This is a Memorandum for Record from the Department of the Air Force's Oklahoma City Air Logistics Complex (AFMC) at Tinker Air Force Base, dated 18 September 2025, detailing a request for equipment procurement by the 76th MXSG/MXSS unit. The procurement includes one GelSight Mobile Series GS Max device (part number GSS-MB2-00041) with multiple accompanying components such as a calibration target kit, cleaning kit, hard case, USB cable, and fourteen GelSight GS Max Gel Cartridges.
The equipment request also encompasses a PRO-Perpetual Software License (part number GS-SFT-PER-PRO) with GS Mobile v3.8 software and twelve months of maintenance, one Microsoft Surface Pro Package with Surface Pro Tablet, Windows 11 PRO, keyboard cover, and rugged case, and additional training components including a two-hour virtual training session for 3-5 users and an eight-hour onsite training for up to three people. All calibration equipment is required to conform to ISO 17025 standards, with specific detailed requirements for calibration plates, target kits, and validation processes.
View the file
Other files for this federal contract opportunity
| File | Type | Posted |
|---|---|---|
| M25-024 GelSight Max Mobile - CSSFA812625Q0063Rev5_092225.pdf | ||
| M25-024 GelSight Max Mobile - BNJ_Redacted.pdf |
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
e. Quantity of one (1) Part No: RSP-VC002 Validation Plate with Certificate of Calibration.
(1) Shall Conform to ISO 17025 Standards
(2) One (1) Calibration Target Kit for system-level calibration for each time a new gel cartridge is installed. The Calibration Target Kit shall include:
(a) G-250 calibration plate with Certificate of Calibration for X, Y, and Z dimensions
(b) Ball Grid Array (BGA) D5 S10 (ball diameter: 0.5mm, ball diameter tolerance:
±0.1mm, ball material: chrome steel)
(c) GSS-CAL-0019 (glass calibration artifact)
(3) One (1) GS Max Validation Plate with Certificate for verifying if the system is still operating within specified tolerances.
f. One (1) Virtual Training via Teams, for two (2) hours for 3-5 users. (Part No: GS-TRN-1001)
g. Onsite Training for up to 3 people. Training duration for eight (8) hours (Part No: GS-
TRN-1002)