4.03 N00244-25-Q-S078 Q&A.docx

DOCX document 16 KB Posted

Attached to
Closed Cycle Cryostat for Microscope
Solicitation number
N00244-25-Q-S078
Issued by

About this file

This document is a Questions and Answers (Q&A) file for Request for Quote (RFQ) N00244-25-Q-S078, clarifying technical specifications for a closed-cycle cryostat. The Q&A confirms that the requirement is specifically for a cryostat with precise technical requirements, including an optical microscopy interface with a direct viewing optical window compatible with a Leica DM2500 microscope. The cryostat must accommodate a 24-pin dual inline pin package and include an electrical feedthrough with at least 24 pre-installed electrical wires for sample characterization, enabling thermal and electrical analysis of nanomaterials across a wide temperature range for the Naval Postgraduate School's Mechanical and Aerospace Engineering Department.

View the file

Other files

File Type Posted
4.03 N00244-25-Q-S078 Q&A (II).docx DOCX document
4.02 SF 1449 - N0024425QS078 - updated.pdf PDF
4.02 SF 1449 - N0024425QS078.pdf PDF

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

RFQ N00244-25-Q-S078 Q&A

1. The solicitation states it is for a Closed Cycle Cryostat for Microscope however the SF1449 section “3.0 Minimal Technical Requirements/Salient Characteristics” includes a mix of specifications for a cryostat and a microscope. Is the requirement for both a microscope and cryostat or just a cryostat? And, if for just the cryostat, may we please have specifications for the cryostat only?

Answer: The requirement is just for a cryostat.

· Must have optical microscopy interface (i.e., sample holder beneath optical window). The cryostat must include an optical window that allows direct viewing of the sample through an optical microscope (e.g., Leica DM2500, part of an existing Raman microscopy system). The sample holder must be positioned beneath this window and within the working distance of the microscope objective to enable optical access.

· Must be able to accommodate 24 pin dual inline pin package & Must have at least 24 electrical wires for sample characterization.

· The cryostat must include an electrical feedthrough with at least 24 pre-installed electrical wires for sample characterization. Additionally, the sample holder must be compatible with a standard 24-pin dual in-line package (DIP) and capable of securely accommodating and thermally anchoring it to the cryostat’s sample stage.