Questions from Solicitation Notice ID N0025921Q0014

Received: 12/16/2020

In reference to the stated notice id, can you answer a few questions for us?

1. How many pipettes are single channel versus multi-channel pipettes?

Answer: All are Single Channel

a. Are all channels for multis requiring calibration?

b. Answer: N/A – No multi-channel

2. How many volumes are required?

Answer: Dependent on single/multi-volume (no more than 3 volumes) pipette (majority are single volume)

3. How many readings are required?

Answer: Minimum of 4 readings per volume

4. Is the service requiring ISO/IEC certifications?

Answer: No, must be ISO/IEC and or ANSI/NCSL Z540.3 compliant at a minimum