This document contains technical specifications and drawings for an InnoWave Unity steam sterilizer. The specifications provide details on the physical dimensions, weight, water supply requirements, drain configuration options, electrical power supply requirements, and operating environment requirements for the sterilizer. Key specifications include external dimensions of 42.5 inches height (closed), 58.9 inches height (open), 34.8 inches width, and 25.8 inches depth. The sterilizer requires a 208V or 230V electrical supply, water hardness below 125 ppm, water flow rate between 3 to 8 gallons per minute, drain capacity of 8 gallons per minute, and indoor operating location at altitudes up to 6,562 feet.
The related federal contract opportunity is a solicitation from the Department of Veterans Affairs for steam sterilizers. Responses are due by August 24, 2021 at 8:00 AM Central Time. The solicitation is set aside for service-disabled veteran-owned small businesses and will result in a firm fixed-price contract award. The North Texas VA Healthcare System is the contracting office located in Dallas, Texas.
View the file
Other files
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
T I T L E
Corporation SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER: REVISION:
This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.
SHEET
A
OF 71
NTS TJ C1101174610982297
Equipment Specifications InnoWave Unity
Physical dimensions and weight
External height (lid closed)
42.5 in. 1078 mm
External height (lid open)
58.9 in. 1495 mm
External width 34.8 in. 882 mm
External depth
25.8 in. 655 mm
Net weight (dry)
298 lb
135 kg
Operating weight (tank filled)
397 lb
180 kg
Max. operating weight (incl. filled tank, chemicals, and instruments to be cleaned)
430 lb
195 kg
External Dimension and Operating Clearances
34.8in. [882mm]
42.5in. [1078mm]
58.9in. [1495mm]
25.8in. [655mm]
T L E
Corporation SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER: REVISION:
This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.
SHEET
A
OF 72
NTS TJ C1101174610982297
Equipment Specifications InnoWave Unity
Drain Configuration Option 1 i n m m i n m m
2.38 in. [60 mm]
5.13 in. [130 mm]
13.88 in. [352 mm]
21.63 in. [549 mm]
2.75 in. [70 mm]
4 in. [100 mm] MIN
PRINTER CONNECTION
POWER MAINS
ENTRY
DRAIN 3/4" HOSE TAIL
HOT WATER 3/4"
COLD WATER 3/4"
OVERFLOW
OPEN FLOOR SINK
(components & installation by others)
ETHERNET
2 in. [51 mm] MIN
2 in. [51 mm] MIN
FOR SEISMIC ANCHORING
6.0 in. [152.4mm] MIN
FOR SEISMIC ANCHORING
6.0 in. [152.4mm] MIN
FOR SEISMIC ANCHORING
5.31 in. [135 mm] MIN
FOR SEISMIC ANCHORING
5.31 in. [135 mm] MIN
1.25 in. [33 mm]
1.25 in. [33 mm] i n m m
M
I N
4 in. [102 mm]
POSITION CENTRAL TO MACHINE
i n m m
DRAIN
FLEXIBLE PLASTIC
HOSE
OPEN FLOOR SINK
(components & installation by others) i n m m
5.88 in. [150 mm] MIN
A
R
G
A P
F
E Q
U I R
D
B
Y
L
O C
L
G
U L
A T
I O
N S
1.5 in. [40 mm] MIN
4.0 in. [100mm] MIN
26.75 in. [679.5 mm]
OVERFLOW
FLEXIBLE PLASTIC
HOSE
T L E
Corporation SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER: REVISION:
This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.
SHEET
A
OF 73
NTS TJ C1101174610982297
Equipment Specifications InnoWave Unity
Drain Configuration Option 2
DRAIN
FLEXIBLE PLASTIC
HOSE
OVERFLOW
FLEXIBLE PLASTIC
HOSE
4 in. [100mm] MIN.
(components & installation by others)
A
G
P
Q
U I R
D
B
Y
L O
C A
L
R
E G
U L A
T I O
N S
Ø1.5 IN. [40mm] MIN
DRAIN PIPE
6 in. [150 mm] MIN. 26 in. [655 mm] i n m i n
Ø40mm [1.5] MIN
DRAIN PIPE
i n m m
M
A X
Ø40mm [1.5]
MIN DRAIN POSITION VARIABLE
(components & installation by others)
18.75 in. [475 mm]
T L E
Corporation SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER: REVISION:
This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.
SHEET
A
OF 74
NTS TJ C1101174610982297
Equipment Specifications InnoWave Unity
Drain Configuration Option 3
DRAIN
FLEXIBLE PLASTIC
HOSE
OVERFLOW
FLEXIBLE PLASTIC
HOSE
(components & installation by others)
SUITABLE WALL BRACKETS
i n
Ø4.0 in. [110 mm] MIN.
REDUCER
6 in. [150 mm] MIN.
26 in. [655 mm] i n m m
M
A X
Ø1.5 in. [40 mm] MIN. DRAIN
POSITION VARIABLE
(components & installation by others)
22.75 in. [577 mm]
Ø4.0 in. [110 mm] MIN
REDUCER
(com ponents & installation by others)
D
R A
N
L L
Ø m m
M
I N
S U
P P
O R
T
U
S I N
G
W
A L L
P
I P
E
B
R A
C K
E
S
T L E
Corporation SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER: REVISION:
This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.
SHEET
A
OF 75
NTS TJ C1101174610982297
Equipment Specifications InnoWave Unity
Water supply requirements:
Water hardness (CaCO3), quality and purity, hot and cold supplies
< 125 ppm with water quality and purity tested and approved for disinfection as per ANSI/AAMI ST 15883-1:2009 (item 6.4.1 and
6.4.2) or AAMI TIR 34
Pressure (hot and cold) 30 to 70 psi (2.0 to 5.0 Bar)
Flow rate (hot and cold) 3 to 8 gal/min. (11 to 30 litres/min.)
Temperature (hot and cold) Cold: +41 to 86 ºF (+5 to 30 ºC) Hot: +122 to 176 ºF (+50 to 80 ºC)*
Connections
3/4 in. BSP male thread Rubber gasket is required. Do not use tape or any sealant, 3/4" NPT facility connections required. Hose connections will not work.
Position (hot and cold) Within 3 ft (1 m) of proposed unit location
*The unit is able to operate without hot water connection or with hot water supply cooler than recommended minimum temperature of +122 ºF (+50 ºC). Using only cold water or low-temperature hot water will result in longer cycle operation as water takes longer to heat up in unit.
Site drain requirements The site drain should meet the following requirements:
Drain connection
3/4 in. hose tail outlet from unit.
Drain must be fitted with suitable traps.
No loops in the drain hose, cut the hose to length to required.
Position
Floor sink or floor drain is preferred. Standpipe and wall drains are options for connection to drain. Adhere to local code.
Capacity 8 gal/min. (30 litres/min) Maximum discharge to drain including overflow.
Overflow Drain
An air tight seal of overflow to drain may cause incorrect unit operation.
An air gap must be fitted if required by code.
Effluent temperature up to +140 ºF (+60 ºC) Max.
Typical water consumption per wash
Water consumption will vary depending on volume of instruments loaded and temperature of inlet water supply. The water consumption below describes complete wash cycle at +104 ºF (+40 ºC), with no instruments loaded.
Maximum water consumption for complete wash cycle
Stage
Cold water Hot water
Total for stage
Pre-wash
10.6 gal (40 litres) nil
10.6 gal (40 litres)
Wash
4.2 gal (16 litres) 6.3 gal (24 litres) 10.6 gal (40 litres)
Rinse
10.6 gal (40 litres) nil
10.6 gal (40 litres)
Final Rinse
10.6 gal (40 litres) nil
10.6 gal (40 litres)
Total water consumption 35.9 gal (136 litres) 6.3 gal (24 litres) 42.3 gal (160 litres)
T L E
Corporation SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER: REVISION:
This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.
SHEET
A
OF 76
NTS TJ C1101174610982297
Equipment Specifications InnoWave Unity
Electrical power supply requirements:
Requirement
Model
IW523036 IW523046
Electrical supply voltage
230 V AC 1P+N+PE
(Single-phase, neutral, and protective earth)
Frequency: 60 Hz
208 V AC 3P+PE
(3-phase and protective earth)
Frequency: 60 Hz
Supply current
Site Current Capacity > 15 A per phase
Dedicated circuit is required with a circuit breaker rating of 16A or greater.
Electrical power requirement
4.0 kW 3.0 kW
Allowable supply voltage fluctuation ±10% of nominal voltage
Over-voltage category II Transient over-voltages typically present on mains supply
Electrical connections
Model Conductor
Color and Marking
IW523046 (208 V AC 3P+PE)
supplied with 2.5 m (8.5 ft) 4-core power cable.
Protective earth Green
Live 1 Black 1
Live 2 Black 2
Live 3 Black 3
IW523036 (230 V AC 1P+N+PE)
suppliedwith 2.5 m (8.5 ft) 3-core power cable
Protective earth Green
Live 1 Black 1
Neutral or Live 2 Black 2
Typical power consumption per wash
Power consumption will vary depending on cycle settings and temperature of inlet water supply. The power consumption below describes complete wash cycle at +104 ºF (+40 ºC).
Maximum power consumption for complete wash cycle
Stage Total for stage
Pre-wash 10.7 W.hr
Wash 57.2 W.hr
Rinse 6.8 W.hr
Final Rinse 6.8 W.hr
Total power consumption
81.5 W.hr
T L E
Corporation SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER: REVISION:
This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.
SHEET
A
OF 77
NTS TJ C1101174610982297
Equipment Specifications InnoWave Unity
Operating / Working environment location requirements
General and environmental requirements
Proposed operating location for unit much meet or exceed following requirements
Description Requirements
Operating surface
On level floor, capable of supporting a weight of at least 430 lb
(195 kg).
Location requirement Altitude Indoors only, altitude up to 6562 ft (2000 m)
Ambient operating temperature range
+41 ºF to 104 ºF (+5 ºC to 40 ºC)
Maximum humidity
80% relative humidity for up to +88 ºF (+31 ºC) decreasing linearly to 50% relative humidity at +104 ºF (+40 ºC)
Site services
Within 6 ft (1.8 m) of electrical isolator.
Within 3 ft (1.0 m) of hot and cold water connections.
Within 6 ft (1.8 m) of external site drain for effluent.
Ambient lighting 500 lux (46.5 foot-candle) minimum room lighting
Noise generated during operation 61 dBA at distance of 3.3 ft (1 m) from front of unit
Heat output during operation 205 W (700 BTU/hr) during wash cycle
Shipping crate dimensions and weight
Shipping crate height
49.2 in. 1250 mm
Shipping crate width
41.0 in. 1040 mm
Shipping crate depth
28.7 in. 730 mm
Shipping crate gross weight
375 lb
170 kg
Shipping information
Transportation and storage
Make sure package marking instructions and environmental conditions are always followed during transport and storage of InnoWave™ Unity.
Lifting
Lifting and maneuvering the InnoWave™ Unity in shipping crate must be done with pallet truck or forklift truck able to carry load in excess of 375 lb (170 kg).
Storage markings
The markings on shipping crate stipulate unit's storage conditions and handling information.
Keep dry This way up
File details come from the government source that posted it. Updated .