SOW (1).docx
DOCX document 22 KB Posted
- Attached to
- Precision IR Calibrators
- Solicitation number
- N0042126Q1277
About this file
This is a Statement of Work (SOW) document for a federal procurement opportunity issued by the Department of the Navy Naval Air Systems Command Naval Air Warfare Center under Solicitation Number N0042126Q1277. The Government is seeking to procure Fluke Precision IR Calibrators on a firm fixed-price basis under Federal Acquisition Regulation (FAR) Part 12 authority, with awards to be made on a brand name basis. All respondents must be authorized distributors of the specified brand and registered in the System for Award Management (SAM) Database prior to offer submission.
Key requirements for offerors include: submission of brand name quotes with technical information sufficient for evaluation by the deadline of 31 July 2026 at 5:00 PM Eastern Standard Time; items must be brand new and exact to specifications, covered under manufacturer's warranty, with no used or refurbished items accepted; TAA compliance is mandatory; and award will be made on an all-or-none basis. Quotes must include FOB shipping cost, DUNS number, CAGE code, estimated delivery date, company point of contact information, phone, email, and total price. Offerors must complete the required certification (52.209-11, Representation by Corporations Regarding Delinquent Tax Obligations). All quotes and responses should be submitted via email to Michael Coffey (michael.j.coffey52.civ@us.navy.mil). The solicitation closing date is firm, and proposals received after the deadline will not be accepted.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| Solicitation - N0042126Q1277.pdf |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Statement of Work A. Background The use of NAVAIR METCAL equipment funding is to procure Calibration Standards (CALSTDs) and ancillary equipment for Aviation Fleet Intermediate Calibration Activities, Fleet Training Activities and Aviation Navy Calibration Laboratories (NCLs) to support Aviation Organizational and Intermediate (I) level maintenance assets. All Cal Standards listed for procurement must first go through an approval process. This process includes: initial submission by the fleet; review and prioritization by the Fleet TYCOM’s; laboratory workload justification and standardization review by a NAVAIR METCAL technical review team; overall procurement strategy review to assess overall program objectives, priorities and funding.
REQUIREMENTS:
1. Mfr. one year warranty with sticker affixed to each unit plus a warranty certificate.
2. Calibration Certificate and calibration sticker (as applicable for each unit)
3. Units shall be serialized.
4. Units (over $5,000) shall have an IUID Label applied per DFAR 252.211-7003.
5. Manufacturer’s Standard Operating and servicing manuals (CD version acceptable) shall be provided with each unit. The DFARS clause listed below must be referenced in all contracts and the manufacturer made aware that the use/reproduction of tech data/manuals (for government use only) will follow these use guidelines.
DFARS SUBPART 227.71 - Rights in Technical Data
7. The US government shall have limited technical data rights for all technical manuals and documents furnished with the procured commercial items, for the purpose of Government use only in duplicating or posting technical documentation to limited access web sites for use by US Government Personnel and Contractors engaged in maintenance, calibration or related logistics activities associated with the procured commercial items in accordance with DFARS 252.227-7015 Technical Data--Commercial Items B. Tasking:
The contractor shall provide the following:
1) P/N 4180 Precision IR Calibrator Qty 3 To include the following:
1. Temperature range (@ 23°C ambient, 0.95 emissivity): –15°C to 120°C
2. Display accuracy:
i. ±0.40°C at –15°C
ii. ±0.40°C at 0°C
iii. ±0.50°C at 50°C
iv. ±0.50°C at 100°C
v. ±0.55°C at 120°C
3. Stability:
i. ±0.10°C at –15°C
ii. ±0.05°C at 0°C
iii. ±0.10°C at 120°C
4. Uniformity (5.0 in dia of center of target):
i. ±0.15°C at –15°C
ii. ±0.10°C at 0°C
iii. ±0.25°C at 120°C
5. Uniformity (2.0 in dia of center of target):
i. ±0.10°C at -15°C
ii. ±0.10°C at 0°C
iii. ±0.20°C at 120°C
6. Heating time:
i. 15 min: –15°C to 120°C
ii. 14 min: 23°C to 120°C
7. Cooling time:
i. 15 min: 120°C to 23°C
ii. 20 min: 23°C to –15°C
8. Stabilization time: 10 minutes
9. Nominal emissivity: 0.95
10. Thermometer emissivity compensation: 0.9 to 1.0
11. Target diameter: 152.4 mm (6 in)
12. Computer interface: RS-232
13. Power:
i. 115 V AC (±10%), 6.3 A,
ii. 50/60 Hz, 630 W
iii. 230 V AC (±10%), 3.15 A,
iv. 50/60 Hz, 630 W
14. Fuse(s):
i. 115 V AC 6.3 A, 250 V, slow blow
ii. 230 V AC 3.15 A, 250 V, T
15. Size (H x W x D): 356 x 241 x 216 mm (14 x 9.5 x 8.5 in)
16. IUID Label
2) P/N 4181 Precision IR Calibrator Qty 2 To include the following:
1. Temperature range (@ 23°C ambient, 0.95 emissivity): 35°C to 500°C
2. Display accuracy:
i. ±0.35°C at 35°C
ii. ±0.50°C at 100°C
iii. ±0.70°C at 200°C
iv. ±1.20°C at 350°C
v. ±1.60°C at 500°C
3. Stability:
i. ±0.05°C at 35°C
ii. ±0.20°C at 200°C
iii. ±0.40°C at 500°C
4. Uniformity (5.0 in dia of center of target):
i. ±0.10°C at 35°C
ii. ±0.50°C at 250°C
iii. ±1.00°C at 500°C
5. Uniformity (2.0 in dia of center of target):
i. ±0.10°C at 35°C
ii. ±0.25°C at 250°C
iii. ±0.50°C at 500°C
6. Heating time:
i. 45 min: 35⁰ C to 500⁰ C
7. Cooling time:
i. 100 min: 500°C to 35°C
ii. 40 min: 500°C to 100°C
8. Stabilization time: 10 minutes
9. Nominal emissivity: 0.95
10. Thermometer emissivity compensation: 0.9 to 1.0
11. Target diameter: 152.4 mm (6 in)
12. Computer interface: RS-232
13. Power:
i. 115 V AC (±10%), 6.3 A,
ii. 50/60 Hz, 630 W
iii. 230 V AC (±10%), 3.15 A,
iv. 50/60 Hz, 630 W
14. Fuse(s):
i. 115 V AC 6.3 A, 250 V, slow blow
ii. 230 V AC 3.15 A, 250 V, T
15. Size (H x W x D): 356 x 241 x 216 mm (14 x 9.5 x 8.5 in)
16. IUID Label
C. Deliverables Delivery of the equipment shall be provided at the following address:
COMMANDER
NAVAIR Calibration Standards MCAS, Attn: David Swiger (UIC N48535)BLDG 612, Bay 9 Beaufort, SC 29904-5002
D. Delivery Date
Delivery to be no later than 4 weeks after award of contract
***PARTICAL SHIPMENT ACCECPTED***
F. Invoicing
All invoicing shall be submitting into Wide Area Workflow for payment after ALL material is received in the condition as stated in section B.
VENDOR – IF ELECTRONIC EQUIPMENT, PLEASE MARK BOX/PACKAGE CONTAINING ELECTRONIC EQUIPMENT WITH THE “DO NOT SCAN” STICKER