Purchase Description (W80RA6-33-25C).docx

DOCX document 16 KB Posted

Attached to
Torque Calibration System
Solicitation number
PANRSA25P0000014826
Issued by

About this file

This document is a Purchase Description and related Request for Quote (RFQ) for a Torque Calibration System for the U.S. Army Test, Measurement and Diagnostic Equipment Activity (USATA). The solicitation (PANRSA25P0000014826) seeks one Torque Calibration System as a life-cycle replacement for an existing Lebow torque calibrator, to be delivered to Building 5435 at Redstone Arsenal. The system must have a maximum measurement capability of 50K lbf-in, utilize data acquisition software, provide automated servo-controlled hydraulic operation, include custom calibration adapters, and have a NIST traceable transfer standard torque cell.

The RFQ is structured as a Firm-Fixed-Price (FFP) procurement with no specific set-aside. Vendors must submit a comprehensive quote including unit pricing, delivery terms, company details, business size, socio-economic status, point of contact information, CAGE code, Unique Entity Identifier, and GSA Schedule number if applicable. The torque calibrator must meet specific technical requirements, including an uncertainty of 0.1% or better at 95% confidence level, and capability to perform calibrations in both clockwise (compression) and counterclockwise (tension) modes. On-site installation and personnel training are also required as part of the procurement.

View the file

Other files

File Type Posted
014826 Questions and Answers.pdf PDF

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Purchase Description Procurement of Torque Calibration System

The U.S. Army Test, Measurement and Diagnostic Equipment Activity (USATA) has a requirement to procure a brand name requirement to identify potential sources that can provide one (1) Torque Calibration System or equivalent for a life-cycle replacement for an older Lebow torque calibrator. The equipment shall be delivered to APSL, located at Building 5435, Fowler Road Redstone Arsenal. The Torque Calibration System module shall include the following specifications or equivalent (if not brand name):

1.0 Requirements:

· Maximum measurement capability of 50K lbf-in

· Utilize data acquisition software to output a calibration report,

· Provide automated servo controlled hydraulic operation,

· Supply custom calibration adapters to complete calibrations on Lebow model 2351-102 36K lb-in torque cells

· Provide NIST traceable 50K lbf-in transfer standard torque cell

· On-site installation/training for personnel.

· The torque calibrator must have an uncertainty of 0.1% or better at 95% confidence level.

· The torque calibrator must also provide calibrations for the unit under test in clockwise (compression) and counterclockwise (tension) modes.

2.0 Delivery Requirements

Shipping Address: U.S. Army TMDE Activity, Attn: Bradford Suggs, Building 5435 Fowler Road, Redstone Arsenal, AL 35898

File details come from the government source that posted it. Updated .