Attachment 1 - Statement of Work.pdf

PDF 54 KB Posted

Attached to
High Precision Teslameter with Calibrated 3-Axis Hall Probe
Solicitation number
MB479982
Issued by

About this file

This is a Statement of Work for the procurement of a high precision Teslameter with a calibrated 3-axis Hall probe, classified as a Medium Procurement. The equipment is intended for measurement of the magnetic field in the Wien Filter spin rotator at Brookhaven National Laboratory (BNL), with requirements for high precision and reliability in dipole field measurements.

The Teslameter system must meet specific technical parameters: measurement range of ±100 mT, DC resolution of 0.5 µT (RMS), sampling rate exceeding 5 SPS, DC accuracy of 100 ppm full scale, temperature stability less than 20 ppm/°C, and frequency bandwidth of DC–2.5 kHz. The 3-axis Hall probe must be calibrated with mutual orthogonality between axes of less than 0.1°, and the system must include USB/Ethernet digital interfaces for data logging. The contractor shall deliver the complete Teslameter system with the calibrated 3-axis Hall probe, a calibration certificate, and operating documentation within 10–12 weeks, with delivery to ARO (Atmospheric and Radiation Organization). All work shall be performed at the contractor's facility with no onsite support, installation, commissioning, or training required at BNL. Acceptance is based on review of the calibration certificate and confirmation that the delivered equipment meets all specified requirements. No special intellectual property or data rights requirements are identified.

View the file

Other files

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

STATEMENT OF WORK

High Precision Teslameter with Calibrated 3-Axis Hall Probe

Procurement Category: Medium Procurement

Scope Procurement of a high precision Teslameter with a calibrated 3-axis Hall probe for measurement of the magnetic field in the Wien Filter spin rotator. High precision and reliability are required for dipole field measurements.

Technical Requirements Parameter Requirement

Measurement Range ±100 mT

DC Resolution (RMS) 0.5 µT

Sampling Rate >5 SPS

DC Accuracy 100 ppm FS

Temperature Stability <20 ppm/°C

Frequency Bandwidth DC–2.5 kHz

Hall Probe Calibrated 3-Axis

Mutual orthogonality between axes

Digital interface

Deliverables / Milestones Deliverable Description Due Approval

Teslameter System

Teslameter and calibrated

10–12 weeks

ARO

Yes

<0.1°

USB/Ethernet for data logging

3-axis Hall probe

Calibration Certificate

Calibration documentation

With delivery

Yes

Documentation Operating documentation

With delivery

Yes

Deliverable Description Due Approval

Acceptance Criteria Acceptance shall be based on review of the calibration certificate and confirmation that the delivered equipment meets the specified requirements.

Work Location Work shall be performed at the contractor facility. No onsite support, installation, commissioning, or training at BNL is required.

IP and Data Rights N/A. No special intellectual property or data rights requirements are identified.

Scope