Vacuum Furnace Requirements - Minimum Specifications REVISED.pdf
PDF 89 KB Posted
- Attached to
- Vacuum Muffle Furnace Federal contract opportunity
- Solicitation number
- N0017326RFIGF02
About this file
This document outlines technical requirements for a vacuum furnace, specifying detailed specifications for a laboratory equipment purchase. The vacuum furnace must have 208V or 220V power, accommodate a 12-inch cubic microfluidics board, and reach a maximum temperature of 1200 Celsius. Additional critical requirements include a digital vacuum gauge, compact dimensions of 30 × 30 × 36 inches or less, capability for both inert gas and vacuum operations, and the ability to reach a vacuum of 5E-4 torr or lower. The furnace must also include a compatible vacuum pump with a pumping speed of 300L/min or greater, ensuring compatibility with existing laboratory infrastructure and meeting precise scientific research needs.
View the file
Other files for this federal contract opportunity
| File | Type | Posted |
|---|---|---|
| Vacuum Furnace Requirements - Minimum Specifications.pdf | ||
| RFI_Sources Sought Instructions - Vacuum Muffle Furnace.pdf |
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Vacuum Furnace Requirements:
• 208V or 220V power to interface with existing electrical connections
• Interior capable of holding a 12 inch x 12 inch x 12 inch assembly to accommodate largest considered microfluidics board.
• Max Temp 1200 Celsius (1473K) or greater.
• Digital Vacuum gauge included
• Dimensions: 30 inch × 30 inch × 36 inch or less to fit in existing lab space
• Capability for both inert gas and vacuum operation.
• In Vacuum mode, capable of reaching 5E-4 torr or lower.
• A compatible vacuum pump, with pumping speed of 300L/min or greater.
File details come from the government source that posted it. Updated .