Wiped-Film Short Path Molecular Still
Required Minimum Specifications
This requirement is for a laboratory wiped film short path (WFSP) distillation system which shall include:
• feed flask;
• feed pump;
• heating for evaporator, cooling for condenser;
• collection vessel for residue and distillate;
• vacuum system;
• mobile frame;
• platform for the system;
Features, Components, Accessories, Modules, Software and/or Services to be included:
• Jacketed glass WFSP body with internal condenser;
• Variable speed, direct drive for the wiper;
• Wiper speed controller with digital display;
• Feed flask and metering pump to feed sample to the wiped film evaporator;
• Heater for evaporator with temperature controller and digital display;
• Heating/cooling fluid re-circulator for internal condenser with temperature controller and digital display;
• Receivers for residue and distillate;
• Rotary vane and diffusion vacuum pump system with cold trap, vacuum gauge, and vacuum regulator;
• Cooling for vacuum's cold trap;
• Components are mounted on frame and platform with casters and leveling feet.
• WFSP Still: Jacketed borosilicate glass body; evaporator surface area at least 0.03 m2.
• Feed flask, 1L or larger, with valve and adapter for degassing.
• Feed pump: 0-1.0L/h; flow control with digital display; adapter to connect to the WFSP evaporator.
• Wiper assembly: direct drive, variable speed controller up to 500 RPM; digital speed display;
PTFE wiper blades.
• Evaporator heater with temperature control and digital display; Delivers heat up to 300oC.
• Internal condenser heater/cooler circulator with temperature controller and digital display;
temperature range -25 to 180 degrees C.
• Distillate and residue receivers with same volume as the feed flask.
• Vacuum system: rotary vane pump and diffusion pump pressure to 0.001 torr.
• Digital vacuum gauge measuring range 0.001 to 1000 torr.
• Bleed valve to regulate vacuum pressure.
• Cold trap for vacuum.
• Provision for cooling the cold trap, -60 degrees C or better.