This document is a detailed Requirements/Specifications document for a Sample Holder for Thermo Fisher (S)TEM from the National Institute of Standards and Technology (NIST). The primary item requested is a heating/biasing holder for in operando studies of electrical and optical devices, with extensive technical specifications including:

Key technical requirements include: variable temperature range from ambient to ≥900°C, temperature accuracy of ≥95%, image stability of ≤1 nm/min at 900°C, controllable heating/cooling rates up to 100°C/ms, bias voltage range of 0-40V, current range of 2 pA to 100 mA, and simultaneous temperature and field application. The solicitation also includes a separately priced software option with features like offline image analysis, sample drift correction, and synchronous parameter recording. Additional requirements cover installation by vendor technician, on-site training for three personnel in operations/maintenance/troubleshooting, one-year warranty, and planned government acceptance testing over 7 business days. The document specifies compatibility with Thermo Fisher Spectra and Talos S/TEM instruments and compliance with the manufacturer's safety certifications.

View the file

Other files for this federal contract opportunity

Other files attached to Sampler Holder for TEM, newest first.
File Type Posted
Questions and Answers.pdf PDF
Revised RFQ BV Attach A Evaluation Criteria1357.docx DOCX document
Revised Requirements and Specs Document.docx DOCX document
Revised 1 RFQ Best Value1357.docx DOCX document
Atch B - Experience Submittal1357.pdf PDF
RFQ Best Value1357.docx DOCX document
RFQ BV Attach A Evaluation Criteria1357.docx DOCX document
Requirements and Specifications (1).pdf PDF
Statement of Work.docx DOCX document

On GovTribe

Work with this file on GovTribe

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

R

Requirements / Specifications: Sample Holder for Thermo Fisher (S)TEM 1 Version 1.0 (2019) Double-click on for further assistance.

Use this document to detail specifications/requirements for an item and/or ancillary services.

TITLE: Sample Holder for Thermo Fisher (S)TEM Requesting Lab/Division/Group: PML/Applied Physics Division/ Quantitative Nanostructure Characterization Group

General Statement of Need The heating/biasing (S)TEM sample holder is needed for in operando studies of a wide variety of electrical and optical devices.

Items and/or Ancillary Services

Item 1: Heating-Biasing Holder for Thermo Fisher (S)TEM Quantity: one Specifications: 1.The holder shall have a variable temperature range of operation from ambient to ≥900 °C (higher is better).2.The holder shall have a temperature accuracy of ≥95 % (higher is better).3.Image stability with a holder temperature of 900 °C shall be ≤1 nm/min (lower is better). 4.The holder shall have controllable heating and cooling rates up to 100 °C/ms.5.The temperature on the holder shall be determined either by in-situ temperature measurement or from calibration of the individual heating device on each microelectromechanical systems (MEMS) chip.6.The holder shall have a continuously controllable bias voltage range of 0 V to ≥40 V (higher is better).7.The holder shall have ripple and noise ≤10 mV.8.The holder shall have a current range of 2 pA to 100 mA.9.The holder shall be capable of simultaneous application of temperature and field up to 40 V at 900 °C.10.The holder shall include electrodes to perform 4-point probe electrical measurements simultaneously with sample heating.11.The system shall include software for temporally recording electrical and temperature control and measurement. 12.The holder shall have an α-tilt range of at least ± 30 ° (higher is better).13.The holder shall have a β-tilt range of at least ± 15 ° (higher is better).14.The holder shall have friction-free tilting in both alpha and beta directions to prevent electrical contacts on the chip from scraping against the probes and to preserve heating and electrical integrity.15.The holder (including the microelectromechanical systems (MEMS) chips) shall not block more than 30 % line of sight to the energy dispersive X-ray spectrometer (EDS) when the holder is at 0 ° tilt.16.The holder (including the microelectromechanical systems (MEMS) chips) must be compatible with electron energy loss spectroscopy (EELS).17.Microelectromechanical systems (MEMS) chips specimen holders must be compatible with focused ion beam (FIB) sample and contact preparation. 18.MEMs chips specimen holders must have minimal electrical leakage (lower is better).19.MEMS chips specimen holders must be available with sample support films that are stable at the maximum specified temperature: holey carbon, silicon nitride, and through-hole (no film).20.The contractor shall demonstrate scientific viability of the holder through a record of publications based on using the tool.21.The contractor shall have a dedicated service team. The contractor shall respond to requests for service by phone or email within 48 hours of initial contact.

22.The contractor shall provide searchable electronic copies of manuals, troubleshooting guides, and application notes, as well as real time tracking and updates on all service requests, and online ordering of MEMS devices.23.The system shall include all hardware necessary to operate the biasing holder, including any necessary power supply(ies), electronics, and cabling.24. Compatibility with Thermo Fisher Spectra and Talos S/TEM instruments and software.25.Compliance with the TEM manufacturer’s certification for X-ray safety, mechanical tolerance and vacuum safety.

Requirements / Specifications: Sample Holder for Thermo Fisher (S)TEM 2

Separately Priced Options and Services:

Item 2: 1. Software that allows off-line analysis of images and their metadata (so processing does not have to be done on the TEM). 2.Software that actively corrects sample drift by analyzing the live data and utilizing either/both physical means (stage and beam) and digital means for the correction.

3.Software that synchronously records in-situ heating and electrical parameters and TEM parameters (such as spot size, dwell time, exposure time, etc.) directly into the metadata of each image. 4.Software that will recommend the optimal temperature ramp rate to achieve maximum resolution for a desired field of view.

Quantity: one Specifications: [Enter quantity here.]

[List any other item(s) or service requirements here using the same format above.]

Service Requirements / Item Specifications

Section Requirements/Specifications Planning Considerations ☒ N/A

☐ List any visits required to allow potential vendors to see site specific constraints.

[Explain requirement here] ☐ US citizen required on-site (foreign nationals require preregistration 30 days prior) ☐ Vendor rep on-site more than three days ☐ Other considerations: [List other considerations to ensure an accurate quote]

Shipping & Delivery ☐ N/A

[Enter shipping and delivery information by item here.] Shipping criteria:

☐ Partial delivery acceptable ☐ Direct delivery to building required [Explain need for requirement here] Building has: ☐ Loading dock ☐ Containerization preference: [Choose container] Other requirements: [List any other requirements here] Delivery date criteria:

☒ Delivery of goods, installation, and training shall be completed no later than: 2 months ARO ☐ Deliver no earlier than: [Insert date here] ☐ Other: [List other shipping, delivery, or specil requirements]

Electronic Media ☐ N/A

[Enter electronic media requirements by item here.] ☒ Software ☒ Electronic manuals ☐ License Key ☐ Other: [List any other Electronic Media requirements here] [Enter email address for software delivery here]

Installation ☐ N/A

[Enter installation requirements by item here.] ☒ Vendor technician on site ☐ Rigging ☐ Uncrating / Unpackaging ☐ Removal of packaging material

Requirements / Specifications: Sample Holder for Thermo Fisher (S)TEM 3

☐ Equipment set up ☐ Start-up services ☐ Turn key installation ☐ Other [Enter details here.] Note: When installation is required, TPOC or COR shall be the coordination point.

Facility / Utility Considerations ☒ N/A

[List site specific utility considerations by item here.] ☐ OFPM Work Order Number(s) [List any Work Order Number(s) here] ☐ Power available Voltage: [Enter voltage here] (V) Max Current: [Enter max current draw here] (A) Phase: [Choose phase] Frequency: [Typically 60Hz] (Hz) NEMA Plug Type: [Enter plug type here] (i.e.: NEMA L5-20P) Environment: [Choose environment] Other power consideration: [List any other power considerations here] ☐ Utility connections: [List relevant available utilities, pressure(s), and flowrate(s)] ☐ Other considerations: [List other considerations to ensure an accurate quote]

Warranty ☐ N/A

[Choose warranty type here.] [If you require an Extended Manufacturer Warranty, describe the warranty requirements per item here.] What is included?

☐ Labor ☐ Parts ☐ Travel ☐ Cost and liability for returns ☐ Other: [Enter other requirements here] ☒ Warranty length: one year Period Length:[Ex. month, 6 month, year etc.]

Training ☐ N/A

[Training requirements (who, when, what, where) here. If on-site, is travel necessary?] ☒ On-site ☐ Off-site location: [Enter location(s) here] ☒ Number of trainees: three Scope of training:

☒ Operations ☒ Maintenance ☒ Troubleshooting ☒ Safety ☐ Other: [Enter other scope requirements here] ☐ Training materials to be provided: [Describe requirement here] ☐ Other: [Enter other training requirements here] Note: When training is required, TPOC or COR shall be the coordination point.

Requirements / Specifications: Sample Holder for Thermo Fisher (S)TEM 4

Inspection and Acceptance ☐ N/A

[Enter inspection requirements per item here] ☒ Planned government acceptance testing (7 business days) ☐ Gov’t acceptance expected to take longer than 7 business days: [Duration] ☐ Inspection and acceptance plan: [Explain here] ☐ Other: [Enter other requirements here]

Maintenance ☒ N/A

[Describe your maintenance requirements per item here] Period Length: [Ex. month, 6 month, year etc.] What is included?

☐ Labor ☐ Spare parts ☐ Consumables ☐ Travel ☐ Cost and liability for returns ☐ Software updates ☐ Manuals ☐ Service plan length: [Enter service plan length here] ☐ Other: [Enter other requirements here] ☐ Requirements: [Enter requirements here] (i.e.: onsite within 72 hours, etc.)

☐ Option Period 1 ☐ Option Period 2 ☐ Option Period 3 ☐ Option Period 4

Government Furnished Material / Equipment ☒ N/A

[List any government furnished material/equipment per item here] ☐ Samples for testing ☐ Government equipment which will leave gov't site (TPOC must coordinate with division property custodian) Serial Number: [Enter Serial Number here] Description: [Enter description here] NIST property number: [Enter NIST property number here] ☐ Other: [Enter other details here]

Travel ☒ N/A

[Describe any travel requirements per item here] ☐ Explain: [Explain here] ☐ Other: [Enter other requirements here] Vendor: To the maximum extent possible, include as part of appropriate sections above (maintenance, repair, etc.). Otherwise, it will be a separate line item, reimbursable per the Federal Travel Regulation.

Provide suggestions to improve this form by clicking here.

https://oaam.nist.gov/OAAMAcquisitions/FormFeedback/default.aspx

Use this document to detail specifications/requirements for an item and/or ancillary services.
TITLE: Sample Holder for Thermo Fisher (S)TEM
Requesting Lab/Division/Group: PML/Applied Physics Division/ Quantitative Nanostructure Characterization Group
General Statement of Need
Provide suggestions to improve this form by clicking here.

File details come from the government source that posted it. Updated .