This specification sheet outlines product requirements for a state and local contract opportunity involving two laboratory instruments to be shipped to locations in Anaheim and Sacramento. Line item 1 details a Titrator (to be shipped to Anaheim), which is an Automatic Titration System capable of performing Reserve alkalinity (ASTM D1121) and chloride analysis (ASTM D3634) in a single benchtop unit. The system must include in-line degassing capability, starter consumables, and be compatible with 120V outlets. Technical specifications include potentiometric sensor inputs with specific measurement ranges, temperature sensing capabilities, a touch screen terminal control, and precise burette resolution.
Line item 2 specifies an Automatic Pensky-Martens closed cup flashpoint tester, with one unit to be shipped to Anaheim and another to Sacramento. This instrument is designed to perform ASTM D93 testing and features a cooling fan to reduce analysis time, an electric igniter, boost mode cooling, autodetection via thermoelement, a stainless steel multidetector, and compatibility with 120V outlets. The specification sheet provides detailed technical parameters for both instruments, focusing on their precise measurement capabilities, operational features, and specific testing standards compliance.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| RFQ_P24-1343_SB_Option.pdf |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Specification Sheet:
Line item 1: Titrator (Ship to Anaheim)
Automatic Titration System capable of performing Reserve alkalinity (ASTM D1121) and chloride analysis (ASTM D3634) in one unit. Must include in-line degassing capability.
Benchtop unit. Includes starter amount of instrument consumables needed to perform analysis. 120V outlet plug.
Technical Specifications:
Potentiometric sensor inputs
Measurement Range Plus or minus 2000 mV Resolution/Error Limit 0.1 mV/0.2mV
Temperature-Sensor Measuring Range -20 … 130C Resolution/Error limit 0.1 /0.2C
Terminal Control Touch screen Burette Resolution 10 mL 0.5 uL
Line item 2: Flashpoint (Ship one unit to Anaheim and one unit to Sacramento)
Automatic Pensky-Martens closed cup flashpoint tester with cooling fan to decrease the amount of time between analysis. Capable of performing (ASTM D93). Electric igniter.
Boost mode cooling. Autodetection via thermoelement. Stainless steel multidetector. Plug for 120V outlets.