FDA-17-SOL-1186710 – Amendment 0001 Page 1 of 1 Precision and Analytical Lab Balance PM and Calibrations

Solicitation Section Questions and Government Responses Would it be possible to have an onsite visit to get feedback?

Response: Due to time constraints, we are unable to provide onsite visits to all potential offerors.

Part 3, Specific Requirements, #2f

Does this include the adjustment of the internal weight?

Response: No

Requirements, #2j

Unsure as to what you are referencing with pan supports and plastic spacers. Do you have part numbers or images?

Response: Example: Oahu’s Part#11108115

Requirements, #2k

There are no 17025 standards as they do not issue tolerance. Can you provide the expected tolerance?

Response: Tolerance should meet balance manufacturer’s specifications.

Requirements, #3

What is the criteria of the performance test and at what tolerance will be used to ensure calibration by FDA?

Response: If found out of spec after the calibration services are rendered the Contractor will need to come in again to recalibrate.

Tolerance would be balance manufacturer’s specification.