This document outlines requirements for an Infusion System for VISN 12. The requirements specify that the system must include a Large Volume Pump, Syringe Pump, End Tidal CO2 Monitor, and Patient Controlled Analgesia Pump. Detailed technical requirements are provided for each component and the overall system. Key requirements include FDA approval, FIPS 140-2 certification, wireless connectivity, integration with Cerner EHR software, battery backup, alarms and indicators, service and maintenance support, training, and a Dose Error Reduction System. Responses to the related Request for Information are due no later than July 28, 2022 and should include company information, capabilities statement, and availability for follow up calls the week of August 2nd. Information received will be used to determine acquisition approach and potential small business set-aside.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| V12 Infusion System Technical Matrix.xlsx | XLSX spreadsheet |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
VISN 12 Infusion System Minimum Requirements for Acquisition Planning
Specifications for Central Core Component
• System devices must run on an operating system supported by the Operating System
Manufacturer and be in accordance with VA Directive 6550
• Must be FDA approved (Reference: Federal Food Drug & Cosmetic Act (Fd&C Act), Title 21 Cfr Parts 800-1299)
• Must be FIPS 140-2 certified with plan to transition to FIPS 140-3 (Reference: FISMA Federal Information
Security Management Act, Federal Information Processing Standards Publications 140-2 And 140-3 (FIPS Pub), VA Directive 6550)
• Must have wireless communication - (Reference: VA Directive 6550, Industry Standard According to This Publication:
ECRI. Evaluation Background: Large-Volume Infusion Pumps. Device Evaluation 2022 Mar 3.)
• Required to have Cerner capability (Reference: VA DUSHOM Memo 10022017 -- Electronic Health Record Modernization -
Biomedical Devices Clinical Systems and Software Integration); Cerner CareAware Device List Q1 2022 pages 7,8,9; ECRI Institute.
Infusion pump integration: why is it needed, and what are the challenges? [guidance article]; Health Devices 2013 Jul;42(7):210-21; aami-infusion-anthology-10-8-20.pdf)
• Must have automatic switching to battery power upon line-power interruption (Reference: ANSI/AAMI ES60601- 1:2012 “Medical electrical equipment – Part I: General requirements for basic safety and essential performance.”)
• Must have ability to run multiple infusions concurrently with minimal control units
• Must send infusion data wirelessly from equipment to central repository for analysis (Reference:
Joint Commission Policy: NPSG.03.06.01 – MM.08.01.01, PI.01.01.01, EPs 12, 13, Industry standard according to publication: ECRI.
Evaluation background: large-volume infusion pumps. Device Evaluation 2022 Mar 3.)
• Must have primary and secondary infusion modes (Reference: Industry standard according to this publication:
ECRI. Evaluation background: large-volume infusion pumps. Device Evaluation 2022 Mar 3.)
• Must have a minimum warranty of 1-year parts & labor (Reference: Industry standard according to this publication: ECRI. Evaluation background: large-volume infusion pumps. Device Evaluation 2022 Mar 3.)
• Contractor cannot propose a device that does not have anticipated parts support for the next 10 years.
• Must be portable – transportable with battery backup (Reference: 8 hours (standard battery) 4 hours (wireless module) FDA-2006-D-0300 Radio Frequency Wireless Technology in Medical Devices - Guidance for Industry and FDA Staff
• Must be mountable to an IV pole (Reference: Industry standard according to this publication: ECRI. Evaluation background: large-volume infusion pumps. Device Evaluation 2022 Mar 3.)
• Service & Maintenance – o Must be cleanable with hospital grade disinfectant o Must come with manuals (user, service, maintenance) o Must provide manufacturer training and service onsite o Must provide training for clinical end-users, BioMed and pharmacy staff prior to implementation o Must have company staff on site during implementation o Must provide 24/7 phone line for troubleshooting o Must have parts to support equipment repair at VAMC
Modules:
Large Volume Pump (LVP)
• Must be FDA Approved (Reference: Federal Food Drug & Cosmetic Act (Fd&C Act), Title 21 Cfr Parts 800-1299)
• Must have FIPS 140-2 wireless capability
• Must be able to download drug library updates wirelessly
• Must be validated with Cerner’s CareAware smart pump programming, infusion management, and event management workflows (Reference: VA DUSHOM Memo 10022017 -- Electronic Health Record Modernization - Biomedical Devices Clinical Systems and Software Integration); Cerner CareAware Device List Q1 2022 pages 7,8,9; ECRI Institute.
Infusion pump integration: why is it needed, and what are the challenges? [guidance article]; Health Devices 2013 Jul;42(7):210-21; aami-infusion-anthology-10-8-20.pdf)
• Must have automatic switching to battery power upon line-power interruption (Reference: ANSI/AAMI ES60601- 1:2012 “Medical electrical equipment – Part I: General requirements for basic safety and essential performance.”)
• Infusion data sent wirelessly from equipment to central repository for analysis (Reference: Joint Commission Policy: NPSG.03.06.01 – MM.08.01.01, PI.01.01.01, EPs 1, 5, Industry standard according to publication: ECRI. Evaluation background: large-volume infusion pumps. Device Evaluation 2022 Mar 3.)
• Must have bolus capability
• Must have fluid resistance ingress protection
• Device must have flow rate specifications capability to range from 0.1-999 mL/hr. and in incremental rate minimum of 0.001 mL/hr. (Reference: Infusion Nurses Society Third Edition page 442 & 437; ECRI comparison chart: https://www.ecri.org/Components/HPCS/Pages/Infusion-Pumps,-Large-Volume.aspx?tab=2 ; ECRI publication: ECRI Evaluation background: large-volume infusion pumps. Device Evaluation 2022 Mar 3)
• Must default to the Dose Error Reduction System (DERS) upon startup
• Must have free flow prevention mechanism
• Must have the ability to set manual parameters within drug library
• Must have anesthesia mode
• Must have ability to change to different drug libraries
• Must have secondary infusion capability
• Must have automatic switching to battery power upon line-power interruption
• Must have patient tampering safety feature
• Must have loading dose capability
• Must have alarm volume control and alarm momentary silence
• Must be cleanable with hospital grade disinfectant
Syringe Pump
• Must be FDA Approved (Reference: Federal Food Drug & Cosmetic Act (Fd&C Act), Title 21 Cfr Parts 800-1299)
• Must have FIPS 140-2 wireless capability
• Must be able to download drug library updates wirelessly
• Must be validated with Cerner’s CareAware smart pump programming, infusion management, and event management workflows (Reference: VA DUSHOM Memo 10022017 -- Electronic Health Record Modernization - Biomedical Devices Clinical Systems and Software Integration); Cerner CareAware Device List Q1 2022 pages 7,8,9; ECRI Institute.
Infusion pump integration: why is it needed, and what are the challenges? [guidance article]; Health Devices 2013 Jul;42(7):210-21; aami-infusion-anthology-10-8-20.pdf)
• Must have automatic switching to battery power upon line-power interruption (Reference: ANSI/AAMI ES60601- 1:2012 “Medical electrical equipment – Part I: General requirements for basic safety and essential performance.”)
• Infusion data sent wirelessly from equipment to central repository for analysis (Reference: Joint Commission Policy: NPSG.03.06.01 – MM.08.01.01, PI.01.01.01, EPs 12, 13, Industry standard according to publication: ECRI. Evaluation background: large-volume infusion pumps. Device Evaluation 2022 Mar 3.)
• Must have fluid resistance ingress protection
• Device must have flow rate specifications capability to range from 0.1-99mL/hr. and in increments of 0.1 mL/hr. (Reference: Infusion Nurses Society Third Edition page 442 & 437; ECRI comparison chart:
https://www.ecri.org/Components/HPCS/Pages/Infusion-Pumps,-Large-Volume.aspx?tab=2 ; ECRI publication: ECRI Evaluation background: large-volume infusion pumps. Device Evaluation 2022 Mar 3)
• Must default to the Dose Error Reduction System (DERS) upon startup
• Must have free flow prevention mechanism
• Must have the ability to set manual parameters within drug library
• Must have patient tampering safety feature
• Must have loading dose capability
• Must be cleanable with hospital grade disinfectant
• Must have alarm volume control and alarm momentary silence
• Must be able to accommodate various syringe volumes (minimum 10mL & maximum 60mL volumes)
End-tidal Carbon Dioxide (EtCO2) Monitoring (capnography monitory function)
• Must be FDA Approved (Reference: Federal Food Drug & Cosmetic Act (Fd&C Act), Title 21 Cfr Parts 800-1299)
• Must have FIPS 140-2 wireless capability
• Must have automatic switching to battery power upon line-power interruption
• Must have industry standard default alarms
• Must be able to configure alarm settings
• Must have ability to silence, override alarms (anesthesia mode)
• Must have alarm volume control and alarm momentary silence
• Must be cleanable with hospital grade disinfectant
• Must include list of consumables required and starter supplies
Patient Controlled Analgesia (PCA)
• Must be FDA Approved (Reference: Federal Food Drug & Cosmetic Act (Fd&C Act), Title 21 Cfr Parts 800-1299)
• Must have FIPS 140-2 wireless capability
• Must be able to download drug library updates wirelessly
• Must be validated with Cerner’s CareAware smart pump programming, infusion management, and event management workflows (Reference: VA DUSHOM Memo 10022017 -- Electronic Health Record Modernization - Biomedical Devices Clinical Systems and Software Integration); Cerner CareAware Device List Q1 2022 pages 7,8,9; ECRI Institute.
Infusion pump integration: why is it needed, and what are the challenges? [guidance article]; Health Devices 2013 Jul;42(7):210-21; aami-infusion-anthology-10-8-20.pdf)
• Must have automatic switching to battery power upon line-power interruption (Reference: ANSI/AAMI ES60601- 1:2012 “Medical electrical equipment – Part I: General requirements for basic safety and essential performance.”)
• Must have low battery, depleted battery alarms
• Infusion data sent wirelessly from equipment to central repository for analysis (Reference: Joint
Commission Policy: NPSG.03.06.01 – MM.08.01.01, PI.01.01.01, EPs 12, 13, Industry standard according to publication: ECRI. Evaluation background: large-volume infusion pumps. Device Evaluation 2022 Mar 3.)
• Must have fluid resistance ingress protection
• Device must have flow rate specifications capability to range from 0.1-99mL/hr. and in increments of 0.1 mL/hr. (Reference: Infusion Nurses Society Third Edition page 442 & 437; ECRI comparison chart:
https://www.ecri.org/Components/HPCS/Pages/Infusion-Pumps,-Large-Volume.aspx?tab=2 ; ECRI publication: ECRI Evaluation background: large-volume infusion pumps. Device Evaluation 2022 Mar 3)
• Must default to the Dose Error Reduction System (DERS) upon startup
• Must have free flow prevention mechanism
• Must have multiple infusion capability
• Must have patient tampering safety feature
• Must be cleanable with hospital grade disinfectant
• Must have accumulated dose limit monitoring
• Must have patient pendant for PCA dose requests
• Must have bolus capability
• Must have loading dose capability
• Must have alarm volume control and alarm momentary silence
• Must be able to accommodate various syringe volumes
Dose Error Reduction System (DERS)
• Must be FDA Approved (Reference: Federal Food Drug & Cosmetic Act (Fd&C Act), Title 21 Cfr Parts 800-1299)
• Must be configurable with soft limits, hard limits, and/or soft limits nested within hard limits
• Must have capacity to download drug library updates wirelessly
• Must utilize Integrating the Healthcare Enterprise (IHE) Integration Profiles for standardized communication with hospital information systems
• Must be able to customize DERS and drug library o Drug library and health record need to match what is in drug library in pumps
• Must allow medication drug library catalog updates to be wirelessly pushed to devices
• Must have ability to program custom dilutions for critical/dangerous medications
• Must have anesthesia mode to run without EtCO2 monitoring
• Must be able to program DERS limits and infusion devices which actively use DERS for:
o Medication doses ….
o Define minimum # of meds/care areas that need to be programmed (ECRI greater than
10 care areas needed)
• Number of drug entries is greater than 400
• With options for different care therapies (same drug different limits)
• Medication concentration limits
• Infusion rate limits
• Titration limits
• Continuous infusion limits
• Intermittent and secondary infusion limits
• bolus dose and loading dose infusion limits
• Patient weight for weight-based dosing
• DERS Software must be common across all pumps o PCA o Large Volume o Syringe
• Configurable soft limits, hard limits, and/or soft limits nested within hard limits o Configurable alarm volume limits o Configurable infusion pauses duration limits and delayed start duration limits o Configurable clinical alerts to clinicians to be acknowledged during infusion device programming (example “requires central line”)
• Must provide the following data analytics/DERS compliance reports:
o Med specific compliance and alert reports o Total rates of alerts by med
• Reports of meds with low use frequency and high alert rate
• Rates of overrides by med
• Infusion parameter specific alerts reports (dose-rate, concentration, duration alerts)
• Facility compliance o Programming outside of DERS o Event logs - In the event of patient incident or near miss/close call event logs need to be available
• Downloadable and printable after download using
MS office or Adobe PDF
• Events stored o Key presses o Dose limit o Key presses, program settings, alarms, volume infused, dose limit warnings, o All timestamped
• Must have DERS safety features available on infusion devices during network downtime
• Must be validated with Cerner’s CareAware smart pump programming, infusion management, and event management workflows (Reference: VA DUSHOM Memo 10022017 -- Electronic Health Record Modernization - Biomedical Devices Clinical Systems and Software Integration); Cerner CareAware Device List Q1 2022 pages 7,8,9; ECRI Institute. Infusion pump integration: why is it needed, and what are the challenges? [guidance article]; Health Devices 2013 Jul;42(7):210-21; aami-infusion-anthology-10-8-20.pdf) o Must provide the following EHR integration data analytics:
• Integration compliance (rate of usage of smart pump workflows)
• Auto-documentation mismatch
• Failed auto-programming attempts
• Adhere to IHE model for EHR infusion Pump Integration
• Must provide network security documentation:
o Provide a list of Ports, Protocols, and Services required for network communications.
Alarms & Indicators for Overall System
• Fault condition alarms
• Vein and line blocked
• Clamped secondary line alarm
• Air is in the line for LV, PCA
• Infusion completed alert
• Infusion near end
• Empty reservoir
• Flow error
• Set loaded improperly
• Empty syringe
• Plunger disengaged
• Filter replacement
• Pressure Sensor (high pressure/occlusion)
• Occlusion upstream and downstream o Less than 15 psi for downstream o Detection mechanism
• Door/latch open sensor
• System Malfunction
• Low battery/ Depleted battery
• Volume control capability
• Momentary silence less than 2 minutes
Pre-Deployment Requirements
• Must provide all necessary documentation to meet VA Directive 6550 and Enterprise Risk Assessment requirements o MDS2 for each component
• Topology Diagram
• Hardware/Software Inventory
• Provide a list of Ports and Protocols the systems use to communicate on the network
• 6550 Appendix A for each component
• Must follow all required VA Handbook 6500 requirements to provide remote support of server components of the system including but not limited to background investigations, MOU/ISA, PIV badge issuance, check-in requirements.
• Must provide a listing of any required Government Furnished Equipment (GFE) and/or software if required including but not limited to server specifications, workstations specifications, software prerequisites that would be necessary for the infusion system and its associated information support systems.
• Must participate in pre-implementation meetings with VA information security professionals to review system operation to obtain an ERA. A current ERA is required for these systems to operate on the VA network
Training and Support:
• Must certify local Biomedical Engineering staff to perform maintenance and repairs on all equipment.
• Must provide phone technical support during normal business hours 7am-5pm CT, Mon-Fri.
• Must provide a list of approved sources to purchase necessary repair parts to ensure equipment operates to original manufacturer specifications. Parts availability should be generally available next-day shipment.
• Must comply with current FDA regulations pertaining to Recalls and Medical Device correction
• notices (Reference: site FDA regulation -- https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatoryassistance/overview-device-regulation)
• Must provide training for clinical users o Initial training for frontline users will be onsite and conducted by company representatives o Remote training must be available over digital platform, recorded and available for future reference o Must provide access to clinical user education/reference materials via online company site and physical user manual and instructions for use. o Must have 1:1 support available o Must provide notices for software, IFU, user manuals updates o Must provide education on updates via company online site and physical user manual and instructions for use