Amendment 2 Answers to Questions.docx

DOCX document 14 KB Posted

Attached to
4 Kelvin Closed Cycle Cryostat Federal contract opportunity
Solicitation number
NB686000-26-01896
Issued by
Department of Commerce National Institute of Standards and Technology

About this file

This file is a Questions and Answers document for Amendment 2 of a federal contract solicitation. As a template or reference document without specific substantive content detailed in the provided text, this file serves to address inquiries submitted by potential offerors regarding the solicitation NB686000-26-01896 for a 4 Kelvin Closed Cycle Cryostat. The National Institute of Standards and Technology (NIST), Acquisition Management Division (AMD), Team E-Boulder issued this amendment to clarify requirements and specifications for interested contractors. Potential offerors should reference this document alongside the Combined Synopsis/Solicitation and Requirements and Specifications documents to understand the complete scope of work and technical requirements for the equipment procurement.

The solicitation is structured as a Request for Quotation (RFQ), with all communications and submissions required to reference the solicitation number NB686000-26-01896. Amendment 1 previously modified the delivery requirement from 12 weeks to 12 months, and Amendment 2 posts answers to contractor questions without altering other terms and conditions. Contractors responding to this RFQ should review all amendments and supporting documentation to ensure accurate and compliant quotations for the 4 Kelvin Closed Cycle Cryostat equipment.

View the file

Other files for this federal contract opportunity

Other files attached to 4 Kelvin Closed Cycle Cryostat, newest first.
File Type Posted
Revised RFQ BV NB686000-26-01896.docx DOCX document
Experience Project Data Sheet.docx DOCX document
RFQ BV NB686000-26-01896.docx DOCX document
NB686000-26-01896 Requirements and Specifications.pdf PDF

On GovTribe

Work with this file on GovTribe

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Amendment 2:

Answers to Questions 4K Closed Cycle Cryostat NB686000-26-01896

What is the orientation of the magnet with respect to the optical axis?

The magnetic field must be perpendicular to the optical axis.

Is 110 mm "bore" just radial or also z?

110 mm open access from the top.

Do you require temperature sweeps with the magnet in place?

yes

Is there a rough mass range for the samples wanting to be tested?

Typical sample <100g

And when ‘upright sample space’ is given, does that mean an inverted cold head? Or that the sample must be held in a fixed, upright orientation regardless of the cryostats form factor?

Inverted cold head configuration is required.

What is the size of the split in the magnet, and are there any homogeneity requirements?

110 mm open access from the top

Do you have a range for the required temperature sweeps?

100K