This document outlines the specifications for a Gas Chromatograph (GC) to be used as an emergency response unit by the Illinois Department of Natural Resources Office of Mines and Minerals Analytical Laboratory. The GC must be capable of quickly and continuously analyzing up to 20 samples per hour to detect permanent gases and light hydrocarbons in mine atmospheres. It must be linked to a computer for data analysis and storage, with easily replaceable or repairable components within each channel. The GC must be able to analyze four channels simultaneously using different column and carrier gas configurations to detect a range of compounds including methane, carbon monoxide, carbon dioxide, oxygen, nitrogen, hydrogen, and various hydrocarbons.
The document specifies detailed technical requirements for the GC, including injection volumes, oven temperature range, detector characteristics, and power and gas supply needs. The GC must also be transportable, weighing less than 40 lbs and occupying no more than 1.5 cubic feet. The procurement includes the GC unit, power supply, computer, software, connections, setup, and training. A trade-in discount is offered for a Varian 4900 micro GC.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| Forms A Section V.18.1~50.docx | DOCX document | |
| Standard Terms and Conditions 1~1.pdf | ||
| Forms B V.20.1~1.docx | DOCX document |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Illinois Department of Natural Resources
Office of Mines and Minerals
Analytical Laboratory
Gas Chromatograph Specifications
This Gas Chromatograph(GC) will be used as an emergency response unit for the detection of permanent gases and light hydrocarbons in mine atmospheres. The GC must be fast enough to run 20 samples per hour and rugged enough to sustain this throughput continuously for several weeks. The GC unit should be linked to a computer for data analysis and storage.
Malfunctioning parts within the GC must be able to be quickly and easily be replaced or repaired independently of other components within each channel. Exact specifications are as follows:
Gas Chromatograph
Software selectable injection volumes of 1 µL to 10 µL
Backflush capable
Oven capable of temperatures of 30 to 180 degrees Celsius
Dual-channel four-filament micro thermal conductivity detector
Linear range of five orders of magnitude or greater
Less than one percent RSD repeatability
Minimum of twenty samples per hour throughput
Single injection for the determination of all compounds
Capable of injection by internal pump or by luer-lock syringe
Individual replaceable components (detector, injector, column, etc.) within each of the four channels
Simultaneous analysis of four channels:
Molesieve 5A with argon carrier
Moleseive 5A with helium carrier
Poraplot U with a helium carrier
Poraplot Q with a helium carrier
Compounds of Interest and Detection Ranges
Methane (10ppm/100%)
Carbon monoxide (10ppm/10%)
Carbon dioxide (10ppm/100%)
Oxygen (100ppm/30%)
Nitrogen (10ppm/100%)
Hydrogen (10ppm/10%)
Ethane (10ppm/15%)
Ethylene (10ppm/10%)
Acetylene (10ppm/10%)
Propane (10ppm/10%)
Propylene (10ppm/10%)
Argon (by calculation) (100ppm/5%)
Power and Gas Supply
Main power 110 volt AC, 60 Hz
Maximum 250W power consumption
Inlet connections for two external gas supplies
GC Controls and Data Analysis
A computer must be included and should control all GC functions
The computer performs data analysis and storage
GC software compatible with Microsoft Office
Transportability
Must be able to be transported by simply disconnecting the power and gas supply
(no internal disassembly or securing)
GC must weigh less than 40 lbs
Maximum size of GC of 1.5 cubic feet
These specifications are the absolute minimum requirements. The system should include the GC unit, power supply, a computer, software, all connections, setup of GC, and training of personnel. Trade-in discount for a Varian 4900 micro GC preferred.