This document provides questions and answers related to solicitation number N62645-22-R-0017 from the Naval Medical Readiness Logistics Command seeking high speed immunochemistry screening analyzer systems. The solicitation requests six analyzer systems with full data review capabilities and specified water supply requirements to be located at various Department of Defense drug testing laboratories. An option to acquire seven additional lower throughput systems to expand production capacity at several laboratory locations is also included. The proposed contract award date is December 28, 2022 utilizing commercial acquisition procedures. Offerors must register with the System for Award Management to participate. The contracting specialist is identified for any questions.
View the file
Other files
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
N62645-22-R-0017 Screening Analyzer Questions & Answers
Page 1 of 3 Revised 9/22/2022 V2
Item
Date Page
Section/
Paragraph Questions Answers Status
1 9/9/22
We would like to ask for an extension for Solicitation N6264522R0017.
Granted, extended thru 9/28/2022 Resolved
2 9/15/22
Missing Form -The contractor shall comply with the enclosed other contracting requirements. As part of the proposal, the contractor shall submit the enclosed Naval Medical Logistics Command (NMLC) C&A Initial Technical Questionnaire with the Blue Section completely filled out
Form added to solicitation for download\ Added as Attachment 3. Saved as:
Att 3 -Form_MDERAv6.0CyberLOG- ICS.pdf
Resolved
3 9/21/22 Na Na Will the DOD be including the Addendum to 52.512-1 – Instructions to Offerors in solicitation N6264522R0017 Screening Analyzer?
There is no Addendum for 52.512-1.
This is a typo; also, the page numbers are corrupted. 52.212-1 is on Page 34. Will repost corrected document
Resolved
4 9/21/22 2
2) System Performance
With Option 1 – Will you be keeping all your current equipment?
All current equipment will be retained while validation on the new equipment is conducted.
Resolved
5 9/21/22 2
2) System Performance – 2.8
Does this mean the actual system holds that many samples at a time?
Yes. That means the system shall hold that many samples at a time.
Resolved
6 9/21/22 2
2) System Performance – 2.10
That is only one system. Depending on how many systems are needed at each site?
Correct, some sites will only obtain one system.
Resolved
7 9/21/22 3
5. LIMS Interface Requirements – 5.6
Please clarify what is meant by LAN and batch number.
LAN and batch numbers are the identifying numerical that are provided to the samples to keep track of them.
This is for 5.9 not 5.6.
Resolved
Page 2 of 3 Revised 9/22/2022 V2
8 9/21/22 4 Enhancing Factors - 10
What is the meaning of Full Data Review?
That means that we shall have access to any storage device on the machine that writes/saves information to it. Any drive shall be fully accessible by us. This is Enhancing Factors 19 not 10.
Resolved
7 9/21/22 Does the DOD intend to utilize its current water system? And if so, can you provide specifications on the type of water system used?
Yes, prefer to use current source at labs.
Water Specs:
DI conductivity
- 2.0 uS/cm or less through a 0.5 micrometer or less filter
Pressure
- 0.49x10^5 to 3.92x10^5 Pa (7 to 57 psi)
Consumption
- Avg 124 L/hr (2 module system) and 186 L/hr (3 module system)
Max Demand
- 3.5 L/min (2 module system) or 4.0 L/min (3 module system)
Temp – between 5 - 28 degC or 41 to 83 degF (wide room temp range)
Supply facility
- located within 10m (33 ft) of outlet, minimized air bubbles
Min Supply hose
- braided hose 12mm (ID) x 18 mm (OD), L=10m (33ft), 1 piece
Resolved
Page 3 of 3 Revised 9/22/2022 V2
Water Specs Repeated:
DI conductivity - 2.0 uS/cm or less through a 0.5 micrometer or less filter Pressure - 0.49x10^5 to 3.92x10^5 Pa (7 to 57 psi) Consumption - Avg 124 L/hr (2 module 5820) and 186 L/hr (3 module 5830) Max Demand - 3.5 L/min (5820) or 4.0 L/min (5830) Temp - between 5 - 28 degC or 41 to 83 degF (pretty wide room temp range) Supply facility - located within 10m (33 ft) of outlet, minimized air bubbles Min Supply hose allowance - braided hose 12mm (ID) x 18 mm (OD), L=10m (33ft), 1 piece