This document provides details for a federal solicitation seeking a steam sterilizer. The Department of Veterans Affairs North Texas VA Healthcare System is soliciting quotes for a steam sterilizer to be provided under a firm fixed-price contract. The solicitation is set aside for service-disabled veteran-owned small businesses, who must be registered in SAM to submit a quote. Quotes are due by August 24, 2021 at 8:00 AM Central Time and should be submitted to Rolando Pasayan at rolando.pasayan@va.gov. The NAICS code for this procurement is 339113. The contracting office issuing this solicitation is located in San Antonio, Texas.
View the file
Other files
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
80-3/4 [2051]
UNIT
HEIGHT
31 [787]
LOADING
HEIGHT
26-1/4 [667]
LOADING
CLEARANCE
HEIGHT
8-1/2 [216]
[381]
91 [2311]
RAISED DOOR
HEIGHT
34 [864]
MAX LENGTH
32 [813]
UNIT LENGTH
90 [2286]
SERVICE AREA
HEIGHT
45 [1143]
WALL OPENING
*(NOTE F)
1-1/2 [38]
MINIMUM
94 [2388] *(NOTE G)
RECOMMENDED
98 [2489] *(NOTE H)
MAX 1 [25]
ADJUSTMENT WITH
WASHER LEGS
(UNIT HEIGHT 81-3/4 [2077] MAX)
MINIMUM
90 [2286] *(NOTE G)
RECOMMENDED
96 [2438] *(NOTE H)
42 [1067]
UNIT WIDTH
T I T L E
EQUIPMENT DRAWING OF
AMSCO® 7052HP SINGLE-CHAMBER WASHER/DISINFECTOR
ELECTRIC HEATED, W/O ACU-RINSE RES., UNIT OFFSET FROM WALL
Corporation
SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER
& REVISIONA N/A Jean Tanguay 10292225
This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.
SERVICE AREA
REQUIRED FOR
DRYING MOTOR
REMOVAL
TYPICAL INSTALLATION
NOTES:
(A) UNLESS OTHERWISE SPECIFIED, ALL DIMENSIONS ARE IN INCHES [mm].
(B) SERVICE ACCESS SIDE OF UNIT IS RECOGNIZED BY THE TOUCH SCREEN AT THE RIGHT OF
WASH CHAMBER. ELECTRICAL AND CONTROL COMPONENTS, AS WELL AS MOST OF
SERVICABLE PARTS ARE ACCESSIBLE FROM THAT SIDE OF UNIT. NON-SERVICE SIDE IS
EQUIPPED WITH THE TOUCH SCREEN ON THE LEFT OF THE WASH CHAMBER.
(C) CUSTOMER MUST BE SURE THAT MACHINE STANDS ON A NON-COMBUSTIBLE FLOOR.
(D) WEIGHTED EQUIVALENT SURFACE SOUND PRESSURE LEVEL: 64.0 dB(A) CALCULATED AS DESCRIBED IN ISO 3746.
(E) VERTICAL BARRIER WALL FLANGES ARE PROVIDED WITH UNIT TO FILL GAP BETWEEN
UNIT AND WALL ON BOTH SIDES. DIMENSIONS ARE 3-1/2 in [89 mm] WIDTH X 80 in [2032 mm] LONG.
(F) HORIZONTAL BARRIER WALL FLANGES ARE PROVIDED WITH UNIT TO FILL GAP BETWEEN
UNIT AND WALL ON BOTH SIDES. DIMENSIONS ARE 49 in [1245 mm] WIDTH X 16 in [406 mm] HEIGHT.
(G) MINIMUM REQUIRED DIMENSION FOR DOOR REMOVAL WITH LIMITED ADJUSTMENT AVAILABLE: +0 in [0 mm] THE STANDARD BARRIER WALL FLANGES CAN'T BE USED AT THAT DIMENSION. -3/4 in [19.1 mm]
(H) RECOMMENDED DIMENSION FOR DOOR REMOVAL WITH FULL LEVELING ADJUSTMENT AVAILABLE: +1 in [25.4 mm] THE STANDARD BARRIER WALL FLANGES CAN BE USED AT THAT DIMENSION. *(NOTE E AND F) -3/4 in [19.1 mm]
FRONT VIEW
SERVICE ACCESS
SIDE OF UNIT
*(NOTE B)
TOP VIEW
PASS THROUGH
DOOR
SIDE VIEWLEVEL FLOOR
*(NOTE C)
W A
LL
O
P E
N
IN
G H
E
IG
H T
C E
IL
IN
G H
E
IG
H T
FO
R
D O
O R
R E
M O
V A
L
[102]
Page 1 of 10 10292225.C
Released vbravo Cloud+ vbravo Cloud+ 96" Required. 90" will NOT work for the trim decon side for this mode. Min. Ceiling to use trim is 98" AFF.
2X 33-7/16 [849]
2X 27-11/16 [704]
5-11/16 [144]
7-1/8 [181]
4X 15 [381]
1-7/8 [47]
17-3/8 [441]
T I T L E
EQUIPMENT DRAWING OF
AMSCO® 7052HP SINGLE-CHAMBER WASHER/DISINFECTOR
ELECTRIC HEATED, W/O ACU-RINSE RES., UNIT OFFSET FROM WALL
Corporation
SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER
& REVISIONA N/A Jean Tanguay 10292225
This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.
*NOTES:
(A) CONCRETE FLOOR MUST REMAIN FREE OF DEFECTS IN THESE AREA, AS PER ESR-1917 & ESR-3187.
FOR INSTALLATION OF SEISMIC ANCHORS (NOT PROVIDED BY STERIS).
(B) THE FLOOR SINK AND DRAIN SHALL NOT OCCUPY THESE AREAS. 5 in x 5 in [127 mm x 127 mm] FLOOR SINK WITH Ø4 in [102 mm] DRAIN IS RECOMMENDED WHEN USING SEISMIC MOUNTS.
SEISMIC MOUNTS
SEISMIC ANCHORAGE
FOR DETAIL SEE FD354
QTY 4 ANCHORS
FINISHED WALL
PREFERENCE
LOCATION OF
FLOOR SINK
*(NOTE A AND B)
Page 2 of 10 10292225.C
Released
80-3/4 [2051]
7-1/8 [181]
[381]
31-1/8 [791]
[432]
19-1/32 [483]
16-17/32 [420]
15-1/2 [394]
3-7/8 [99]
HW HOT WATER
CW COLD WATER
PW PURE WATER
A COMPRESSED AIR
D FLOOR DRAIN LOCATION
T I T L E
EQUIPMENT DRAWING OF
AMSCO® 7052HP SINGLE-CHAMBER WASHER/DISINFECTOR
ELECTRIC HEATED, W/O ACU-RINSE RES., UNIT OFFSET FROM WALL
Corporation
SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER
& REVISIONA N/A Jean Tanguay 10292225
This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.
UTILITY (PLUMBING)
CONNECTION POSITION
FRONT VIEW
A
D
PW
HW
CW
SIDE VIEW
HW
PW
CW
A
D
TOP VIEW
HW
CW
PW
A
D
Page 3 of 10 10292225.C
79-1/4 [2013]
35-13/16 [910]
18-3/8 [467]
[534]
38-1/8 [967]
39-5/8 [1007]
E ELECTRICAL
N NETWORK (SEE NOTE)
T I T L E
EQUIPMENT DRAWING OF
AMSCO® 7052HP SINGLE-CHAMBER WASHER/DISINFECTOR
ELECTRIC HEATED, W/O ACU-RINSE RES., UNIT OFFSET FROM WALL
Corporation
SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER
& REVISIONA N/A Jean Tanguay 10292225
This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.
UTILITY (ELECTRICITY)
CONNECTION POSITION
NOTE:
NETWORK CONNECTION (IF APPLICABLE)
- ACTIVE NETWORK DROP WITH CAT 5/6 CABLE WITHIN 10 ft [3 m] OF UNIT
- STATIC OR DHCP RESERVED IP ADDRESS (FOR EACH UNIT)
- NETWORK CABLE (SUPPLIED BY OTHER)
FRONT VIEW SIDE VIEW
E
N
E
N
TOP VIEW
E
N
Page 4 of 10 10292225.C
6-5/32 [156]3
[76]
81-29/32 [2080]
T I T L E
EQUIPMENT DRAWING OF
AMSCO® 7052HP SINGLE-CHAMBER WASHER/DISINFECTOR
ELECTRIC HEATED, W/O ACU-RINSE RES., UNIT OFFSET FROM WALL
Corporation
SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER
& REVISIONA N/A Jean Tanguay 10292225
This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.
EX VENTILATION EXHAUST
UTILITY (VENTILATION)
CONNECTION POSITION (OPTIONAL)
FRONT VIEW SIDE VIEW
EX
SUGGESTED DUCTWORK
(BY OTHERS)
EX
DRAIN 3/4 in [20 mm] O.D.
(ROUTED TO WASHER DRAIN)
DRY CONTACT (SEE PAGE 10 *(NOTE 3))
VAPOR CONDENSATE
COLLECTING POINT
TOP VIEW
EX
VENTILATION CONNECTION (OPTIONAL)
(SEE PAGE 9 *(NOTE 1))
EX
BY OTHERS
Page 5 of 10 10292225.C
60 [1524]
REF
8-1/2 [216]
132 [3350]
MAX.
17 [432]
INSTRUMENT
(With PUC HP)
UTENSIL
(With PUC HP)
CHEMICAL
(CH)
ENZYME CONSUMPTION PER CYCLE
(SEE PAGE 7; *(NOTE 4)) 0.74 oz [22 ml] 0.74 oz [22 ml]
DETERGENT CONSUMPTION PER CYCLE
(SEE PAGE 7; *(NOTE 4)) 0.74 oz [22 ml] 0.74 oz [22 ml]
LUBRICANT CONSUMPTION PER CYCLE
(SEE PAGE 7; *(NOTE 4)) 0.2 oz [6 ml] N/A
T I T L E
EQUIPMENT DRAWING OF
AMSCO® 7052HP SINGLE-CHAMBER WASHER/DISINFECTOR
ELECTRIC HEATED, W/O ACU-RINSE RES., UNIT OFFSET FROM WALL
Corporation
SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER
& REVISIONA N/A Jean Tanguay 10292225
This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.
*NOTES:
(A) MAXIMUM CONTAINER HEIGHT FROM FLOOR IS 24 in [610 mm]. AMBIENT TEMPERATURE SHALL NOT EXCEED 40 °C [104 °F]
DOSING STATION
CEILING
ROUTE CHEMICAL LINES
AND SIGNAL CABLE
THROUGH PIPE TO REACH
BOTTOM OF UNIT
CHEMICAL CONTAINERS
*(NOTE A)
12 ft [3.6 m] OF INJECTION LINE
(FROM CONTAINERS TO
DOSING STATION)
DOSING STATION
WEIGHT : 15 lb [6.8 kg]
2X FD349 SHOWN
UP TO 5 TUBES
100 ft [30 m] OF INJECTION LINE
AND SIGNAL CABLE
DETERGENT LINES SHALL NOT BE IN CONTACT WITH HOT SURFACES.
STERIS RECOMMENDS CUSTOMER TO ROUTE CHEMICAL LINES AND
SIGNAL CABLE THROUGH A 3 in [76 mm] ID CONDUIT.
EXCESS LENGTH MUST BE CUT BEFORE CONNECTING TO
WASHER/STATION TO AVOID FORMING BUNDLES THAT COULD
AFFECT THE CHEMICAL INJECTION SYSTEM'S PERFORMANCE
(DO NOT CUT SIGNAL CABLE).
Page 6 of 10 10292225.C
UTILITY REQUIREMENTS CHART (PLUMBING)
UTILITY PARAMETER REQUIREMENT NOTES
HOT WATER
(HW)
CONNECTION SIZE / TYPE 1/2 in / FEMALE NPT [1/2 in / MALE BSPT]
1. STERIS REQUIRES TO INSTALL WATER HAMMER
ARRESTOR, UNION & PRESSURE GAUGE.
2. CUSTOMER MUST PROVIDE UTILITY
CONNECTIONS WITH SHUTOFFS DISCONNECTS
WITHIN 2 FEET OF THE PERIMETER OF THE
EQUIPMENT AND BELOW THE CEILING DECK.
*STERIS OFFERS FLEXIBLE HOSES
ACCESSORY (FD088 OR FD037)
WITH INSTRUCTIONS.
3. SAMPLING VALVES INSTALLATION AS CLOSE AS
POSSIBLE TO WASHER WATER CONNECTIONS IS
RECOMMENDED.
PRESSURE RANGE
15 - 50 psig DYNAMIC
[103-345 kPa] MAX. 90 psig STATIC [620 kPA]
OPERATING FLOW RATE 6.8 US GPM [25.7 L/min] TO
14.1 US GPM [53.4 L/min]
TOTAL CONSUMPTION PER
INSTRUMENT CYCLE
*(NOTE 4)
10.5 US GAL [39.8 L]
WITH ACU-WASH RESERVOIR:
7.0 US GAL [26.5 L]
TEMPERATURE RANGE 110 °F [43 °C] TO 150 °F [65.5 °C]
WATER QUALITY *(NOTE 5) HARDNESS (CaCO3) : 50-80 ppm (MAX. 120 ppm)
COLD WATER
(CW)
*(NOTES 6)
CONNECTION SIZE / TYPE 1/2 in / FEMALE NPT [1/2 in / MALE BSPT]
4. CONSUMPTION BASED ON INSTRUMENT CYCLE WITH
PROLYSTICA ULTRA CONCENTRATE HIGH
PERFORMANCE CHEMICALS (PUC HP) WITH
THE FOLLOWING UTILITY CONDITIONS:
THREE-LEVEL MANIFOLD RACK
(CW) COLD WATER PRESSURE: 30 psig DYNAMIC [207 kPa]
(CW) COLD WATER TEMPERATURE: 70 °F [21 °C]
(HW) HOT WATER TEMPERATURE: 150 °F [65 °C]
(PW) PURE WATER TEMPERATURE 70 °F [21 °C]
UTENSIL CYCLE CONSUMPTION:
TWO-LEVEL MANIFOLD RACK
CW: 11.0 US GAL [41.6 L]
HW: 10.2 US GAL [38.6 L]
PW: 6.8 US GAL [25.7 L]
5. STERIS RECOMMENDS THE FOLLOWING WATER
QUALITY PARAMETERS:
TOTAL DISSOLVED SOLIDS: 100-200 ppm (MAX. 500 ppm) TOTAL ALKALINITY (CaCO3) : 70-120 ppm (MAX. 180 ppm) pH : 6.8-7 (MAX. 7.5) TOTAL SILICA : 0.1-0.5 ppm (MAX. 1 ppm)
PRESSURE RANGE
30 - 50 psig DYNAMIC
[206-345 kPa] MAX. 90 psig STATIC [620 kPa]
OPERATING FLOW RATE 10.7 US GPM [40.5 L/min] TO
14.1 US GPM [53.4 L/min]
TOTAL CONSUMPTION PER
INSTRUMENT CYCLE
*(NOTE 4)
FOR SUMP FILLING: 11.1 US GAL [42.0 L]
(ADD THE FOLLOWING CONSUMPTION
FOR OPTIONS/CONFIGURATIONS)
NON-VENTED SYSTEM:
20.7 US GAL [78.4 L]
DRAIN COOLDOWN:
14.5 US GAL [54.9 L]
DRAIN ACU-COOLDOWN:
10.9 US GAL [41.3 L]
TEMPERATURE RANGE 70 °F [21 °C] MAXIMUM
WATER QUALITY *(NOTE 5) HARDNESS (CaCO3) : 50-80 ppm (MAX. 120 ppm)
PURE WATER
(PW)
CONNECTION SIZE / TYPE 1/2 in / FEMALE NPT [1/2 in / MALE BSPT]
6. BACK FLOW PREVENTER AGAINST BACK
SIPHONAGE MAY BE REQUIRED (LOCAL
PLUMBING CODE)
7. PW IS OPTIONAL. IF UNAVAILABLE, UNIT WILL
PROCESS THERMAL RINSE USING HOT WATER
SUPPLY.
PRESSURE RANGE 5 - 30 psig DYNAMIC [30 - 206 kPa] MAX. 90 psig STATIC [620 kPa]
OPERATING FLOW RATE 5.5 US GPM [21 L/min] TO
12.6 US GPM [48 L/min]
TOTAL CONSUMPTION PER
INSTRUMENT CYCLE
*(NOTE 4 AND NOTE 7)
7.0 US GAL [26.5 L]
TEMPERATURE RANGE 70 °F [21.1 °C] TO 150 °F [65.5 °C]
RESISTIVITY
(CONDUCTIVITY)
0.1-0.5 M -cm (2-10 µS/cm) pH 6.8 - 7.5
T I T L E
EQUIPMENT DRAWING OF
AMSCO® 7052HP SINGLE-CHAMBER WASHER/DISINFECTOR
ELECTRIC HEATED, W/O ACU-RINSE RES., UNIT OFFSET FROM WALL
Corporation
SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER
& REVISIONA N/A Jean Tanguay 10292225
This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.
Page 7 of 10 10292225.C
UTILITY REQUIREMENTS CHART (PLUMBING)
COMPRESSED
AIR
(A)
CONNECTION SIZE / TYPE 1/8 in / FEMALE NPT [1/8 in / MALE BSPT]
1. AIR PURITY PER ISO-8573-1: 2010 [7:4:4]
MAX. PARTICLE DENSITY: 10 mg/m³
MAX. DEW POINT FOR WATER CONTENT:
37 °F [3 °C]
MAX. OIL CONCENTRATION FOR THE OIL
CONTENT: 5 mg/m³
2. STERIS RECOMMENDS A REFRIGERATED AIR
DRYER WHERE ENVIRONMENTAL DEW POINT
CONDITIONS ARE HIGHER THAN
RECOMMENDED.
3. CUSTOMER MUST PROVIDE UTILITY
CONNECTIONS WITH SHUTOFFS DISCONNECTS
WITHIN 2 FEET OF THE PERIMETER OF THE
EQUIPMENT AND BELOW THE CEILING DECK.
PRESSURE RANGE MIN 80 psig DYNAMIC [552 kPa] MAX. 125 psig STATIC [860 kPa]
OPERATING FLOW RATE 1.75 SCFM [50 L/min]
DRAIN
(D)
FLOOR DRAIN SIZE MIN 4 in [102 mm] O.D. WITH 8 in x 8 in [204 mm x 204 mm] FLOOR SINK
4. WASHER DRAIN OUTLET SIZE 1-1/2 in / MALE NPT
5. MAX OFFSET OF 3/4 in [19 mm] FROM
OUTLET (PAGE 3, D).
6. FOR DRAIN REQUIREMENTS WHEN USING
SEISMIC MOUNTS, SEE PAGE 2.
OPERATING FLOW RATE GRAVITY DRAIN : 50 US GPM
[189.3 L/min]
T I T L E
EQUIPMENT DRAWING OF
AMSCO® 7052HP SINGLE-CHAMBER WASHER/DISINFECTOR
ELECTRIC HEATED, W/O ACU-RINSE RES., UNIT OFFSET FROM WALL
Corporation
SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER
& REVISIONA N/A Jean Tanguay 10292225
This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.
CHEMICAL
DOSING
SYSTEM
CEILING ROUTING
CONDUIT 3 in CONDUIT
11. STERIS RECOMMENDS TO ROUTE CHEMICAL
LINES AND SIGNAL CABLE THROUGH A
3 in [76 mm] ID CONDUIT (BY OTHERS).
100 ft [30 m] OF INJECTION LINE AND
SIGNAL CABLE ARE PROVIDED WITH THE UNIT.
Page 8 of 10 10292225.C
29-1/2 [749]
[406]
[305]
20-3/8 [518]
24-1/4 [616]
[178]
26 [660]
24 [610]
22 [559]
24 [610]12-1/2 [318]
[152]
[51]
T I T L E
EQUIPMENT DRAWING OF
AMSCO® 7052HP SINGLE-CHAMBER WASHER/DISINFECTOR
ELECTRIC HEATED, W/O ACU-RINSE RES., UNIT OFFSET FROM WALL
Corporation
SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER
& REVISIONA N/A Jean Tanguay 10292225
This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.
UTILITY REQUIREMENTS CHART (ELECTRIC)
UTILITY
PARAMETER AND REQUIREMENT NOTES
CONNECTION
SIZE/TYPE
NOMINAL
AMPERAGE
RECOMMENDED
CIRCUIT
ELECTRICITY
(EL)
208V 60 Hz, 3-Ph
CONVERTED TO
460-480V
THROUGH TRANSFORMER
(OPTIONAL) FD787
N/A N/A
(NOTE 7)
N/A
(NOTE 7)
1. ALL CONNECTIONS SHOULD BE IN
ACCORDANCE WITH LOCAL CODES.
2. NEUTRAL WIRE NOT REQUIRED.
3. ALWAYS FOLLOW LOCAL ELECTRICAL CODES
AND SAFETY-RELATED WORK PRATICES TO
DETERMINE THE VALUE OF PROTECTION.
4. WIRING ON THE EQUIPMENT TERMINATES AT
A JUNCTION BOX. WIRING BETWEEN JUNCTION
BOX AND BUILDING SERVICE LINES ARE NOT
SUPPLIED BY STERIS.
5. DISCONNECT SWITCHES WITH OFF POSITION
LOCKOUT (SUPPLIED BY OTHERS) MUST BE
INSTALLED IN ELECTRIC SUPPLY LINES NEAR
THE EQUIPMENT.
6. STERIS RECOMMENDS A LIGHT IN SERVICE
AREA.
7. IT IS CUSTOMER RESPONSABILITY TO
CALCULATE NOMINAL AMPERAGE REQUIRED AT
TRANSFORMER INPUT. NOMINAL AMPERAGE
REQUIRED AT TRANSFORMER OUTPUT TO
SUPPLY THE WASHER CAN BE FOUND IN THIS
TABLE FOR THE SPECIFIC ELECTRICAL
CONFIGURATION OF THE WASHER.
200V 50 Hz, 3-Ph
CONVERTED TO
380-400-415V
THROUGH TRANSFORMER
(OPTIONAL) FD78-800
N/A N/A
(NOTE 7)
N/A
(NOTE 7)
460-480V 60 Hz, 3-Ph
1 in [27 mm] 1-1/4 in [35 mm]
CONDUIT
28.5 A 40 A
380-400V 60 Hz, 3-Ph
1 in [27 mm] 1-1/4 in [35 mm]
CONDUIT
32 A 40 A
380-400-415V 50 Hz, 3-Ph
1 in [27 mm] 1-1/4 in [35 mm]
CONDUIT
30.5 A 40 A
208V, 60Hz 200V, 50Hz POWER (kVA) 45
INPUT (HV) (V 3, 60Hz) 208/240/600 208
OUTPUT (LV) (V 3, 60Hz) 480 231 / 400
WEIGHT LB (KG) 467 (212) 343 (156)
NOISE LEVEL (dBa) 45
NO LOAD LOSS (WATTS) 241.2 319
LOAD LOSS
338°f (170°C)
(WATTS) 1062.6 2461
TOTAL LOSSES (WATTS) 1303.8 2780
EXCITATION
CURRENT (%) 2.54 3.26
INSTALLATION WALL/FLOOR
SIZE IN (MM) SEE RIGHT
PRIMARY CONNECTORS
PER PHASE
(MCM) 1x 250 MCM
SECONDARY CONNECTORS
PER PHASE
(AWG) 1x #2 AWG-#14 AWG
TEMPERATURE
CLASS
°F (°C) 428° (220°)
WINDING (HV) (LV) COPPER
ENGINEERING DATA-RECOMMENDED
TRANSFORMER
TRANSFORMER
Page 9 of 10 10292225.C
UTILITY REQUIREMENTS CHART
VENTILATION
EXHAUST
(EX)
OPTIONAL
CONNECTION SIZE 3 in [76 mm]
OUTSIDE DIAMETER
1. NOT REQUIRED WITH NON VENTED UNIT.
2. STERIS REQUIRES CORROSION-RESISTANT AND
WATER-TIGHT DUCT:
- MIN OPERATING TEMPERATURE 210 °F [99 °C]
- SLOPED TOWARDS THE WASHER AND NO DEAD LEGS
- INSTALL DRAIN PART AT LOWEST POINT NEAR UNIT
3. EXHAUST FAN NORMALLY OPEN-AUXILIARY CONTACTS
IS AVAILABLE TO SUPPLY A SIGNAL TO THE FACILITY HVAC
CONTROL SYSTEM WHENEVER THE WASHER VENTILATION
EXHAUST FAN IS IN OPERATION.
PRESSURE RANGE 1/2 in [50 mm]
VACUUM AT CONNECTION
OPERATING FLOW RATE 60 SCFM [1.7 m³/min]
TEMPERATURE RANGE MAX. 240 °F [116 °C]
PHYSICAL SPECIFICATIONS
SHIPPING DIMENSION W x H x L 50 in x 91 in x 42 in [1270 mm x 2311 mm x 1067 mm]
1. LEVELING FEET ARE PROVIDED FOR PROPER INSTALLATION.
2. PERMISSIBLE ENVIRONMENTAL CONDITIONS.
THIS EQUIPMENT IS DESIGNED TO GIVE OPTIMAL RESULTS
UNDER THE FOLLOWING CONDITIONS:
- INDOOR USE ONLY;
- ALTITUDE OF OPERATION UP TO 6562 ft [2000 m];
- TEMPERATURE OF 41 TO 104 °F [5 to 40 °C];
- MAXIMUM RELATIVE HUMIDITY IS 80% FOR TEMPERATURE UP
TO 88 °F [31 °C], DECREASING LINEARLY TO 50% RELATIVE
HUMIDITY AT 104 °F [40 °C];
- POLLUTION DEGREE 2: EQUIPMENT MUST BE INSTALLED IN AN
ENVIRONMENT WHERE NORMALLY ONLY NON-CONDUCTIVE
POLLUTION OCCURS BUT WHERE OCCASIONAL, TEMPORARY
CONDUCTIVITY CAUSED BY CONDENSATION CAN BE EXPECTED
(AS DETERMINED ACCORDING TO INTERNATIONAL STANDARD
EN/IEC 61010-1).
3. STERIS ASSUMES NO RESPONSIBILITY FOR CHANGES MADE TO,
OR FAILURE TO OBSERVE THE SPECIFICATIONS ON EQUIPMENT
DRAWING AND NOTE PAGES. SPECIFICATIONS AND
DESCRIPTIONS ARE SUBJECT TO CHANGE WITHOUT NOTICE.
SHIPPING WEIGHT 1268 lb [575 kg]
OPERATING DIMENSION W x H x L 42 in x 80.75 in x 32 in [1067 mm x 2051 mm x 813 mm]
OPERATING WEIGHT
(WITH WATER AND FULLY LOADED) 1254 lb [569 kg]
AVERAGE FLOOR LOAD
OVER WASHER FOOTPRINT 0.93 psi [6.43 kPa]
MAX. FLOOR LOAD
AT WASHER FOOT PADS
(3 in [76.2 mm] DIAMETER PADS) 67 psi [462 kPa]
NOISE LEVEL:
A-WEIGHTED EQUIVALENT
SURFACE SOUND PRESSURE LEVEL
60.6 dB(A)
CALCULATED AS DESCRIBED
IN ISO 3746.
OPENING AREA FROM
CLEAN SIDE TO SOILED SIDE
151.8 in² [979 cm²]
*(NOTE 4)
4. OPENING AREA CAN BE REDUCED TO 26 in² [168 cm²]
WITH INSTALLATION OF AIR MANAGEMENT PANELS
ACCESSORY (FD352)
CHAMBER DIMENSION W x H x L
VOLUME
26.5 in x 26.25 in x 25.5 in [673 mm x 667 mm x 648 mm]
10.3 ft³ [0.29 m³]
T I T L E
EQUIPMENT DRAWING OF
AMSCO® 7052HP SINGLE-CHAMBER WASHER/DISINFECTOR
ELECTRIC HEATED, W/O ACU-RINSE RES., UNIT OFFSET FROM WALL
Corporation
SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER
& REVISIONA N/A Jean Tanguay 10292225
This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.
TYPICAL HEAT LOSS AT 75 °F [24 °C]
CONTINUOUS OPERATION
NON-VENTED UNIT
WITHOUT ACU-RINSE RES.
VENTED UNIT (OPTIONAL)
WITHOUT ACU-RINSE RES.
CLEAN SIDE 2049 Btu/h [600 W]
2049 Btu/h [600 W]
SOILED SIDE 6557 Btu/h [1920 W]
4747 Btu/h [1390 W]
TOTAL 8606 ± 854 Btu/h [2520 ± 250 W]
6796 ± 854 Btu/h [1990 ± 250 W]
VALUES INCLUDE HEAT LOSS DURING DOOR OPENING FOR LOADING AND UNLOADING OPERATIONS. "TOTAL" WILL VARY DEPENDING ON DISINFECTION TIME AND TEMPERATURE, DRYING TIME AND TEMPERATURE AND USE RATE OF UNIT.
AN ADDITIONAL HEAT LOSS FOR LOAD COOL DOWN IN CLEAN SIDE CAN BE ADDED TO THE TOTAL HEAT LOSS OF THE WASHER:
MAXIMUM 1713 Btu [1807 kJ]
Page 10 of 10 10292225.C
| Sheet 1 |
| Views |
| new_view_35 |
| new_view_37 |
| new_view_57 |
| new_view_66 |
| Zones |
| A1 |
| Sheet 2 |
| Views |
| new_view_55 |
| Zones |
| A1 |
| Sheet 3 |
| Views |
| new_view_38 |
| new_view_39 |
| new_view_40 |
| Zones |
| A1 |
| Sheet 4 |
| Views |
| new_view_44 |
| new_view_45 |
| new_view_46 |
| Zones |
| A1 |
| Sheet 5 |
| Views |
| new_view_41 |
| new_view_42 |
| new_view_43 |
| new_view_65 |
| Zones |
| A1 |
| Sheet 6 |
| Views |
| new_view_47 |
| Zones |
| A1 |
| Sheet 7 |
| Zones |
| A1 |
| Sheet 8 |
| Zones |
| A1 |
| Sheet 9 |
| Views |
| new_view_64 |
| Zones |
| A1 |
| Sheet 10 |
| Zones |
| A1 |
File details come from the government source that posted it. Updated .