M25-008 Bruker Spectrometer - Item Description.pdf

PDF 181 KB Posted

Attached to
Bruker Spectrometer Federal contract opportunity
Solicitation number
FA812625Q0021
Issued by
Department of the Air Force Materiel Command Air Force Sustainment Center

About this file

This document is a Memorandum for Record from the Department of the Air Force Headquarters Air Force Sustainment Center (AFMC) at Tinker Air Force Base, Oklahoma, detailing the specifications for a Bruker Spectrometer procurement. The 76th MXSG/MXSS/MXDTAA is requesting funds to purchase an INVENIO-S Fourier Transform Infrared Spectrometer with specific technical requirements, including a laser device for cell alignment, a deuterated triglycine sulfate (DTGS) detector, spectral resolution better than 0.4 cm⁻¹, and a mid-infrared KBr beamsplitter with a spectral range of 8,000-380 cm⁻¹.

The procurement includes one Bruker INVENIO-S Spectrometer, one Bruker Data System (CS81/25+), one Bruker OPUS Spectroscopy Software (O/IR9+), on-site installation and performance verification, and 8 hours of on-base training for three personnel. The equipment comes with a 1-year onsite warranty covering all parts and labor, and includes specialized maintenance tools and a user manual. The document is dated 6 January 2025 and provides a detailed itemization of the required scientific instrument and associated components.

View the file

Other files for this federal contract opportunity

Other files attached to Bruker Spectrometer, newest first.
File Type Posted
Solicitation - FA812625Q0021.pdf PDF
M25-008 Bruker Spectrometer - SOW_Redacted.pdf PDF
Solicitation - FA812625Q0021.pdf PDF
Appendix C-Redacted.pdf PDF

On GovTribe

Work with this file on GovTribe

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

DEPARTMENT OF THE AIR FORCE

HEADQUARTERS AIR FORCE SUSTAINMENT CENTER (AFMC)

TINKER AIR FORCE BASE OKLAHOMA

6 January 2025

MEMORANDUM FOR RECORD

SUBJECT: Item Description, M25-008 Bruker Spectrometer

1. The 76th MXSG/MXSS/MXDTAA is requesting utilization of funds for the purchase of an INVENIO-S Fourier Transform Infrared Spectrometer.

2. The requested equipment item descriptions/specifications are outlined below.

• One (1) each, Brucker, INVENIO-S Fourier Transform Infrared Spectrometer, with the following specifications:

Laser device for cell alignment and path length verification Capable of transferring current laboratory calibration spectra seamlessly between software generations Fitted with a deuterated triglycine sulfate (DTGS) detector Achieves spectral resolution better than 0.4 cm⁻¹ Optical system utilizing a mid-infrared KBr beamsplitter, offering a spectral range limited to 8,000-380 cm⁻¹ Fixed interferometer that maintains alignment without requiring software or manual adjustment Electronically stabilized infrared source Provides a wavenumber accuracy of 0.01 cm⁻¹ at 2000 cm⁻¹ or higher 1-year onsite warranty covering all parts and labor Specialized tools required for routine maintenance User Manual

• One (1) each, Bruker, Data System, CS81/25+

• One (1) each, Bruker, OPUS Spectroscopy Software, O/IR9+

• On-site Installation and Performance Verification

• On-Base Training 8 hours for three (3) persons.

File details come from the government source that posted it. Updated .