This Statement of Work (SOW) details NIST's requirements for an Ion Gauge Calibration Service Heater Enclosure for the Fundamental Thermodynamics Group. The document specifies technical requirements for a new vacuum bake-out enclosure to upgrade the existing ionization gauge calibration service (IGCS), with a focus on increasing automation and calibration turnaround time.
Key requirements include a heater enclosure design with 4-8 removable side panels, capable of withstanding temperatures up to 200°C, and featuring heating elements, pneumatic valve ports, and a proportional-integral-differential (PID) controller. The enclosure must rest on an IGCS support table, accommodate pneumatic valves, and operate at temperatures up to 135°C. Delivery is specified for 12 weeks after order, with the enclosure to be delivered to NIST in Gaithersburg, MD. The government will conduct a 30-day inspection and acceptance testing period, with payment to be made 100% after installation and acceptance by the NIST Technical Point of Contact (TPOC).