container handling methods remain flexible based on vendor proposal recommendations.
View the file
Other files
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Below find our questions to RFP # 134- 202627-001:
Isolator:
5.2. Specifications B. Containment Isolator and C Utilities
Question: Our integrated HVAC design usually requires chilled water, this is mentioned under utilities “no water” and we need to know the conditions of the surrounding clean room (Temperature etc.) no chilled water could result in higher temperature inside of the Isolator
ANSWER: The intent of adding "No water" to the Utilities section was to eliminate the requirement for piping clean water to the fill machine for cleaning/flushing. The surrounding room will be maintained as a standard laboratory environment that only simulates an ISO 7 room, so it may not be as cool as an actual ISO 7 room. If incoming air must be cooled with chilled water to maintain the environment in the isolator, that is acceptable. However, since the current design does not include piping chilled water to the fill room, the requirement for chilled water could be a negative factor (small) during evaluation if other isolators don't require chilled water.
Environmental monitoring 5.2. Specifications B. Containment
Question: Very common on the market is a combination of active and passive sampling in combination and particle monitoring.. But only passive can also be done by Optima.
We should know the amount of sampling heads for particle counting and settle plates etc… usually at least one sampling point at the de-lidding, Filling, stoppering and capping
Off course this has a bigger impact to the budget.. Since this is a training machine etc.. we could also just make one spot to train settle plate handling.. please let us know how many sampling points for each is needed and if active air sampling is also interesting for training.
ANSWER: Only one sampling point is required for training purposes. That single sampling point can be utilized for both passive sampling (passive is preferred) and particle monitoring.
Filler:
Dosing system.. nothing mentioned of the type.. Optima would assume Peristaltic Pump dosing system because product supply via Bag and weigh cell is mentioned.. Please confirm type of dosing system.
Container unload into separate open Trays would be much easier.. loading back automated into TUB/Nest configuration is also possible but results in bigger effort … this could be done also manual please advise what is preferred.
ANSWER: While we're open to a variety of dosing methods, dosing with a peristaltic pump is preferred for training purposes.