About this file

The document provided is the DLAR Necropsy Table Specifications for a federal contract opportunity. It outlines the technical requirements for a stainless-steel downdraft necropsy table with a minimum 60x25 inch ventilated work space. Key features include a self-contained ventilation system, sloped pan with water flush, removable work surface panels, dual GFCI outlets, and the ability to connect to existing water, electrical, and drainage systems. The table must be stationary, non-portable, and have a height range of 37-42 inches. The related federal contract opportunity is Solicitation Number HU000124Q0016, issued by the DOD Uniformed Services University of the Health Sciences, for a Necropsy Table. The solicitation requests that offerors refer to the attached documentation, which contains the detailed technical specifications provided in this document.

View the file

Other files for this federal contract opportunity

Other files attached to Necropsy Table, newest first.
File Type Posted
SF-30 Amendment 2.pdf PDF
DLAR Necropsy Table Specifications Rev 1.docx DOCX document
SF-30 Amendment 1.pdf PDF
Buy American Certificate (Feb 2024).pdf PDF
RFQ HU000124Q0016 Necropsy.pdf PDF
Provisions and Clauses.pdf PDF

On GovTribe

Work with this file on GovTribe

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

DLAR Necropsy Table

Specifications of the necropsy table for USU vivarium are as follows:

· We need a stainless-steel downdraft necropsy table that provides at least 60x25 inches of ventilated work space with the following features:

· Self-contained ventilation system that filters out odor, vapors, and chemical fumes down away from the user appropriate for BSL-1 and BSL-2 animal necropsies.

· Sloped pan with built-in water pan flush system and hand spray attachment with hose.

· Removable stainless-steel work space panels for cleaning.

· Dual GFCI outlets on table.

· Total table dimensions must not exceed 90x35 inches.

· Must be able to connect to existing water, electrical, and drainage systems in room.

The main utility requirements are:

· the table must connect to a 1/2" flex connection water line

· drainage that is able to function with our existing room floor drain

· be able to either plug into a wall outlet or be hard-wired for an electrical connection onto one 20-amp circuit

· table's exhaust filtration system must be able to filter out formalin fumes

· Must not be portable (i.e. not on wheels)

· Stationary table with height ranges from 37” to 42”