This is a Salient Characteristic List for Contract Line Item Number (CLIN) 001 specifying requirements for an OPD-Scan III Visual System or equivalent optical diagnostic device.

The equipment is a 10.4-inch color LCD touch panel optical imaging system with dimensions of 286mm (W) x 525mm (D) x 530mm (H) and a mass of 23kg. The system must operate on 100-240 VAC 50/60 Hz power supply with 110 VAC power consumption. The device shall include integrated components such as printer paper, power cord, dust cover, chinrest paper, fixing pins, spherical model eye, touch pen, touch pen stand, ferrite core, and installation CD with manual for the OPD Web Viewer System. The finish color shall be tank/cream.

Functionally, the system must provide combined capabilities for auto-refraction, auto-keratometry, topography, aberrometry, and pupillometry with a built-in thermal type line printer for data output and optional external color printer for map printing. Key performance specifications include spherical power range of -20.00 to +22.00 diopters, cylindrical power of 0.00 to ±12.00 diopters, axis range of 0 to 180 degrees, curvature radius of 5.00 to 10.00mm, and refractive power of 33.75 to 67.50 diopters. The system shall support multiple mapping methods including axial, instantaneous, refractive, elevation, Zernike graph, PSF, and MTF graph with a working distance of 75mm, auto-tracking in X-Y-Z directions, and measurement area of 14 x 11mm. Corneal topography measurement capability requires 33 vertical and 39 horizontal measurement rings across a 0.5 to 11.0mm area with more than 11,880 data points. Measuring time shall be less than 10 seconds using automated objective refraction via dynamic skiascopy. Delivery is required by 15 February 2027 with a minimum 1-year manufacturer warranty.

View the file

Other files

File Type Posted
SOW_Marco_OCT 2026.pdf PDF

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Salient Characteristic List

Key Salient Physical Characteristics: CLIN 001

a. Commercial Description OPD-Scan III Visual System or equivalent

b. Dimensions and Size 286 (W) x 525 (D) x 530 (H) mm; Mass 23kg

c. Material Composition 55lbs

d. Components and Parts

Printer paper, Power cord, Dust cover, Chinrest paper, Fixing pins for chinrest paper, Spherical model eye, Touch pen, Touch pen stand, Ferrite core, Installation CD for OPD Web Viewer System, Installation manual for OPD Web Viewer System

e. Compatibility and Connections Power Supply: 100 to 240 Vac 50 / 60 Hz Power Consumption: 110 VAC

f. Color and Finish Tank/Cream Color

g. Product Design Features 10.4-inch color LCD touch panel

Key Salient Functional Characteristics: CLIN 001 continued

a. Compatibility and Integration

Capability with Combined Functionality of:

i) Auto-refraction

ii) Auto-keratometry

iii) Topography

iv) Aberrometry

v) Pupillometry b Performance Metrics Built-in thermal type line printer for data print External color printer (optional) for map print

c. Operational Utility Power Supply: 100 to 240 Vac 50 / 60 Hz Power Consumption: 110 VAC d Service/Installation Requirements N/A e Quality Assurance Standards: N/A k.

Other pertinent information that describes the item, material or service required:

Curvature radius 5.00 to 10.00 mm, Refractive power 33.75 to 67.50 D (n = 1.3375), Astigmatism 0 to ±12.00 D, Axis 0 to 180°

Key Aspects of VA Performance Salient Characteristics: CLIN 001 continued

a. Functional/Performance Capabilities

Mapping Methods: Axial, Instantaneous, “Refractive”, Elevation, Zernike graph, PSF, MTF graph, Visual Acuity General Information Working Distance: 75 mm Auto Tracking: X-Y-Z directions Observation Area: 14 x 11 mm

b. Operational Suitability Spherical Power Range: -20.00 to +22.00 D Cylindrical Power: 0.00 to ±12.00 D

Axis: 0 to 180° Measurement Area: 2.0 to 9.5 mm (7 zone measurement) Data Points: 2,520 points (7 x 360) Measuring time: < 10 seconds Measurement Method: Automated objective refraction (dynamic skiascopy) Mapping Methods: OPD, Internal OPD, Zernike graph, PSF, MTF graph, Visual Acuity Corneal Topography Measurement Rings: 33 vertical, 39 horizontal Measurement Area: 0.5 to 11.0 mm (r = 7.9) Dioptric Range 33.75 to 67.5 D Axis Range: 0 to 359˚ Data Points: More than 11,880

c. Delivery & Reliability Time of delivery: 15 February 2027

d. Service & Support Minimum 1-year manufacturer warranty

e. Quality Standards N/A