Requirements Doc.docx
DOCX document 36 KB Posted
- Attached to
- Whole Body Counter Calibration-Maintenance
- Solicitation number
- 75F401-24-Q-1280533
About this file
This document is a Request for Quote (RFQ) for the calibration and maintenance of a Whole Body Counting (WBC) system used by the Department of Health and Human Services (HHS) Food and Drug Administration (FDA) Office of Regulatory Affairs (ORA) at the Winchester, MA facility.
The key requirements include repair of a faulty Cryo-Pulse 5 cooling unit on one Broad Energy Germanium (BEGe) 6530 detector, as well as annual calibration of the WBC system using BOMAB, Livermore Lung, and ANSI neck phantoms with corresponding radioactive sources. The contractor must have experience calibrating similar instrumentation at facilities with a Broadscope Type A Radioactive Materials Nuclear Regulatory Commission License and access to the required calibration sources. The period of performance is one year from the date of award, with delivery and acceptance at the FDA ORA WEAC location in Winchester, MA. Offers are due by August 30, 2024 at 11:59 PM EDT and will be evaluated on the basis of lowest price technically acceptable.
View the file
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Statement of Work Calibrations of Whole Body Counting System
Background WEAC’s Whole Body Counting system provides radiation exposure data to ORA personnel who have inadvertently ingested or inhaled radioactive material. The capability of having a Whole Body Counting system (WBC) is a requirement per ORA's NRC License #20-08361-01.
Objectives
The WBC uses an array of HPGe spectrometers for operation which use Cryo-Pulse 5 mechanical cooling to remain at operating temperature. One detector in the array has a faulty cooling unit requiring repair by an authorized vendor. WEAC is requesting repair of the faulty components of the detector cooling system and restore WEAC’s full capability to perform in-vivo assays using the WBC.
Additionally WEAC’s WBC requires calibration annually. The calibrations require the use of the BOMAB, Livermore Lung, and ANSI neck phantom with corresponding radioactive sources. This will include shipping of the phantoms to/from the calibrator's chosen destinations. Vendor must provide history of calibrating similar instrumentation at facilities with a Broadscope Type A Radioactive Materials Nuclear Regulatory Commission License with access to required calibration sources (BOMAB, Livermore Lung, and ANSI neck phantom). The on-site calibration must occur after repair and delivery of the faulty detector.
Scope
1. The Contractor shall provide repair of the Cryo-Pulse 5 cooling unit on one Broad Energy Germanium (BEGe) 6530 detector. The Cryo-Pulse 5+ cooling unit shall satisfy the performance criteria listed below:
Minimum Technical Requirements
The equipment shall have the following characteristics:
Compatible with existing Cryo-Pulse 5+ (CP-5+) controllers
| Electrical: | AC Power: 110-120 Vac, 50/60 Hz (auto ranging power supply) | |
| <100 W nominal | ||
| <200 W max. (during cool down) | ||
| Connectors: | 2x BNC-F High Voltage Inhibit | |
| RS-232 Connector for remote control and status read-out | ||
| DB15-F-Relay output for cooler status and warm/cold status |
Environmental:
| Forced air cooling |
| +5 to + 40°C operating temperature |
| <55 dB(A) at 1 m noise |
Trade and Service Specifications
1. The instrument repair must include newly manufactured components, not used and refurbished or previously used for demonstration.
1. FOB Point destination to include inside delivery and clean-up of area after installation.
1. The entire system must be warranted for parts and labor for 12 months from the date of formal government acceptance. The vendor must also be capable of servicing the instrument through the covered warranty period. Any equipment repair and maintenance work shall be performed by an OEM-trained engineer. This factory-trained engineer shall have (verified by the OEM) the following: 1) access to OEM factory telephone support; 2) access to the most current OEM factory training for both hardware and software components; and 3) access to all current OEM factory parts, not build-to-order parts. The OEM-trained service engineer shall not use salvaged parts from other instruments for performing maintenance and repairs. All parts used in PM and repairs must be guaranteed, factory-tested, OEM quality parts.
1. Instrument operators shall have access to a technical representative call center at no additional charge, for technical assistance and troubleshooting, which is staffed by senior engineers to provide a high level of expertise for troubleshooting the instrument.
2. The government requires calibration of the Whole Body Counting System, which includes Whole Body Orientation and Lung Orientation:
· The WBC:
· Whole Body
· Contractor shall conduct energy calibration
· Contractor shall conduct efficiency calibration utilizing a BOMAB (Eu-152 source)
· Contractor shall conduct efficiency confirmation counts
· Contractor shall conduct QC counts
· Lung:
· Contractor shall conduct energy calibration
· Contractor shall conduct efficiency calibration utilizing the Livermore Lung Phantom with chest overlays using 5 different chest wall thicknesses.
· Contractor shall conduct efficiency confirmation counts
· Contractor shall conduct QC counts
3. The government requires calibration of the Thyroid System as combined unit and as two separate units:
· Contractor shall conduct energy calibration
· Contractor shall conduct efficiency calibration utilizing an ANSI neck phantom and mixed gamma source
· Contractor shall conduct efficiency confirmation counts
· Contractor shall conduct QC counts
The above calibrations require the use of the BOMAB, Livermore Lung, and ANSI neck phantom with corresponding radioactive sources. This will include shipping of the phantoms to/from the calibrator's chosen destinations.
Vendor must provide history of calibrating similar instrumentation at facilities with a Broadscope Type A Radioactive Materials Nuclear Regulatory Commission License with access to required calibration sources (BOMAB, Livermore Lung, and ANSI neck phantom)
Deliverables
| Deliverable |
| Description |
| Quantity |
| Award/delivery |
| 1 |
| Repair of Cryo Pulse 5+ Unit (BEGe 6530) |
| 1 |
| Immediate upon award. |
| 1 |
| WBC System calibration |
| 1 |
| After repair of detector |
| 1 |
| Thyroid System Calibration |
| 1 |
| After repair of detector |
Security
Check in with Security at front entrance. Deliveries accepted weekdays 8:00am to 4:30pm.
Shipping location
Delivery must be coordinated with ORA/WEAC at least 48 hours in advance:
U. S. FDA Winchester Engineering and Analytical Center Attn: POC to be identified at time of award 109 Holton Street Winchester, MA 01890
Place of Performance
U.S. FDA Winchester Engineering and Analytical Center 109 Holton Street, Winchester, MA 01890
Period of Performance
The Period of Performance begins the date of contract award execution and continues for one year from the date of formal government acceptance, or longer depending on how long the OEM offered warranty period is. Instrument repair shall be immediate. Calibration must occur after repair. Shipment of calibration equipment to site to occur prior to calibration.