This document appears to be the necropsy table specifications for the Uniformed Services University of the Health Sciences (USU) vivarium. The key requirements are for a stainless-steel downdraft necropsy table with a minimum work surface of 60x25 inches, a self-contained ventilation system, a sloped pan with a built-in water flush, removable stainless-steel work panels, dual GFCI outlets, and the ability to connect to the existing room's water, electrical, and drainage systems. The table must be stationary, with a working height between 37-42 inches, and able to filter out formalin fumes. The related federal contract opportunity, Solicitation Number HU000124Q0016, is for the procurement of this necropsy table.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| SF-30 Amendment 2.pdf | ||
| SF-30 Amendment 1.pdf | ||
| DLAR Necropsy Table Specifications Rev 1.docx | DOCX document | |
| Provisions and Clauses.pdf | ||
| Buy American Certificate (Feb 2024).pdf | ||
| RFQ HU000124Q0016 Necropsy.pdf |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
DLAR Necropsy Table
Specifications of the necropsy table for USU vivarium are as follows:
· We need a stainless-steel downdraft necropsy table that provides at least 60x25 inches of ventilated work space with the following features:
· Self-contained ventilation system that filters out odor, vapors, and chemical fumes down away from the user appropriate for BSL-1 and BSL-2 animal necropsies.
· Sloped pan with built-in water pan flush system and hand spray attachment with hose.
· Removable stainless-steel work space panels for cleaning.
· Dual GFCI outlets on table.
· Total table dimensions must not exceed 90x35 inches.
· Must be able to connect to existing water, electrical, and drainage systems in room.
The main utility requirements are:
· the table must connect to a 1/2" flex connection water line
· drainage that is able to function with our existing room floor drain
· be able to either plug into a wall outlet or be hard-wired for an electrical connection onto one 20-amp circuit
· table's exhaust filtration system must be able to filter out formalin fumes
· Must not be portable (i.e. not on wheels)
· Stationary table with working height between 37” to 42”