Open-Path NH3 analyzer

Required Minimum Specifications:

• The analyzer must be a commercially available, single-unit system capable of measuring NH3 concentration, with a lower detection limit of 8 ppm-m and upper detection of 5,000 ppm-m with a 1% reading accuracy and resolution of 0.2%.

• Path length of coverage range from 5-500 m.

• Remote diagnostics and data transmission via wireless communications.

• Dual thermoelectric cooler controller enables a laser temperature stability to +/-0.001C.

• Practically no temperature related drift from -40 to 50C.

• Dynamic Real-time Pressure and Temperature Compensation via Internal Pressure &

Temperature sensors

• Store up to 20 years of data

• Touchscreen interface with graphical data displays, including spectral waveforms.

• NH3 TDL instrument must be able to measure NH3 concentration at a minimum of 8 ppm-m with an upper boundary of 5,000 ppm-m.

• Open-Path NH3 TDL instrument path length of operation should be between 5 to 500 m.

• NH3 TDL instrument must be able to store data for months at a time.

• NH3 TDL instrument must be able to have remote diagnostics and data transmission via wireless communications

• NH3 TDL instrument must be able to compensate for pressure and temperature from internal pressure and temperature sensors.

• NH3 TDL instrument must include Onboard Software (no computer required)

• NH3 TDL instrument must have little to no temperature drift from -40 to 50C.

• NH3 TDL instrument must have a touchscreen interface with graphical data displays, including spectral waveforms.

• NH3 TDL instrument must be compatible with existing Boral Lasers currently in use laboratory (sharing mirror and tripods).

• Tripod for unit, compatible with GasFinder3-OP and the X-Y Aiming Mount.