This is a product specification document for the PDS-100G-GN/ID Spectrometric Personal Radiation Detector manufactured by Mirion Technologies, a device designed for the State of Maine. The PDS-100G-GN/ID is a pocket-sized, dual-capability radiation detection system available in two versions: the PDS-100G/ID for gamma detection and the PDS-100GN/ID for combined gamma and neutron detection. The device features embedded spectrum acquisition and identification capabilities to detect, locate, quantify, and identify radioactive sources while discriminating between naturally occurring radioactive materials, medical isotopes, industrial sources, and special nuclear materials. The detector is intended for use by first responders, law enforcement, customs inspectors, and security personnel at critical infrastructure facilities and incorporates automated and manual identification modes with visual, audio, and vibration alarms, wireless communication interfaces via IRDA and Bluetooth technology, and data storage for later transmission of spectra and identification results.
The PDS-100G-GN/ID utilizes a cesium iodide detector for gamma detection with sensitivity exceeding 400 counts per second per millisievert per hour for Cesium-137, and includes a lithium iodide detector for neutron detection in the GN version. The device measures 123 x 74 x 43 millimeters, weighs 300 grams, and operates on two AA batteries with a typical operational lifetime of 100 hours. It functions across a temperature range of -20°C to 50°C and is resistant to electromagnetic interference, shock, vibration, drop, and water exposure. The detector can identify isotopes across four categories: naturally occurring radioactive materials, medical isotopes, industrial sources, and special nuclear materials including uranium-235, uranium-238, and plutonium-239. Accessories available include PDSmass and SMI software for remote display, spectra analysis, and identification replay, along with protective cases, belt clips, and external power supplies.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| Opportunities Municipalities and Schools.doc | DOC document | |
| Specifications.doc | DOC document | |
| CONTRACT TERMS.rtf | RTF text file | |
| Solicitation_terms.doc | DOC document |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
health physics A Mirion Technologies Division
Featuring:
PDS-100G-GN/ID
Detection and Identifi cation - Risks Prevention
PDS-100G-GN/ID
Spectrometric Personal Radiation Detector
The PDS-100G/ID and PDS-100GN/ID are the most ultimate evolution of the gamma and gamma/neutron radiation detectors offering embedded spectrum acquisition and identifi cation.
These sensitive pocket-sized devices are designed to detect, locate, quantify and identify radioactive sources to discriminate on the spot Naturally
Occuring Radioactive Material (NORM) and medical isotopes against industrial sources or Special Nuclear
Materials.
High sensitivity provides better spectra in a shorter time. Fully automated acquisition and identifi cation as well as manual modes are provided.
Identfi cation results are displayed as a list of isotopes with categories and confi dence levels. Spectra and ID results are memorized for later transmission.
PDS-100G/ID and PDS-100GN/ID have been designed specially for First Responders, Law
Enforcement, Customs inspectors and for Personnel and Site security in critical infrastructures.
VERSIONS:
PDS-100G/ID: Gamma version•
PDS-100GN/ID: Gamma and Neutron version•
FEATURES
Detection, Search and Identifi cation of nuclear material for
Homeland Security applications.
Small, rugged, compact, user-friendly ●
High sensitivity and fast response time ●
Embedded identifi cation, automated & manual mode ●
Visual, audio and vibration alarms ●
Wireless communication interface ●
Mirion Technologies (MGPI) Inc
5000 Highlands Parkway Suite 150 Smyrna Georgia 30082
USA
T +1.770.432.2744 F +1.770.432.9179
Mirion Technologies (MGPI) SA
BP 1
F-13113 Lamanon France
T +33 (0) 4 90 59 59 59 F +33 (0) 4 90 59 55 18
Mirion Technologies (RADOS) Oy
P.O. Box 506 FIN-20101 Turku Finland
T +358 2 468 4600 F +358 2 468 4601
Mirion Technolgies (RADOS) GmbH
Ruhrstrasse 49 DE-22761 Hamburg Germany
T +49 (0) 40 851 93-0 F +49 (0)40 851 93 256
Health Physics Division
Since norms, specifi cations and designs are subject to occasional change, please ask for confi rmation of the information given in this publication.
PDS-100G-GN/ID
Detection and Identifi cation - Risks Prevention www.mirion.com
144215EN-B
DETECTION
Detector gamma CsI(Tl): ● > 400 cps per μSv/h for 137Cs Detector neutron LiI(Eu) (GN version only) ●
Gamma dose rate display 0.01 μSv/h ● to 100 μSv/h (1 μrem/h to 10 mrem/h) Gamma count rate display: 0 to 99 999 cps ●
Neutron count rate display: 0.0 to 999 cps ●
Gamma alarm response time: ●
Standard threshold: 0.5 μSv/h step, alarm within 1 s ●
Sensitive threshold: 0.05 μSv/h step, alarm within 3 s ●
Neutron alert response time: mean time to detect ●
20000 n/s Cf252 at 10 cm: <2 sec.
SEARCH
1s integration time with chirp ●
SPECTROMETRY AND IDENTIFICATION
512 / 1024 channels spectra : 30keV to 1.7 MeV ●
Automated alarm triggered mode with auto ● confi rmation Manual mode with preset time and/or counts and ● resume capability Identifi cation by NMD algorithm ● up to 4 isotopes mixed ○ detectabiliby grade, unknown or ID unsure indication ○ identifi cation time at 1 μSv/h typical 1 minute ○ designed to exceed ANSI N42-48 SPRD standard ○
FUNCTIONAL FEATURES
Detection, search and/or identifi cation mode ●
Source indication alarm and danger alarm ●
Visual, audible and silent alarms (vibration, earphone) ●
Easy-to-read display (OLED technology) ●
Memory of 100/50 512 ch/1024 ch. spectra ● and >1000 events IRDA and Bluetooth® technology communication ●
ELECTRICAL & MECHANICAL CHARACTERISTICS
Power supply: 2 x AA batteries ●
(Lithium, Alkaline or Ni-MH); lifetime typical 100 hours Dimensions (l x w x h): 123 x 74 x 43 mm ●
(4.84 x 2.91 x 1.69 in); weight: 300 g (10.58 oz)
ENVIRONMENTAL CHARACTERISTICS
Temperature range:- 20°C to 50°C (-4°F to 122°F) ●
ID: -15°C to +45°C (5°F to 115°F) EMI, shock, vibration, drop and water resistant ●
ACCESSORIES
PDSmass software for remote display, spectra and ● historic retrieve, parameters settings SMI software for spectra analysis and identifi cation replay ●
Silicon protection ●
Belt Clip ●
Pouch with belt clip ●
External power supply / battery charger ●
Auto ID in progressAAAuto ID in progress
Detection mode screen
Search mode screen
Manual ID in progress
NMD result screen
Spectra view with PDSmass
MMMManManMan lllualualual IDIDIDIDID ID iiiinin in prpr progrogrogressessess
Auto ID in progress
NORM 40K, 226Ra and daughters, 232Th and daughters
Medical 18F, 51Cr, 67Ga, 75Se, 99mTc, 111In, 123I, 131I, 201Tl
Industrial 22Na, 57Co, 60Co, 133Ba, 137Cs, 152Eu, 192Ir, 241Am
SNM 235U, 238U, 239Pu
In bold, isotopes addressed by ANSI for SPRD