Coagulation SOW - Enclosure A Site Specific information Amendment 0003.xlsx

XLSX spreadsheet 18 KB Posted

Attached to
FY26 HCD-W Coagulation Analyzer CPRR Consolidation Effort
Solicitation number
HT941026Q2004
Issued by

About this file

This file is a site-specific information template for Madigan Army Medical Center, located at Joint Base Lewis-McChord, Washington, in the Indo-Pacific region. The document collects facility requirements and operational details to support the procurement of coagulation analyzers under a Cost Per Reportable Result (CPRR) contract. The anticipated start date for equipment delivery is either October 1, 2026, or October 1, 2027, contingent upon the MTF's decision to exercise Option 2, with the specific analyzer quantity pending MTF request.

The site-specific requirements captured include equipment specifications (maximum allowable dimensions of 4 feet by 3 feet by 6 feet), power requirements (100-240V at 50-60Hz with maximum power consumption of 65W, or 95V/115V/230V at 50-60Hz), and operational details regarding coagulation testing panels, quality control frequencies, batch testing procedures, and heparin monitoring protocols. The template also addresses facility constraints, including whether the MTF has adequate space to accommodate new equipment while maintaining existing equipment, laboratory hours of operation, access to reagent-grade rinse water, and whether the site requires onsite vendor support for lot-to-lot assistance and regression reports—a requirement mandated for Naval MCSD facilities. This information collection ensures that the contracted coagulation analyzer solution aligns with Madigan Army Medical Center's specific operational, space, power, and support requirements.

View the file

Other files

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Site Specific Information

MTFNetworkAnticipated Start Date of DeliveriesAnalyzer QuantityMaximum Space Dimensions (Height x Width x Depth) for Analyzer EquipmentPower Requirement (what is the maximum power available)Any site-specific requirements or additional equipment required?Does MTF have room for new equipment while old equipment still there?AddressLaboratory Hours of OperationDIC Panel Question 1: What individual tests do you anticipate making up this panel?DIC Panel Question 2: Are these individual tests included in the other test count totals provided or to be billed a s a CPRR separately?Does your labs already have access to reagent grade rinse water? Y / NOnsite vendor supported “lot to lot” assistance and regression reports are required for Naval MCSD. Does your site also require onsite vendor support for Lot-to-Lot assistance and reports? Y / NWhich tests are performed in batches, if any?If tests are performed in batches, how frequently is the QC performed?What frequency and number of levels of QC are performed on Routine tests? (Example: 3 levels once per day, 2 levels every 8 hours, or Only when ordered)Please list what tests are included in a "DIC Panel".Which MTFs, if any, use the “Activated Partial Thromboplastin Time (APTT)” test to monitor heparin levels in patients? (Does you MTF use APTT to monitor heparin levels in patients?)
Madigan Army Medical Center, Joint Base Lewis-McChord, WAIndo PacificEither 10/1/2026 or 10/1/2027* TBD - Depends on MTF requesting to exercise Option2Site can accommodate a large space - maximum allowable dimensions: 4’ x 3' x 6'100-240V with 50-60Hz and max power consumption 65W and 95V, 115V, or 230V at 50/60Hz and max power <110W (1400VA)N/ANo for one analyzer; yes for one analyzerDepartment of Pathology / Core Laboratory

9040 Jackson Ave.

Tacoma, WA 9843124 hours a day, 7 days a week (holidays included)PT, aPPT, Fibrinogen, D-DimerDIC are billed as a panel; further testing will be billed seperatelyY (currently through NERL water through Omnicell)NoAll Factors assays (V, VII, VIII, IX, X and XI) and Lupus Anticoagulant (LAC)Weekly2 levels every 8 hoursPT, PTT, FIB and D-DimerWe still use PTT for Heparin Therapy but as of October 2024, DHA has added Anti-Xa testing to our Heparin Therapy Power Plans.
Naval Hospital Camp Pendleton, CAPacific Rim2/1/272H 80" x W 50" x L 125"120V, 60 Hz.N/AYES, depending on the size of the new equipment.200 Mercy Circle
Oceanside, CA 9205524 hours a day, 7 days a week (holidays included)PT, aPPT, Fibrinogen, D-DimerYes, all the test are included in the count provided.Y (currently through NERL water through DMLSS)YesN/AN/A2 levels every 8 hoursPT, PTT, FIB and D-DimerNo
Naval Hospital Bremerton, WAPacific Rim10/1/26120.5" x 44.5" x 19.5"120V, 60 Hz.Must NOT exceed defined maxiumum space dimensions for site. Shall NOT require an external water source or drains for operation. Must be Fully automated benchtop. Shall be capable of clot detection, automatically dilution, processing plasma patient samples, and positively identifying patient samples using bar codes.Yes, depending on the size of the new equipment.Laboratory Department

1 Boone Road Code O5L, Building 1

Bremerton, WA 98312-18980700 to 1600, 7 days a week (holidays included)N/AYes, all the test are included in the count provided.Yes (currently through NERL water through Omnicell)NoN/AN/A2 levels every 8 hoursN/AYes
Robert E. Bush Naval Hospital Twentynine Palms, CAPacific Rim10/1/262H 40" x W 96" x L 30"120V, 60 Hz.Does not require an external water source. See max dimension column.Yes, depending on the size of the new equipment.Laboratory Department

1145 Sturgis Rd., Bldg. 1343

Twentynine Palms, CA 9227824 hours a day, 7 days a week (holidays included)PT, aPPT, Fibrinogen, D-DimerWe do not have a DIC panelYes (currently through DMLSS or GPC)NoN/AN/A2 levels every 8 hours for PT, PTT. As needed QC for D-Dimer and Fibrinogen due to low volume.PT, PTT, FIB and D-DimerNo
99th Medical Group, Mike O'Callaghan Military Medical Center, Nellis AFB, NVContinental11/17/2026 OR 11/17/2027 * TBD - Depends on MTF requesting to exercise Option2Site cannot accommodate a large space: There is not extra space available at this time - H 27" x W 31" x D 35"; (one analyzer is countertop)120 V/ 60 HzConnection compatibility with the Aptio LineNoCore Lab

0079 MMDCN Lab 4420 N. Grissom Ave.

Nellis AFB, NV 8919124 hours a day, 7 days a week (holidays included)PT, aPPT, Fibrinogen, D-DimerYes, all the test are included in the count provided.Yes. They are bought via ECAT. Preferably, reagent grade water be included in this contract along with the required consumables.On site support is preferred at first. Can be as needed remote support after.N/AN/A2 levels every 8 hours for PT and PTT. As needed QC for D-Dimer, Fibrinogen, Heparin due to low volume.N/ANo
60th Medical Group, David Grant Medical Center, Travis AFB, CACentral9/1/272130" x70" x 70"The facility is able to supply the maximum power requirements.NoBased on the laboratory layout with all laboratory equipment and when the other contracts are awarded. If multiple contracts are awarded at once the laboratory lacks sufficient space to old and new equipment.Core Lab

101 Bodin Circle, Bldg. 777

Travis AFB, CA 9453524 hours a day, 7 days a week (holidays included)PT, aPPT, Fibrinogen, D-DimerYes, all the test are included in the count provided.Y (currently through NERL water through DMLSS)YesN/AN/A2 levels every 8 hours for PT, PTT. As needed QC for D-Dimer and Fibrinogen due to low volume.PT, PTT, FIB and D-DimerNo
Weed Army Community Hospital, Fort Irwin, CAWest12/30/261L 130" X W 31 ½"110VN/AYesAttention: LAB

390 N Loop Rd.

Fort Irwin, CA 923100730-1630 for phlebotomy and 24hrs for inpatient/ER careN/AN/AY (currently through NERL water through DMLSS)YesN/AN/A2 levels every 8 hoursPT, PTT, FIB and D-DimerNo
Naval Medical Center San Diego, CA
- Core LabPacific Rim10/1/2026 OR 10/1/2027* TBD - Depends on MTF requesting to exercise Option2L 72" x W 36" x H 48"The facility is able to supply the maximum power requirements.Shall NOT require an external water source or drains for operation. Must be Fully automated benchtop. Shall be capable of clot detection, automatic dilution, processing plasma patient samples, positively identifying patient samples using bar codes, providing onboard QC data base for current and historical lots, provide onboard historical patient database of results to include patient demographics, lot info, etc. (for duration of 5 yr contract), and provide lot to lot calibration support with inclusion of final report with easy-to-read data regression and acceptability interpretation.YesNMCSD Core Lab

34800 Bob Wilson Dr.

San Diego, CA 9213424 hours a day, 7 days a week (holidays included)PT, aPPT, Fibrinogen, D-DimerYes, all the test are included in the count provided.YesYesN/AN/A2 levels every 8 hoursPT, PTT, FIB and D-DimerYes
Tripler Army Medical Center, Fort Shafter, HIIndo Pacific10/1/2026 OR 10/1/2027* TBD - Depends on MTF requesting to exercise Option2Site can accommodate a large space - maximum allowable dimensions: H 3' x W 12' x D 2'8"95V, 115V, or 230V at 50/60Hz and max power <110W (1400VA)NoYesDepartment of Pathology

1 Jarrett White Road

Tripler AMC, HI 968590730-1630N/AN/AYesWe do not require on site assistance, but do require assistance with calculating geo mean and normal ranges.Prior to inital testing of patient samples and if exceeds 8 hours of the last acceptable QC results.Prior to inital testing of patient samples and if exceeds 8 hours of the last acceptable QC results.2 levels every 8 hoursN/ANo
56th Medical Group, Luke AFB, AZCentral9/11/271L 72 x W 36" x H48"
Luke does not have a space restriction.120V, 60 Hz.NoYes7219 N Litchfield Rd.
Luke AFB, AZ 8530924 hours a day, 7 days a week (holidays included)PT, aPPT, Fibrinogen, D-DimerDIC are billed as a panel; further testing will be billed seperatelyY (currently through NERL water through Omnicell)NoN/AN/A2 levels every 8 hoursN/ANo
673rd Medical Group, Joint Base Elmendorf-Richardson (JBER), AKCentral8/1/27140 x 100 x 50120V, 60 Hz.NoYES, depending on the size of the new equipment.Lab Department (Core)

5955 Zeamer Ave.

Elmendorf AFB, AK 99506 0730-1630 for phlebotomy and 24hrs for inpatient/ER care N/A N/A Y (currently through NERL water through DMLSS) Yes N/A N/A 2 levels every 8 hours N/A No

Estimated Annual Workload

MTFProthrombin Time (PT)Activated Partial Thromboplastin Time (APTT)FibrinogenD-DimerHeperinLupus AnticoagulantDIC PanelPatient Thrombin TimeMixing StudiesReptilase TimeFactor V AssayFactor VII AssayFactor VIII AssayFactor IX AssayFactor X AssayFactor XI AssayAntithrombin IIIVon Willebrand Factor AntigenFactor II Assay
Madigan Army Medical Center, Joint Base Lewis-McChord, WA120008000100024007015018003525105015110101010000
Naval Hospital Camp Pendleton, CA2000800310600000000000000000
Naval Hospital Bremerton, WA3352111950000000000000000
Robert E. Bush Naval Hospital Twentynine Palms, CA1000800300600000000000000000
99th Medical Group, Mike O'Callaghan Military Medical Center, Nellis AFB, NV810235392701631215200000000000000
60th Medical Group, David Grant Medical Center, Travis AFB, CA4740252024081696000000000000000
Weed Army Community Hospital, Fort Irwin, CA500400100200000000000000000
Naval Medical Center San Diego, CA
- Core Lab140008001240038003720000000300000000
Tripler Army Medical Center, Fort Shafter, HI5891483550811840001120161241124439165
56th Medical Group, Luke AFB, AZ43329600000000000000000
673rd Medical Group, Joint Base Elmendorf-Richardson (JBER), AK14629838468700000000000000
Totals:504633038554071191835541501800464510662745122141439165

File details come from the government source that posted it. Updated .