This document is a Standard Operating Procedure (SOP) for particle size analysis of hardwood chips, issued by the National Center for Toxicological Research's Division of Biochemical Toxicology at the US Food and Drug Administration in Jefferson, AR. The procedure details a method using a Ro-Tap sieve shaker to analyze hardwood chip particle sizes, involving nested sieves (#8 with 2.36 micron openings and #20 with 850 micron openings) and a receiver pan. The process requires weighing 25.00g of sample, shaking the sieves for 3.0 minutes, and then measuring the weight retained on each sieve and pan. Particle size limits are specified as no more than 10% retained on the #8 sieve and no more than 0.7% in the receiver pan, with the procedure designed to comply with FDA Good Laboratory Practice (GLP) regulations as specified by 21 CFR 58.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| NCTR-2026-132197 Sources Sought Notice.pdf |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Uncontrolled when Printed SOP No. NCTR OSC CHEM 316.03, Page 1 of 2
National Center for Toxicological Research Division of Biochemical Toxicology
Chemistry Support Group US Food and Drug Administration, Jefferson, AR
Laboratory Operating Procedures SOP No: NCTR OSC CHEM 316.03
Prepared by: F. McLellen
Approved by OSC/Chemistry Team Leader:
Supersedes: 316.02 Date Approved Page 1 of 2 Effective Date: 00/00/00
Standard Operating Procedure Particle Size Analysis of Harwood Chips
1. PURPOSE:
This procedure is to describe the standard operating procedure (SOP) for analysis of particle size of hardwood chips in compliance with FDA Good Laboratory Practice (GLP) as specified by 21 CFR 58.
2. SCOPE:
This procedure applies to the determination of particle size of hardwood chips utilized at the
NCTR.
3. PRINCIPLE
A sieve shaker is used to shake two sieves of different sizes, and a receiver pan in order to determine the particle size of hardwood chips.
4. MATERIALS and SUPPLIES
a. Balance with accuracy to 0.01g
b. U.S. Standard #8 Sieve: 2.36 micron/0.093 inches
c. U.S. Standard #20 Sieve: 850 micron/0.0331 inches
d. Receiver Pan
e. Ro-Tap sieve shaker (I.E. Ro-Tap RX-812)
f. Scoop
5. Procedures
Note: Prior to weighing operations, the analytical balance is calibrated, and the calibration is documented with NIST traceable weights in the weight range of the intended operation.
A. Weigh each clean, empty sieve pan and record each empty weight.
B. Nest the sieve pans in the following order: #8 on top, #20 in the middle, and receiver pan on bottom. Place the nested sieve pans and receiver pan on the balance.
C. Tare the balance and weigh 25.00g into the #8 (top) sieve pan. Record this weight as the total sample weight.
D. Place the nested sieve pans and receiver pan on the Ro-Tap sieve shaker.
E. Using the Ro-Tap sieve shaker, shake sieves for 3.0 minutes.
F. Stop Ro-Tap sieve shaker and remove sieve pans and receiver pan from the Ro- Tap sieve shaker. .
G. Carefully separate the sieve pans and receiver pan from each other. Be careful not to spill the contents of any of the pans.
Uncontrolled when Printed SOP No. NCTR OSC CHEM 316.03, Page 2 of 2
H. Weigh each pan, to 0.01g, separately on a tared balance. Record each weight as the (sample weight+ pan weight) for that pan.
5. Calculations
Sample Weight after shaking= (Sample Weight + Pan Weight)- Pan Weight
Example: 25.00g of sample were placed in the #8 sieve. The nested sieves were then placed on the Ro-Tap sieve shaker and allowed to vibrate for 3.0 minutes. The results are listed.
Pan Pan Weight Sample Weight+ Pan Weight
Sample Weight after shaking.
#8 Sieve 300.00 301.72 1.72
#20 Sieve 300.00 323.01 23.01 Receiver Pan 300.00 300.29 0.29
% Retained = Sample wt after shaking x4 (1.72 x4 = 6.88 %)
7. Particle Size Limits (Updated 2025)
#8 Sieve – no more than 10% of the original sample weight.
Receiver Pan – no more than 0.7% of the original sample weight.