IGCS_chamber_CAD_drawing.pdf

PDF 164 KB Posted

Attached to
Ion Gauge Calibration Service Heater Enclosure
Solicitation number
1333ND25QNB680246
Issued by

About this file

This document is a technical CAD drawing for a heater chamber assembly, likely for an ion gauge calibration service at the National Institute of Standards and Technology (NIST). The detailed technical drawing specifies precise design requirements for a hexagonal-shaped enclosure mounted on a table, including critical specifications such as:

Key design constraints include: a 51.00 mm height, 22.00 mm radius, and 100-degree configuration; a red-shaded "keep-out volume" where no components can protrude; requirement for internal 2-inch maranite insulation with unblocked table holes; mandatory placement of pneumatic valve actuators outside the heater enclosure; and provisions for routine access by removing or opening side panels without full enclosure disassembly. The drawing also specifies that port holes for valve actuators must have removable insulated covers, and the enclosure must be installable with pneumatic valves already attached to the chamber.

View the file

Other files

File Type Posted
Responses to questions.pdf PDF
SOW.pdf PDF
1333ND25QNB680246 CSS.pdf PDF

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

1.

Table has internal 2 in maranite insulation.

Holes in table must not be blocked.

Actuator section of pneumatic valves must be outside heater enclosure.

R22.00

17.00

100° R28.13

Red shaded region indicates minimum keep-out volume.

Heater enclosure and frame must rest on table.

Routine access without enclosure disassembly by removing or opening side panels is required.

Potential frame and/or enclosure shape (other shapes are acceptable, see requirements).

Port holes to accommodate valve actuators must have removable insulated covers.

Enclosure components cannot protrude into the keep-out volume.

Enclosure must be installable with pneumatic valves attached to chamber and pneumatics must be removable without full disassembly of enclosure.

A A

B B

C C

D D

DRAWN

CHK'D

APPV'D

MFG

Q.A

UNLESS OTHERWISE SPECIFIED:

DIMENSIONS ARE IN MILLIMETERS

SURFACE FINISH:

TOLERANCES:

LINEAR:

ANGULAR:

FINISH: DEBURR AND

BREAK SHARP

EDGES

NAME SIGNATURE DATE

MATERIAL:

DO NOT SCALE DRAWING REVISION

TITLE:

DWG NO.

SCALE:1:50 SHEET 1 OF 1

A4

WEIGHT:

HV-chamber-assembly-on-table

Sheet1
Drawing View4
Drawing View5
Drawing View6