Performance Work Statement_TriCarb and Quant.pdf
PDF 173 KB Posted
- Solicitation number
- 89243224QNE000024
About this file
This document is a Performance Work Statement (PWS) for the support and maintenance of two PerkinElmer Liquid Scintillation Counters owned by the Department of Energy (DOE) Radiological and Environmental Sciences Laboratory (RESL). The PWS includes requirements for annual preventative maintenance, on-site corrective maintenance and repair services, and the procurement and installation of replacement parts as needed to keep the instruments operational. The work will be performed at the contractor's location as well as on-site at the RESL facility in Idaho Falls, ID. The PWS also includes a Quality Assurance Surveillance Plan that outlines the performance standards, acceptance criteria, and monitoring methods for the required services.
The related federal contract opportunity is a notice of intent to award a sole source service maintenance agreement for the RESL's PerkinElmer instrumentation, specifically the Quantulus GCT 110 V and Tri-Carb 5110TR 110 V models. The DOE intends to award the contract to Oxford Instruments America, Inc. as the Original Equipment Manufacturer (OEM) and only known source capable of providing the required comprehensive maintenance and support. The notice allows for potential alternate sources to respond by September 27, 2024 if they can meet the RESL's requirements. The contract was awarded on September 30, 2024 to an unknown vendor for an unknown amount.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| Sole Source Justification (JOFOC) - RESL_Redacted.pdf |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
24NE000340
PERFORMANCE WORK STATEMENT
Department of Energy (DOE)
Radiological and Environmental Sciences Laboratory (RESL)
A. Introduction:
This Performance Work Statement (PWS) is for the support of existing PerkinElmer Liquid Scintillation Counters. It includes preventative maintenance and corrective maintenance services.
The United States Department of Energy (DOE) Radiological and Environmental Sciences Laboratory (RESL) currently owns the following PerkinElmer Liquid Scintillation Counters:
Model: Tri-Carb 5110TR SN# SGLO 3819 1147 Manufacturer: PerkinElmer
Model: Quantulus GCT 6220 SN# SGLO 3117 0554 Manufacturer: PerkinElmer
B. Background:
RESL conducts cost-effective measurement quality assurance programs that help assure that key DOE missions are completed, assuring quality and stability of laboratory measurement systems throughout DOE, and by providing expert technical assistance.
The PerkinElmer equipment is used to support the DOE Laboratory Accreditation Program required by 10 CFR 835 and its Mixed Analyte Performance Evaluation Program.
This PWS includes purchasing service maintenance support for the above-listed equipment, emergency on-site visits, and parts as needed to support RESL’s current equipment and RESL’s future needs using PerkinElmer instrumentation.
C. Scope of Work:
The following describes the work effort for services, as ordered under this PWS.
The contractor shall provide all maintenance and repair services on-site as well as procure and install replacement parts and supplies when necessary to maintain the proper operation of the listed instruments and equipment itemized in section A.
Task 1 – Preventative Maintenance The contractor shall perform preventative maintenance (PM) for all liquid scintillation counters listed in section A on an annual basis. PM shall be performed in accordance with the original manufacturers’ guidelines. Preventative maintenance services include but are not limited to cleaning and operation of the computer boards, cleaning and operation of the photomultiplier tubes, computer system backups, system calibration, sensitivity checks, and associated module checks for proper functionality.
Task 2 – Corrective Maintenance/Service Repairs
Prior to performing any Corrective Maintenance requirements beyond the covered Preventive Maintenance tasks, such as work or repair parts, these will need to be coordinated through the RESL Technical Monitor and then authorized by the issuance of a Task Order by the DOE Contracting Officer.
The contractor shall provide repair services for all instruments and equipment listed in section A. Service repair shall include technical telephone assistance within one workday of notification of a problem. On-site service repair shall begin within three workdays of notification of a problem unless specified otherwise by the Technical Monitor or designee.
The contractor shall procure and install all replacement parts and circuit boards of at least equivalent quality for defective parts and circuit boards found while performing repair services. Replacement parts and circuit boards shall be either shipped to or physically carried by the contractor to the RESL Laboratory, to the attention of the Technical Monitor, within one workday after notification of need for such by the Technical Monitor. The contractor shall be responsible for installation and proper functioning of all replacement parts and circuit boards and any other repair service performed.
D. Quality Surveillance Plan:
The Quality Surveillance Plan shown below will be utilized by RESL in monitoring contractor performance. It clearly delineates the performance standard applicable to particular tasks, the maximum allowable deviation which constitutes acceptance criteria, and the method of surveillance that will be utilized for that activity.
QUALITY ASSURANCE SURVEILLANCE PLAN (QASP)
REQUIRED
SERVICES
PERFORMANCE
STANDARD
MAXIMUM
ALLOWABLE
DEVIATION
METHOD OF
SURVEILLANCE
AMOUNT
INCENTIVE/
DISINCENTIVE
Preventative Preventative None Technical Monitor review of May adversely
Maintenance maintenance is instrument logbooks impact performed in accordance and contractor CPARS with the manufacturers’ service reports. With performance rating guidelines every 12 documentation in months (from effective writing date of contract) unless specified otherwise by the PO or designee.
Corrective Maintenance/Ser vice Repairs
Technical telephone None Technical Monitor review of RESL
May adversely assistance provided on-site observation impact within 1 workday of log. With CPARS notification of a documentation in performance rating problem. On-site repair writing services began within three workdays of notification of a problem unless specified otherwise by the PO or designee.
E. Security Considerations:
Foreign or foreign nationals are not permitted into DOE buildings onto any DOE sites without prior authorization. Technicians will be escorted by a technical representative from RESL for any on-site visits required. Contractor personnel accessing Government facilities will need to wear a DOE-issued access badge; they may also be required to have background checks. The contractor shall comply with all federal and site-specific security safeguards; The DOE Technical Monitor will inform contractor personnel of any specific security safeguards prior to their accessing the Government facility, i.e. badging requirements.
F. Environmental, Safety and Health Considerations:
Technician will be required to read RESL safety pamphlet. The contractor shall comply with all federal and site-specific safety and security safeguards.
H. Travel/Work/Shipping Location:
Work shall be at the contractor’s location as well as for services at RESL, 2251 N.
Boulevard, Idaho Falls, ID 83402
Shipping Address:
Cannon Johnson 1765 N. Yellowstone Highway Bldg RESL601 Idaho Falls, Id 83401
I. Point of Contact/Technical Monitor:
Technical Monitor, (RESL) Cannon Johnson Ph: (208) 690-8150 Email: johnsocm@id.doe.gov
Alternative RESL Contact Michelle Smith Ph: (208) 526-7084 Email: smithml@id.doe.gov mailto:johnsocm@id.doe.gov mailto:smithml@id.doe.gov
PERFORMANCE WORK STATEMENT